Computer-assisted total knee replacement. A controlled cadaver study using a multi-parameter quantitative CT assessment of alignment (the Perth CT Protocol).

Abstract

A controlled study, comparing computer- and conventional jig-assisted total knee replacement in six cadavers is presented. In order to provide a quantitative assessment of the alignment of the replacements, a CT-based technique which measures seven parameters of alignment has been devised and used. In this a multi-slice CT machine scanned in 2.5 mm slices from the acetabular roof to the dome of the talus with the subject's legs held in a standard position. The mechanical and anatomical axes were identified, from three-dimensional landmarks, in both anteroposterior and lateral planes. The coronal and sagittal alignment of the prosthesis was then measured against the axes. The rotation of the femoral component was measured relative to the transepicondylar axis. The rotation of the tibial component was measured with reference to the posterior tibial condyles and the tibial tuberosity. Coupled femorotibial rotational alignment was assessed by superimposition of the femoral and tibial axial images. The radiation dose was 2.7 mSV. The computer-assisted total knee replacements showed better alignment in rotation and flexion of the femoral component, the posterior slope of the tibial component and in the matching of the femoral and tibial components in rotation. Differences were statistically significant and of a magnitude that support extension of computer assistance to the clinical situation.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

S K Chauhan, G W Clark, S Lloyd, R G Scott, W Breidahl, J M Sikorski. 2004. Computer-assisted total knee replacement. A controlled cadaver study using a multi-parameter quantitative CT assessment of alignment (the Perth CT Protocol).. https://doi.org/10.1302/0301-620x.86b6.15456

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

[Dislocation of artificial cartilage (SaluCartialge)].

Treament of focal cartilage defects is challenging, and techniques of bone marrow stimulation or cartilage transplantation are not sufficient to reconstitute hyaline cartilage. Preliminary results of artificial cartilage repair are promising, but no long-term results are available. Here, we report on a case of graft dislocation 6 months after artificial cartilage repair (SaluCartilage). We conclude that this method is not fully developed and more long-term results are required.

Arthroplasty, Replacement, Knee↗

Visualization of femorotibial contact in total knee arthroplasty using X-ray fluoroscopy.

The purpose of this study was to build a visualization technique of the femorotibial contact in fixed-bearing total knee arthroplasty (TKA) using X-ray fluoroscopy, and to apply this technique to a TKA patient during dynamic motion. In vivo kinametcis of the metallic knee implant was determined using a 2D/3D registration technique, which uses computer assisted design (CAD) model of the implant to estimate the 3D pose of radiopaque metallic femoral and tibial components from a single-plane fluoroscopic image. In fixed-bearing TKA, a 3D pose of radiolucent tibial polyethylene insert can be determined from the estimated pose of the tibial component. To visualize femorotibial contact, the proximity between surfaces of femoral component and tibial insert was calculated, and mapped onto the insert surface model. For the clinical application, dynamic states of contact on the tibial insert were observed including axial rotation and unilateral loading during knee flexion, and post-cam contact of posterior stabilized TKA. The present technique provided us new information and enabled us to better understand the relationship between in vivo knee kinematics and articular shape of the implant.

Arthroplasty, Replacement, Knee↗

A novel method for in vivo knee prosthesis wear measurement.

Wear remains an important cause of failure in knee replacement. Of the current methods of early performance assessment or prediction, simulators have been un-physiological, single X-ray film analyses remain limited by accuracy and retrieval and survival methods have a prohibitive time scale. An accurate method is needed to allow a timely assessment of polyethylene component wear in vivo, when a new design is introduced, in order to predict likely outcome. We present a new method for measuring wear in vivo that we believe will allow this prediction of long-term wear. X-ray film pairs were taken of implanted prosthetic metal components. When the X-ray system was calibrated, projections of the appropriate Computer Aided Design (CAD) model could be matched to the shapes on the scanned X-ray films to find component positions. Interpenetration of the metal femoral component into the polyethylene component could then be established and represents our estimate of "wear". This method was used to measure in vivo prosthesis wear to an accuracy of 0.11 mm.