Search PubMed⌕ Search

Biomedical subjects

T V Glotzer

Publications and source records attributed to T V Glotzer.

6 recordsLinked to original sources

Variability of atrial sensing during exercise in a patient with a dual chamber ICD caused inappropriate ICD discharges.

We present the case of a patient with a dual chamber implantable cardioverter defibrillator (ICD) who experienced inappropriate ICD discharges during exercise. Interrogation of the ICD revealed intermittent atrial undersensing during exercise that was responsible for the erroneous classification by the ICD of sinus tachycardia as ventricular tachycardia. Monitoring of the intracardiac electrograms and Marker Channels during an exercise test confirmed a marked decrease in P wave amplitude during exercise. By increasing the atrial sensitivity setting the problem was resolved.

Adult↗

Electromagnetic interference from a muscle stimulation device causing discharge of an implantable cardioverter defibrillator: epicardial bipolar and endocardial bipolar sensing circuits are compared.

This case report is about two patients with two different types of ICDs who underwent electrical muscle stimulation (EMS) therapy. In one patient with an ICD that has epicardial screw-in bipolar sensing leads, electromagnetic interference (EMI) from the EMS device caused the delivery of an inappropriate ICD discharge. In a second patient with an ICD with endocardial true bipolar sensing, there was no evidence of EMI during the EMS therapy despite all of our attempts to reproduce it. The sensing circuits in the two different ICDs are compared.

Aged↗

Atrial arrhythmia following a biatrial approach to mitral valve surgery.

The biatrial approach to exposing the mitral valve during surgery has the potential for improving visualization of the valve with minimal cardiac manipulation. This procedure, involving a right atriotomy and an extended transseptal incision, may isolate the sinus node from its normal blood supply and autonomic innervation. Thirty-eight consecutive patients undergoing this procedure were examined. Twenty-two of these patients (58%) were admitted in normal sinus rhythm and 15 (40%) were in atrial fibrillation (AF) or atrial flutter. Of the 22 patients admitted in normal sinus rhythm, only 3 patients remained in this rhythm at discharge. Fourteen of the 22 patients were discharged in a slow, low atrial rhythm. All of the patients admitted in AF were discharged in AF. Of the 14 patients discharged in a low atrial rhythm, the rhythm persisted in eleven patients (80%) at a mean of 6-month follow-up. The routine use of this transseptal approach to mitral valve surgery needs further assessment in light of the predictable loss of the sinus mechanism.

Adult↗

Mapping reentry around atriotomy scars using double potentials.

Supraventricular arrhythmias, often seen in patients after cardiac surgery, may be associated with scars produced in the atria at the time of surgery. Double potentials, found in the presence of functional or anatomical block, can define the limits and critical regions of a reentrant circuit associated with the atriotomy scars. We describe six patients with seven distinct atrial tachycardias in whom atriotomy scars were successfully mapped during intraatrial reentry utilizing the presence and interelectrogram relationship of observed double potentials. The reentrant circuit was mapped in all patients by following the relationship between double potentials along the surgical scar, assuming that they would be widely split in the middle of the scar and merge into a single continuous fractionated potential at the apex of the scar. At this site, atrial pacing was performed to entrain the tachycardia and confirm the participation of the atriotomy scar in the clinically relevant atrial tachycardia. Radiofrequency ablation was performed from the site of electrogram fusion to the nearest anatomical obstacle. Five of seven atrial tachycardias were successfully ablated utilizing this technique over a mean follow-up of 10 months. We proposed that these double potentials and their interelectrogram relationship are an effective means of mapping atriotomy scars and guiding successful radiofrequency ablation.

Action Potentials↗

Relationship of fluorescence polarization to cell lineage in lymphocytes from normal subjects and patients with chronic lymphocytic leukemia.

Previous investigations have shown differences in fluorescence polarization between normal and chronic lymphocytic leukemia lymphocytes following incubation with the probe 1,6-diphenyl-1,3,5-hexatriene. In the present study, we determined the fluorescence polarization of unseparated or enriched subpopulations of T and B lymphocytes from normal subjects and patients with chronic lymphocytic leukemia. As had been observed by others, the mean polarization (P) value at 25 degrees C for unseparated chronic lymphocytic leukemia lymphocytes, .240 +/- .007 (N = 22), was lower than that of unseparated normal lymphocytes, .248 +/- .005 (N = 18), P less than .001 (Student's t-test). The difference was greater when B-enriched populations were compared. The mean P value of B-cell-enriched chronic lymphocytic leukemia lymphocytes, .240 +/- .007 (N = 5), was significantly lower than that of B-cell-enriched normal preparations, .256 +/- .004 (N = 5), P less than .001. In contrast, no significant difference was found between normal and chronic lymphocytic leukemia T cells. The anomalous fluorescence polarization manifested by chronic lymphocytic leukemia lymphocytes of B-cell origin serves to distinguish this lineage from its normal counterpart and from T cells of either source.

B-Lymphocytes↗