Computerized bone densitometric analysis: operator-dependent errors.

Abstract

PURPOSE: To determine the nature and relative frequency of operator-dependent data analysis errors in dual x-ray absorptiometry. MATERIALS AND METHODS: Over 40 months, 2,528 dual x-ray absorptiometric examinations of the forearm, femoral neck, and lumbar spine were performed by 11 technologists by using standard techniques and software. Each analysis was reviewed by a radiologist; errors were recorded and corrected. RESULTS: There were no forearm analysis errors. There were 24 (0.9%) femoral neck analysis errors, of which 23 resulted from misplacement of the analysis region. There were 33 (1.3%) spinal analysis errors, of which 24 resulted from misplacement of intervertebral disk space markers. Analysis errors of the femur and spine resulted in six misdiagnoses (0.2%). CONCLUSION: Misdiagnosis due to analysis errors is rare. Femoral neck analysis errors were easily detectable, but accurate spinal analyses depended on accurate identification of vertebral end plates and posterior elements. Nonetheless, these potentially serious errors can be detected and corrected if the analyses are reviewed and interpreted by a supervising physician who is familiar with the relevant anatomy, proper analysis techniques, and factors--such as artifacts--that adversely affect the accuracy of the analysis.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

R B Staron, R Greenspan, T T Miller, J P Bilezikian, E Shane, N Haramati. 1999. Computerized bone densitometric analysis: operator-dependent errors.. https://doi.org/10.1148/radiology.211.2.r99ma55467

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

A repeated measures concordance correlation coefficient.

The concordance correlation coefficient is commonly used to assess agreement between two raters or two methods of measuring a response when the data are measured on a continuous scale. However, the situation may arise in which repeated measurements are taken for each rater or method, e.g. longitudinal studies in clinical trials or bioassay data with subsamples. This paper proposes a coefficient for measuring agreement between two raters or two methods of measuring a response in the presence of repeated measurements. We illustrate the methodology with examples comparing (1) 1-hr versus 2-hr blood draws for measuring cortisol in an asthma clinical trial and (2) two measurements of percentage body fat, from skinfold calipers and dual energy X-ray absorptiometry.

Absorptiometry, Photon↗

Crystallization and preliminary X-ray analysis of a domain in the Runx2 transcription factor that interacts with the 1alpha,25 dihydroxy vitamin D3 receptor.

The Runx2 transcription factor is a key regulator of osteoblast differentiation. In response to 1alpha,25 dihydroxy vitamin D3, Runx2 may interact with the 1alpha,25 dihydroxy vitamin D3 receptor (VDR) in the promoter of target genes, producing a synergic activation of their transcription. Previous studies have suggested that the motifs responsible for the VDR-Runx2 interaction are contained within the 230-361 domain of Runx2. In this work, we confirmed by GST-pull down that Runx2(I(209-361)) is sufficient to interact with the VDR. To obtain structural information, GST-Runx2(I(209-361)) protein was overexpressed in Escherichia coli, purified and crystallized using the hanging-drop vapor-diffusion method and polyethyleneglycol as a precipitant. The crystals were found to diffract to a maximum resolution of 2.7 A and a complete data set to a 3.3 A resolution was collected and analyzed. The crystals belong to the tetragonal system, with a space group P4 and unit-cell parameters of a = b = 90.8, and c = 57.2 A. The presence of a monomer of the recombinant GST-Runx2(I(209-361)) in the asymmetric unit gives a V(M) of 2.7 A(3) Da(-1) and a solvent content of 54.8%.

Absorptiometry, Photon↗

Quantifying the treatment effect explained by markers in the presence of measurement error.

Surrogate markers or intermediate markers are important in identifying subjects with high risk of a serious disease or for monitoring disease progression of a subject on treatment. Quantifying the proportion of treatment effect (PTE) explained by markers has been studied extensively. Due to reasons such as biological variation, limited machine precision, etc. markers are generally measured with error. The estimated PTE ignoring the measurement error could be biased, which may lead to incorrect conclusions. In this article, we adjust for the measurement error using regression calibration to construct a less biased estimator of excess relative odds, a quantity to measure the treatment effect explained by markers. The method is applied to data from a clinical study in osteoporosis.

Absorptiometry, Photon↗