Certain imprecisions in the bubble point measurement.
Explore the source record for details and available documents.
Biomedical subjects
Publications and source records attributed to T H Meltzer.
Explore the source record for details and available documents.
Explore the source record for details and available documents.
Current efforts at filter validation involve the need to test under "worst case conditions" the bacterial retentive action of filters utilizing the very drug being processed. These concerns recognize that the adsorptive retentions of organisms vary in their efficiencies, depending upon the filtration conditions and the presence or absence of surfactant. The sieve retention of organisms is independent of such conditions, deriving solely from considerations of organism-size, pore-size relationships. Flow-decay studies, can serve to differentiate between filter actions based on sieve-retentions and adsorptive-retentions. A filter's dependence solely upon sieving effects is demonstrable. What is involved is a plotting of flow rates as a function of time in accordance with established mathematical equations. Appropriate plots would indicate the exclusive sieve retention of organisms, a condition of absolute filter reliability deriving from organism-pore size relationships. The clear demonstration by a filter company that its sterilizing filter retains organisms solely by the sieve retention mechanism would suffice to validate that filter for all pharmaceutical filtrations, eliminating the need for individual user validations. The advantage of this mode of filter validation, one based on structurally-inherent filter pore architecture, is that pharmaceutical manufacturers would be free from the need to validate a given filter for each and every pharmaceutical preparation. This would also simplify the FDA's confirmation of filter validation, since a single inspection of the filter manufacturer's records would suffice for all the filter's users.
The more certain implications to retention based upon definition of the largest pores present in a filter may be had where bubble point testing is possible. Where diffusive airflows interfere with bubble point determinations, the diffusive airflow measurements themselves, correlated to organism retentions, serve as integrity tests. Diffusive airflow measurements have their advantage in addition to serving as indicators of filter integrity. They can more precisely reveal filter incompatibilities. They can be used to gauge the completeness of a filter's wettability, and they are more revealing of a filter's clean water flow properties. In almost all its applications, however, diffusive airflow integrity testing requires defining the diffusive airflow curve by way of multipoint measurements using the liquid product in question. This is especially true in filter validation contexts where product-specific testing is necessitated. Once the full diffusional airflow curve has been elucidated, single-point diffusional flow testing is warranted, as in production integrity testing settings. Single point diffusive flow testing can be used as a standard integrity test method for use in the pharmaceutical production environment.
There are two recognized mechanisms whereby organisms are retained by liquid filters; namely, sieve-retention and adsorption. The efficiency of each may be influenced by the organism, suspending milieu, and by the filtration conditions. Validations of sterilizing filtrations require the use of organism suspensions in product-specific media. However, where the product is bactericidal to the challenge organism(s), surrogate solutions may be required. The ideal surrogate solution would minimize adsorptive retention, ensuring that the sterilizing action of the filter under consideration is the consequence of sieve-retention. This review explores the impact that various physicochemical factors may have on bacterial cell size and cell surface characteristics. An understanding of interactions among challenge bacteria, suspending fluid, and filter medium is essential for the development of surrogate solutions that provide a "worst case" mileu for filter validation studies or a "placebo," non-inhibitory challenge solution.
Explore the source record for details and available documents.
Explore the source record for details and available documents.
The pressure-drop/hold procedure enables the diffusive flow integrity testing of filters to be performed without breaching the system downstream of the filter. It is not necessary to measure volumetrically the diffused gas on the downstream side of the filter. By means of pressure transducers the pressure loss is determined upstream; thus eliminating the threat of sepsis due to down-stream invasions. The pressure decay exercise can be used to characterize the various filter types. A constancy of filter manufacture is required for a given filter type. Unless the pressure drop exceeds the value established as the maximum allowable decay, the filter is judged to be integral. It qualifies as a sterilizing grade filter. Excessive pressure decays will also eventuate from leaks, as from improperly sealed housings. Performed prior to the filtration, the procedure serves to eliminate the wasteful use of an imperfect system, whether caused by faulty sealing, incorrect filter type or flawed filters. Where leaks are detected, the filter can be reexamined for its integrity. To enable the pressure-drop procedure to serve as an integrity test, the measured pressure decays require being correlated with organism retention data. This is made possible by the arithmetic conversion of the pressure decay curve into the conventional diffusive airflow curve established to have such a correlation. The transformation of the pressure-drop curve into the differential airflow plot is automatically performed by certain of the automated integrity test machines. These devices, utilizing pressure transducers, are capable of measuring small pressure drops with requisite sensitivity; gauges commonly are not. Moreover, as previously stated, the measurements are advantageously made on the upside of the filter. The use of automated test machines is, therefore, recommended for the performance of the pressure hold/drop integrity test in furtherance of the practice of filter integrity testing.