IMRT delivery verification using a spiral phantom.

Abstract

In this paper we report on the testing and verification of a system for IMRT delivery quality assurance that uses a cylindrical solid water phantom with a spiral trajectory for radiographic film placement. This spiral film technique provides more complete dosimetric verification of the entire IMRT treatment than perpendicular film methods, since it samples a three-dimensional dose subspace rather than using measurements at only one or two depths. As an example, the complete analysis of the predicted and measured spiral films is described for an intracranial IMRT treatment case. The results of this analysis are compared to those of a single field perpendicular film technique that is typically used for IMRT QA. The comparison demonstrates that both methods result in a dosimetric error within a clinical tolerance of 5%, however the spiral phantom QA technique provides a more complete dosimetric verification while being less time consuming. To independently verify the dosimetry obtained with the spiral film, the same IMRT treatment was delivered to a similar phantom in which LiF thermoluminescent dosimeters were arranged along the spiral trajectory. The maximum difference between the predicted and measured TLD data for the 1.8 Gy fraction was 0.06 Gy for a TLD located in a high dose gradient region. This further validates the ability of the spiral phantom QA process to accurately verify delivery of an IMRT plan.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Susan L Richardson, Wolfgang A Tomé, Nigel P Orton, Todd R McNutt, Bhudatt R Paliwal. 2003. IMRT delivery verification using a spiral phantom.. https://doi.org/10.1118/1.1603965

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Characteristics of dosemeter types for skin dose measurements in practice.

A growing number of papers report deterministic effects in the skin of patients who have undergone interventional radiological procedures. Dose measurements, and especially skin dose measurements, are therefore increasingly important. Methods and acceptable dosemeters are, however, not clearly defined. This paper is the result of a literature overview with regard to assessing the entrance skin dose during radiological examinations by putting a dosemeter on the patient's skin. The relevant intrinsic characteristics, as well as some examples of clinical use of the different detector types, are presented. In this respect, thermoluminescence, scintillation, semiconductor and film dosemeters are discussed and compared with respect to their practical use.

Film Dosimetry↗

Dosimetry limitations and a dose correction methodology for step-and-shoot IMRT.

For the step-and-shoot intensity-modulated radiation therapy (IMRT) technique, the combination of high dose rate, multiple beam segments and low dose per segment can lead to significant differences between the planned dose and the dose delivered to the patient. In this technique, a dose delivery inaccuracy known as the 'overshoot' effect is caused by the dose servo control system. This typically occurs in the first and last beam segments and causes an over- and underdose, respectively. Some dose positional error in the segment sequence is also possible there. Commercial ionization chambers (RK-type) and radiographic Kodak films were used for the measurements. The reported results were obtained using the Pinnacle(3)-V6.2 treatment planning system and a Varian Clinac 21 EX linear accelerator equipped with a 120-leaf Millennium MLC. The dose inaccuracy measurements were based on the comparison of the dose and profiles for reference fields and fields irradiated with the step-and-shoot technique. For our linear accelerators, an 'overshoot' effect ranging from 0.1 to 0.6 MU was found, corresponding to a dose rate from 100 to 600 MU min(-1), respectively. For segments with off-axis distances from 0 to 5.5 cm with >3.5 MU per segment and all dose rates, a MLC leaf-position error of <1 mm was measured. For segments with an off-axis distance of 9.5 cm, a positional error >2 mm was measured for 600 MU min(-1) and 1 MU per segment. The purpose of this study was to find a correction method for segmental dose errors caused by the 'overshoot' effect when small monitor unit and high dose rate are used. To better represent the fluctuation of the segment doses in the beam, a dose ratio between reference and step-and-shoot irradiated fields was defined. A method for the correction of segment dose inaccuracies and a quality assurance programme for the 'overshoot' effect were developed. The ordering of the biggest segment shape in the segment sequence was studied for ten randomly selected prostate patients planned for IMRT. The results of this work can be used to improve the agreement between the planned and delivered doses for IMRT.

Film Dosimetry↗