Fully automated analytical method for codeine quantification in human plasma using on-line solid-phase extraction and high-performance liquid chromatography with ultraviolet detection.

Abstract

A simple, sensitive and fully automated analytical method for the analysis of codeine in human plasma is presented. Samples are added with oxycodone, used as internal standard (I.S.), and directly loaded in the autosampler tray. An on-line sample clean-up system based on solid-phase extraction (SPE) cartridges (Bond-Elut C2, 20 mg) and value switching (Prospekt) is used. Isocratic elution improved reproducibility and allowed the recirculation of the mobile phase. A Hypersil BDS C18, 3 microns, 10 x 0.46 cm column was used and detection was done by UV monitoring at 212 nm. Retention times of norcodeine (codeine metabolite), codeine and oxycodone (I.S.) were 5.5, 6.4 and 9.1 min, respectively. Morphine was left to elute in the chromatographic front. Detection limit for codeine was 0.5 microgram l-1 and inter-assay precision (expressed as relative standard deviation) and accuracy (expressed as relative error) measured at 2 micrograms l-1 were 5.03% and 1.82%. Calibration range was 2-140 micrograms l-1.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

J A Pascual, J Sanagustín. 1999-03-19. Fully automated analytical method for codeine quantification in human plasma using on-line solid-phase extraction and high-performance liquid chromatography with ultraviolet detection.. https://doi.org/10.1016/s0378-4347(99)00015-8

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

A robotics-based automated assay for inorganic and organic phosphates.

Phosphate analyses are fundamental to a broad range of biochemical applications involving inorganic phosphate and organic phosphoesters such as phospholipids, phosphorylated proteins, and nucleic acids. A practical automated method utilizing robotics is described in this report. Five colorimetric methods of phosphate analyses based on formation of a phosphomolybdate complex and compatible with the automated assay were tested, and the fundamental chemistry is discussed. The relative sensitivities are malachite green > crystal violet > quinaldine red > ascorbate reduction > antimony-modified ascorbate reduction, although only a fourfold improvement was observed in going from the modified ascorbate procedure to malachite green. Malachite green was selected to optimize the assay because this dye provided the highest sensitivity. However, where color stability and low blanks are more important than sensitivity, the ascorbate reduction and quinaldine red methods were found to be better choices than malachite green. Automation using a robotic liquid-handling system substantially reduces the labor required to process large arrays of samples. The result is a sensitive, nonradioactive assay of inorganic phosphate with high throughput. A digestion step in an acid-resistant 96-well plate was developed to extend the assay to phosphate esters. The robotic-based assay was demonstrated with inorganic phosphate and a common phospholipid, phosphatidylcholine.

Automation