Diffusion coefficient measurement in a microfluidic analyzer using dual-beam microscale-refractive index gradient detection. Application to on-chip molecular size determination.

Abstract

We report a microchip-based detection scheme to determine the diffusion coefficient and molecular mass (to the extent correlated to molecular size) of analytes of interest. The device works by simultaneously measuring the refractive index gradient (RIG) between adjacent laminar flows at two different positions along a microchannel. The device, referred to as a microscale molecular mass sensor (micro-MMS), takes advantage of laminar flow conditions where the mixing of two streams occurs essentially by diffusion across the boundary between the two streams. Two flows merge on the microchip, one containing solvent only, referred to as the mobile phase stream and one which contains the analyte(s) of interest in the solvent, i.e. the sample stream. As these two streams merge and flow parallel to each other down the microchannel a RIG is created by the concentration gradient. The RIG is further influenced by analyte diffusion from the sample stream into the mobile phase stream. Measuring the RIG at a position close to the merging point (upstream signal) and simultaneously a selected distance further down the microchannel (downstream signal) provides real-time data related to the extent a given analyte has diffused, which can be readily correlated to analyte molecular mass by taking the ratio of the downstream-to-upstream signals. For the dual-beam RIG measurements, a diode laser output is coupled to a single mode fiber optic splitter with two output fibers. Light from each fiber passes through a graded refractive index (GRIN) lens forming a collimated beam that then passes through the microchannel and then on to a position sensitive detector (PSD). The RIG at both detection positions deflects the two collimated probe beams. The deflection angle of each beam is then measured on two separate PSDs. The micro-MMS was evaluated using polyethylene glycols (PEGs), sugars, and as a detector for size-exclusion chromatography (SEC). Peak purity can be readily identified using the micro-MMS with SEC. The limit of detection was 0.9 ppm (PEG at 11 840 g/mol) at the upstream detection position corresponding to a RI limit of detection (LOD) (3sigma) of 7-10(-8) RI. The pathlength for the RIG measurement was 200 microm and the angular LOD was 0.23 micro(rad) with a detection volume of 8 nl at both positions. The average molecular mass resolution was 9% (relative standard deviation) for a series of PEGs ranging in molecular mass from 106 to 22 800 g/mol. With this excellent mass resolution, small molecules such as monosaccharides, disaccharides, and so on, are readily distinguished. The sensor is demonstrated to readily determine unknown diffusion coefficients.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Colin D Costin, Roy K Olund, Bethany A Staggemeier, Ana Kristine Torgerson, Robert E Synovec. 2003-09-26. Diffusion coefficient measurement in a microfluidic analyzer using dual-beam microscale-refractive index gradient detection. Application to on-chip molecular size determination.. https://doi.org/10.1016/s0021-9673(03)01101-4

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

The influence of grafted polymer architecture and fluid hydrodynamics on protein separation by entropic interaction chromatography.

Entropic interaction chromatography (EIC) provides efficient size-based separation of protein mixtures through the entropy change associated with solute partitioning into a layer of hydrophilic homopolymer that has been end-grafted within the pores of a macroporous chromatography support. In this work, surface-initiated atom-transfer radical polymerization (ATRP) is used to prepare a library of EIC stationary phases covering a wide range of grafted-chain densities and molecular weights. Exhaustive chain cleavage and analysis by saponification and GPC-MALLS, respectively, show that the new ATRP synthesis procedure allows for excellent control over graft molecular weight and polydispersity. The method is used to prepare high-density grafts (up to 0.164 +/- 0.005 chains/nm(2)) that extend the range of EIC applications to include efficient buffer-exchange and desalting of protein preparations. Reducing the graft density allows for greater partitioning of high molecular weight solutes, extending the linear range of the selectivity curve. Increasing graft molecular weight also alters selectivity, but more directly affects column capacity by increasing the volume of the grafted layer. Protein partitioning in high-density EIC columns is found to decrease with mobile-phase velocity (u). Although solute mass transfer resistances leading to an increase in plate height can explain this effect, pressure drop data across the column are indicative of weak convective flow through at least a fraction of the grafted architecture. Modeling of the grafted brush properties in the presence of solvent flow by subjecting a self-consistent-field theory representation of the brush to a viscous shear force predicts that the grafted chains will tilt and elongate in the direction of flow. The shear force may therefore act to reduce the number of conformations available to chains, increasing their rigidity without significantly altering the thickness of the grafted layer. A reduction in protein partitioning is then predicted when the dependence on u of the solute entropy loss is stronger than that of the grafted polymer, a condition met at high graft densities.

Chromatography, Gel↗

Effects of gamma radiation on beta-lactoglobulin: oligomerization and aggregation.

The conformational changes and aggregation process of beta-lactoglobulin (beta-LG) subjected to gamma irradiation are presented. Beta-LG in solutions of different protein concentrations (3 and 10 mg/ml) and in solid state with different water activities (a(w)) (0.22; 0.53; 0.74) was irradiated using a Cobalt-60 radiation source at dose level of 1-50 kGy. Small-angle X-ray scattering (SAXS) was used to study the conformational changes of beta-LG due to the irradiation treatment. The irradiated protein was also examined by high performance size exclusion chromatography (HPSEC) and sodium dodecyl sulfate-polyacrylamide gel electrophoresis (SDS-PAGE) under nonreducing and reducing conditions and fluorescence. SAXS analysis showed that the structural conformation of irradiated beta-LG in solid state at different a(w) and dose level was essentially the same as the nonirradiated beta-LG. The scattering data also showed that the irradiation of beta-LG in solution promoted the formation of oligomers. Interestingly, from the data analysis and model building, it could be shown that the formed oligomers are linear molecules, built by linear combinations of beta-LG dimers (tetramers, hexamers, etc). The formation of oligomers was also evidenced by SDS-PAGE analysis and HPSEC chromatograms, in which products with higher molecular mass than that of the dimeric beta-LG were detected. Formation of intermolecular cross-linking between tyrosyl radicals are proposed to be at least partially responsible for this occurrence. From the results it could be shown that the samples irradiated in solution presented some conformational changes under gamma irradiation, resulting in well ordered oligomers and aggregates formed by cross-linking of beta-LG dimers subunits, while the samples irradiated in the solid state were not modified.

Chromatography, Gel↗

Comparison of two gel filtration chromatographic methods for the purification of Lily symptomless virus.

Lily symptomless virus (LSV) occurs frequently in many Lilium species worldwide and often causes developmental abnormalities such as a smaller flower and lower bulb yield. In this study, two moderate and efficient gel filtration chromatography (GFC) methods was compared, these two techniques were, respectively, based on Superdex-200 HR and Sephacryl S-1000 SF. The products purified by the two methods were then characterized by measurements with UV-spectrophotometer, reverse transcriptase (RT)-PCR, transmission electron microscope (TEM), polyacrylamide gel electrophoresis (PAGE), Western blotting and matrix assisted laser desorption-ionisation/time-of-flight mass spectrometry (MALDI-TOF-MS). The final yield of purified LSV by the Superdex-200 HR GFC was 9.4 mg from 50 g of fresh infected tissues of Lanzhou lily. However, from the same amount tissues, only 5.6 mg of LSV were obtained by using Sephacryl S-1000 SF GFC. The Superdex-200 HR method was thus shown to be more suitable for the purification of LSV than the Sephacryl S-1000 SF GFC. The Superdex-200 HR method does not require costly equipment for density centrifugation and ultracentrifugation. Furthermore, it can provide an economical and efficient way to obtain purified products for the preparation of antibodies for serological diagnosis or LSV infection and related investigations.

Chromatography, Gel↗