Electrosurgery in laparoscopy.

Abstract

The two factors that probably contributed most to the bowel burns that have occurred during monopolar laparoscopy were the use of higher-than-necessary voltages and the electrical isolation of the salpinx. Preferably, low-peak voltages should be used. If the two-burn technique is used, desiccation should be from distal to proximal to avoid electrical isolation of the salpinx. Bipolar laparoscopy is unquestionably safer than monopolar, but it is also harder to perform because the salpinx must be grasped repeatedly on both sides to desiccate the required 2 to 3 cm of tissue. Moreover, electrosurgical cutting with bipolar forceps is not very practical because the spark punches through to the opposite jaw and the cutting stops. Bipolar forceps that can make a mechanical cut after desiccation are available for those who desire this feature. Metal trocar sleeves are recommended for use with the monopolar, single-puncture technique only. They are not necessary for the bipolar or double-puncture monopolar techniques.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

F W Harris. 1978. Electrosurgery in laparoscopy.. https://pubmed.ncbi.nlm.nih.gov/151145/

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

9.4T human MRI: preliminary results.

This work reports the preliminary results of the first human images at the new high-field benchmark of 9.4T. A 65-cm-diameter bore magnet was used together with an asymmetric 40-cm-diameter head gradient and shim set. A multichannel transmission line (transverse electromagnetic (TEM)) head coil was driven by a programmable parallel transceiver to control the relative phase and magnitude of each channel independently. These new RF field control methods facilitated compensation for RF artifacts attributed to destructive interference patterns, in order to achieve homogeneous 9.4T head images or localize anatomic targets. Prior to FDA investigational device exemptions (IDEs) and internal review board (IRB)-approved human studies, preliminary RF safety studies were performed on porcine models. These data are reported together with exit interview results from the first 44 human volunteers. Although several points for improvement are discussed, the preliminary results demonstrate the feasibility of safe and successful human imaging at 9.4T.

Burns, Electric↗