Interinstitutional experience in verification of external photon dose calculations.

Under the auspices of NCI contracts, four institutions have collaborated to assess the accuracy of the pixel-based dose calculation methods they employ for external photon treatment planning. The approach relied on comparing calculations using each group's algorithm with measurements in phantoms of increasing complexity. The first set of measurements consisted of ionization chamber measurements in water phantoms in normally incident square fields, an elongated field, a wedged field, a blocked field, and an obliquely incident beam. The second group of measurements was carried out using thermoluminescent dosimeters in phantoms designed to investigate the effects of surface curvature, high density heterogeneities, and low density heterogeneities. The final study tested the entire treatment planning system, including CT data conversion, in an anthropomorphic phantom. Overall, good agreement between calculation and measurements was found for all algorithms. Regions in which discrepancies were observed are pointed out, areas for algorithm improvement are identified and the clinical import of algorithm accuracy is discussed.

Humans

Abutment of high energy electron fields.

The problem of positioning multiple electron fields on the skin surface is complicated by the fact that the scattering properties of electrons result in constriction of the higher value isodoses and bulging out of the lower value isodoses. We have studied the dose distributions in the junction region for adjacent electron fields using our 3-dimensional radiation treatment planning system by investigating the effects of field separation for several field size, air gap, and beam energy combinations. Film dosimetry was used to measure specific test cases. Using our 3-dimensional radiation treatment planning system, we are able to display the calculated dose distribution, calculate dose-volume histograms, and compare calculations with the film dosimetry measurements. Results of our study are presented.

Electrons

Verification data for electron beam dose algorithms.

The Collaborative Working Group (CWG) of the National Cancer Institute (NCI) electron beam treatment planning contract has performed a set of 14 experiments that measured dose distributions for 28 unique beam-phantom configurations that simulated various patient anatomic structures and beam geometries. Multiple dose distributions were measured with film or diode detectors for each configuration, resulting in 78, 2-D planar dose distributions and one, 1-D depth-dose distribution. Measurements were made for 9- and 20-MeV electron beams, using primarily 6 x 6- and 15 x 15-cm applicators at several SSDs. Dose distributions were measured for shaped fields, irregular surfaces, and inhomogeneities (1-D, 2-D, and 3-D), which were designed to simulate many clinical electron treatments. The data were corrected for asymmetries, and normalized in an absolute manner. This set of measured data can be used for verification of electron beam dose algorithms and is available to others for that purpose.

Algorithms