PubMed · 12948131
Interference-free analysis using three-way fluorescence data and the parallel factor model. Determination of fluoroquinolone antibiotics in human serum.
Abstract
Three-way fluorescence data and multivariate calibration based on parallel factor analysis (PARAFAC) are combined for the simultaneous quantitation of three fluoroquinolone anitibiotics (norfloxacin, enoxacin, and ofloxacin) in human serum samples. The three analytes can be adequately determined with limits of detection of 0.2, 3.0, and 0.5 microg L(-1), respectively, with minimum experimental effort. The selected analytical methodology fully exploits the so-called second-order advantage of the employed three-way data, allowing obtaining individual concentrations of calibrated analytes in the presence of any number of uncalibrated (serum) components. In contrast to PARAFAC, less satisfactory results were obtained with a multidimensional partial least-squares (nPLS) model trained with the same calibration set.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Arsenio Muñoz de la Peña, Anunciación Espinosa Mansilla, David González Gómez, Alejandro C Olivieri, Héctor C Goicoechea. 2003-06-01. Interference-free analysis using three-way fluorescence data and the parallel factor model. Determination of fluoroquinolone antibiotics in human serum.. https://doi.org/10.1021/ac026360h
Cite the original work for its findings. Save a collection to share your selection of sources.