Abstract

The aim of this study was to improve the reliability of pedicle screw insertion. Transpedicle screw insertion may cause neurological, vascular, and mechanical complications. Previous studies of surgical procedures have shown a significant rate of incorrect placement of the screw ranging from 10 to 40%. A new technique that combines preoperative computed tomography (CT) imaging with intraoperative passive navigation was used to perform 64 pedicle screw insertions in the thoracolumbar region. At the same time, 64 pedicle screw insertions were performed manually in the same region and on the same vertebral levels. Surgery was followed in all cases by postoperative radiographs and computed tomography examination, which allowed measurements of screw position relative to pedicle position to be performed. A comparison between the two groups showed that six screws in 64 vertebra (9%) had incorrect placement with the computer-assisted technique whereas 28 screws in 64 vertebra (44%) had incorrect placement with manual insertion. The intraoperative accuracy provided by the computer after registration was better than 1 mm. The good results obtained are similar to those reported in the literature. The cortex penetration observed with the computer-assisted technique was not imputed to computer failure. Errors by the surgeon in acquiring data in the pre- and perioperative steps may explain the six incorrect screw placements. This clinical experience confirms that the accuracy and the reliability of this computer-assisted technique are good.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

P Merloz, J Tonetti, L Pittet, M Coulomb, S Lavallée, J Troccaz, P Cinquin, P Sautot. 1998. Computer-assisted spine surgery.. https://doi.org/10.1002/(sici)1097-0150(1998)3%3A6%3C297%3A%3Aaid-igs3%3E3.0.co%3B2-8

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Experimental validation of noninvasive referencing in navigated procedures on long bones.

Navigation procedures in orthopedic surgery require fixation of reference markers to the anatomic region of interest. Inadequate fixation might lead to micromotion or loosening of the reference marker, consequently causing registration failures or errors in navigation. Osseous rigid fixation is usually achieved by minimally invasive Schanz screws or pins. The goal of this study was to evaluate a non invasive external fixation device, a headband so far used in cranial navigation, as an alternative invasive fixation technique to reference markers in the femur. A common navigation system with an adapted trauma software application was used to track the positions of the soft tissue-attached headband relative to an invasive reference marker on the femur during manipulations of the thigh. Relative translative and rotational movements of the headband were measured during defined movements of the hip and knee and manipulations of the headband itself. The results revealed high translative and rotational movements, up to 6 mm and 3 degrees , respectively, due to minor manipulations of the affected lower extremity. Noninvasive soft tissue fixation with a headband does not allow rigid fixation for accurate navigated registration or operative procedures at the femur. Necessary intraoperative movements or manipulations would cause substantial registration failures. Invasive fixation techniques with screws or pins are still the method of choice.

Bone Screws↗

Safety of computer-assisted surgery for cannulated hip screws.

Computer-assisted orthopaedic surgery has developed considerably during the past few years. Several manufacturers produce hardware and software for use in trauma surgery. Validation of these systems before clinical application is mandatory to be sure they work accurately and safely. The accuracy of surgical performance is highly correlated with the cut-out percentages of hip screws. In a standardized operative setting, three cannulated hip screws were inserted in each of 20 sawbones. The screws were positioned either by fluoroscopic navigation technique or by conventional operative technique depending on randomization. Our primary aim was to assess whether computer-navigated screw fixation is equally safe compared with conventional screw fixation using fluoroscopy. To determine safety, we investigated number of drilling attempts, screw position, and radiation time. Secondary to these safety parameters, we also compared the operating time between the two procedures to assess the efficiency of computer navigation. Statistical analysis showed no differences regarding accuracy of screw position. Computer-assisted surgery resulted in fewer drilling attempts and less radiation time, with a similar operation time. We believe the currently used navigation system is safe and accurate.

Bone Screws↗

Techniques of posterior C1-C2 stabilization.

Instability of the atlantoaxial complex may result from inflammatory, traumatic, congenital, neoplastic, or degenerative disorders and often requires surgical stabilization. Initial dorsal wiring techniques allow safe fixation but require rigid external immobilization and have been associated with high fusion failure rates. Rigid screw fixation techniques including transarticular screw fixation and C1-C2 rod-cantilever fixation offer higher fusion rates and less need for rigid immobilization but are more technically demanding. C1-C2 fixation using crossing C2 laminar screws offers rigid fixation but without the technical demands of C2 pars placement. The history and techniques of dorsal fixation of the atlantoaxial complex are reviewed, and the success rates and complications of each are discussed.

Bone Screws↗