[Accidental decrease in the air flow during air/oxygen sevoflurane anesthesia].

Abstract

We describe our experience of an accidental decrease in the air flow during air/oxygen sevoflurane anesthesia. According to the malfunction of an air compressor system, dehydration of the compressed air was inadequate, and the wet air was delivered to the anesthesia machine. The moisture was formed at a flow control valve to cause a gradual decrease in the air flow. The low-pressure alarm did not work at that time, because the pressure of the air supply was normal. There are two types of medical air, one is produced from the ambient air by air compressor system and another is the synthetic air which is a mixture of oxygen and nitrogen. Although the synthetic air is always dry and clean, the former contains dust, bacteria, and moisture. There is a possibility that all of these particles cannot be excluded before use. Accordingly, we must check the air compressor system routinely when a source of the compressed air is used.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

M Okuyama, I Nakamura, O Kemmotsu. 2000. [Accidental decrease in the air flow during air/oxygen sevoflurane anesthesia].. https://pubmed.ncbi.nlm.nih.gov/10885247/

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Volatilisation of triallate as affected by soil texture and air velocity.

The rate of volatilisation of the formulated herbicide triallate was investigated in a wind tunnel under controlled wind-speed conditions. An experimental set-up is described which allows the monitoring of wind speed (w.s.), soil-water content, and the temperature of air and soil. A system controlling soil-water content is also described. The influence of air velocity and soil texture was investigated measuring the cumulative volatilisation losses of triallate from soil. The herbicide volatilisation losses after application ranged from 40% at 3 m/s to 53% at 9 m/s for loam soil and from 60% at 3 m/s to 73% at 9 m/s for sandy soil.

Air↗

"Oxidative stress" response in submerged cultures of a recombinant Aspergillus niger (B1-D).

A recombinant strain of Aspergillus niger (B1-D), engineered to produce the marker protein hen egg white lysozyme, was investigated with regard to its susceptibility to "oxidative stress" in submerged culture in bioreactor systems. The culture response to oxidative stress, produced either by addition of exogenous hydrogen peroxide or by high-dissolved oxygen tensions, was examined in terms of the activities of two key defensive enzymes: catalase (CAT) and superoxide dismutase (SOD). Batch cultures in the bioreactor were generally found to have maximum specific activities of CAT and SOD (Umg x protein(-1)) in the stationary/early-decline phase. Continuous addition of H2O2 (16 mmole L(-1) h(-1)), starting in the early exponential phase, induced CAT but did not increase SOD significantly. Gassing an early exponential-phase culture with O2 enriched (25 vol%) air resulted in increased activities of both SOD and CAT relative to control processes gassed continuously with air, while gassing the culture with 25 vol% O2 enriched air throughout the experiment, although inducing a higher base level of enzyme activities, did not increase the maximum SOD activity obtained relative to control processes gassed continuously with air. The profile of the specific activity of SOD (U mg CDW(-1)) appeared to correlate with dissolved oxygen levels in processes where no H2O2 addition occurred. These findings indicate that it is unsound to use the term "oxidative stress" to encompass a stress response produced by addition of a chemical (H2O2) or by elevated dissolved oxygen levels because the response to each might be quite different.

Air↗

Cyclic fatigue-crack propagation in sapphire in air and simulated physiological environments.

Single-crystal aluminas are being considered for use in the manufacture of prosthetic heart valves. To characterize such materials for biomedical application, subcritical crack growth by stress corrosion (static fatigue) and by cyclic fatigue has been examined in sapphire along (1100) planes in 24 degrees C humid air and 37 degrees C Ringer's solution (the latter as a simulated physiological environment). The relationships between crack-propagation rates and the linear-elastic stress intensity have been determined for the first time in sapphire for both modes of subcritical cracking. It was found that growth rates were significantly faster at a given stress intensity in the Ringer's solution compared to the humid air environment. Mechanistically, a true cyclic fatigue effect was not found in sapphire as experimentally measured cyclic fatigue-crack growth rates could be closely predicted simply by integrating the static fatigue-crack growth data over the cyclic loading cycle.

Air↗