The design of attendant propelled wheelchairs.

The attendant operated wheelchair is propelled by applying forces to handles at the rear of the chair. There are no published data to justify the design of pushing handles on existing wheelchairs. In Dundee, studies of pushing have been conducted in order to obtain subjective preferences for location and design of handles and an understanding of biomechanical factors associated with wheelchair pushing. Preferred positions for handles have been found to be in the region of 0.75 of shoulder height, 1.14 times shoulder width although deviations of +/- 5% in these values are still rated as acceptable. The preferred positions do not correspond to minimum levels of resultant force or with lowest levels of moment in any of the upper body joints. Moments occurring at the lower back are not substantially affected by handle position. The biomechanical analysis so far has not revealed why some handle positions are more comfortable for pushing than others. Further study, involving calculation of resultant moments (rather than just sagittal plane moments) at these joints and at the lower body joints, is a next step in attempting to find the indicators of discomfort. Transferring a patient from or to a wheelchair can be a difficult operation with risks of accidents to the patient through falling and risks to the attendant of strain, particularly to the back. Current footrests on wheelchairs are a major source of the problems during transfer. A new approach to footrest design is described which solves these difficulties by using a footrest that lowers onto the floor. This has other attractive features such as providing good stability and restraint of the chair during transfer.(ABSTRACT TRUNCATED AT 250 WORDS)

Biomechanical Phenomena↗

Design and evaluation of footrests for hospital wheelchairs.

Transferring a patient to or from a hospital wheelchair can be a difficult operation, with some risk to the patient and to the attendant. Studies are described in which the footrests of hospital wheelchairs are assessed and improved in order to reduce the problems of patient transfer. Four different types of footrest were fitted to hospital ward chairs and three types to transit chairs; these were assessed and compared by hospital ward staff and porters respectively. Based on the findings of these trials, a new type of footrest was designed for use on both types of chair. The new footrest lowers down to the ground and is operated by a footlever.

Equipment Design↗

Measurement of the turning, rolling and obstacle resistance of wheelchair castor wheels.

The turning resistance and rolling resistance of a range of wheelchair wheels up to 200 mm in diameter were measured on three types of indoor surface. Rolling resistance values were found by measuring deceleration on a flat surface and also by direct measurement on a treadmill. The horizontal forces required to push the front wheels of a wheelchair over small step obstacles were also found. Measurements were made for a variety of front wheels and the relationship between the horizontal force, step height, load, and type compressibility was investigated.

Biomedical Engineering↗

Shape memory alloy fixator system for suturing tissue in minimal access surgery.

A new technique for suturing human tissue is described in which tissue closure is achieved by means of small fixators made from shape memory alloy. The aim of the development is to provide an alternative to thread suturing in minimal access surgery, which is quicker and requires less skill to achieve the required suturing quality. The design of the fixators is described in terms of the thermal shape recovery of shape memory alloy and a novel form of finite element analysis, which uses a nonlinear elastic element for the material property. Thermal analysis of the fixators and surrounding tissue is used to predict the temperature distribution during and after the application of electric current heating. This was checked in an in vitro experiment, which confirmed that deployment caused no detectable collateral damage to surrounding tissue. In vivo animal studies on the use of the shape memory alloy fixator for suturing tissue are ongoing to establish safety and healing effects.

Alloys↗