Search PubMed⌕ Search

Biomedical subjects

S Saksena

Publications and source records attributed to S Saksena.

At least 91 records · Page 5Linked to original sources

Interaction between electronic article surveillance systems and implantable defibrillators: insights from a fourth generation ICD.

We present a case report of an inappropriate discharge from a fourth generation implantable cardioverter defibrillator (ICD) as a result of exposure to electromagnetic interference. A 60-year-old man implanted with a Medtronic 7219D defibrillator experienced shocks without preceding symptoms while walking through an electronic surveillance system in a store. Though this has been reported, the mechanism of the interaction remains unexplained. The electrogram storage capabilities of this particular device enabled us to establish that this resulted from inappropriate sensing of the electromagnetic interference.

Defibrillators, Implantable↗

Internal atrial defibrillation: effect on sinus and atrioventricular nodal function and implanted cardiac pacemakers.

Internal atrial defibrillation (IAD) has been extensively evaluated for clinical efficacy but the need for concomitant demand pacing and the effect of IAD shocks on pacemaker function is not well studied. We prospectively evaluated: (1) the incidence of bradycardia as a result of IAD shocks; and (2) effect of these shocks on functioning of implanted cardiac pacemakers. Consecutive consenting patients with atrial fibrillation (AF) requiring cardioversion or undergoing electrophysiological study were selected for IAD. IAD shocks were delivered using the right ventricle to right atrium (RV-RA), right ventricle to superior vena cava (RV-SVC), right atrium to axillary patch (RA-AX), and right atrium to left pulmonary artery or coronary sinus (RA-LPA/CS) lead configurations. Mean RR interval before and after the shocks and the time interval from shock delivery to first QRS complex were analyzed for unsuccessful and successful shocks. Pacing and sensing function was analyzed in patients with previously implanted pacemakers. Twenty-five patients, 18 men, mean age 67.9 +/- 10 years were included in the study. A total of 305 shocks (264 unsuccessful, 41 successful) were analyzed. For unsuccessful shocks the mean post-IAD shock RR interval (795 +/- 205 ms) and the time to first post-IAD shock QRS complex (970 +/- 438 ms) were both significantly greater than the pre-IAD shock RR interval (685 +/- 131 ms, P < 0.001). The increase in post-IAD shock RR interval and time to first post-IAD shock QRS complex was seen with all four lead configurations used. With successful shocks the mean post-IAD shock sinus cycle length (1,105 +/- 450 ms) and time to first post-IAD shock QRS complex (1,126 +/- 443 ms) were both also significantly greater than the pre-IAD shock RR interval (766 +/- 172 ms). Nine patients (36%) had episodes of significant bradycardia after shock delivery. Shocks of up to 20 J using the RA-LPA/CS lead configuration did not affect pacemaker function. IAD can result in transient bradycardia related to sinus and atrioventricular nodal effects requiring backup ventricular pacing. Shocks can be safely delivered using RA-LPA or RA-CS lead configurations in patients with implanted bipolar cardiac pacemakers.

Aged↗

Ursodeoxycholic acid in the treatment of liver diseases.

Ursodeoxycholic acid is a dihydroxy bile acid with a rapidly expanding spectrum of usage in acute and chronic liver diseases. The various mechanisms of action of this hydrophilic bile acid include direct cytoprotection, detergent action on dysfunctional microtubules, immunomodulation and induction of hypercholeresis. Its efficacy in primary biliary cirrhosis and primary sclerosing cholangitis as an adjunct to medical therapy has been well established. Newer indications include its use in the management of chronic hepatitis, cirrhosis, post liver transplant rejection, graft-versus-host disease and acute viral hepatitis, where it not only relieves symptoms of cholestasis but also arrests ongoing hepatocyte necrosis.

Biliary Tract Diseases↗

The role of pacemaker therapy in the prevention of atrial fibrillation.

Atrial fibrillation (AF) has been recognized, with increasing concern, as a potentially disabling illness, occurring either as a symptom of many cardiac diseases or as an isolated disorder. It can independently contribute to mortality and morbidity and may have serious prognostic importance in acute or chronic cardiac disease. In patients with symptomatic drug refractory atrial fibrillation, ventricular rate control by atrioventricular nodal ablation or modification commonly results in pacemaker implantation. The concept of AF prevention by pacemaker therapy has been introduced in patients with bradycardia-tachycardia syndrome or vagally mediated bradycardia-dependent AF. In patients with sick sinus syndrome, atrial pacing has proved to be more effective than VVI pacing in maintaining the electrical stability of the atrium in long-term follow up. Recently, the development of new techniques of atrial pacing employing pacing at two atrial sites may improve the effectiveness of the AF prevention by pacemaker therapy. Dual-site right atrial pacing using overdrive stimulation activates simultaneously the high right atrium and the left atrium via the ostium of the coronary sinus. Two main mechanisms have been proposed to explain the favourable effect of this technique. One is the suppression of atrial premature beats initiating AF by the overdrive pacing. The second is the alteration of atrial activation pattern by preexcitation of the area of the coronary sinus ostium which permits earlier recovery of excitability in sites of atrial conduction delay. The combination of drug therapy and pacing is essential for effective AF control. We have hitherto studied 30 patients with single- and dual-site pacing. Single-site pacing was performed at the high right atrium or coronary sinus ostium. The dual-site pacing mode increased the arrhythmia-free intervals, decreased patients' arrhythmia-related symptoms and anti-arrhythmic drug use as compared to the period preceding institution of pacing and incrementally over conventional high right atrial pacing alone. A multicentre randomized trial, Dual-site Atrial Pacing for Prevention of Atrial Fibrillation (DAPPAF), evaluating three pacing modes (dual-site, single-site and support pacing), is now in progress.

Atrial Fibrillation↗

Improved survival with an implanted defibrillator in patients with coronary disease at high risk for ventricular arrhythmia. Multicenter Automatic Defibrillator Implantation Trial Investigators.

BACKGROUND: Unsustained ventricular tachycardia in patients with previous myocardial infarction and left ventricular dysfunction is associated with a two-year mortality rate of about 30 percent. We studied whether prophylactic therapy with an implanted cardioverter-defibrillator, as compared with conventional medical therapy, would improve survival in this high-risk group of patients. METHODS: Over the course of five years, 196 patients in New York Heart Association functional class I, II, or III with prior myocardial infarction; a left ventricular ejection fraction < or = 0.35; a documented episode of asymptomatic unsustained ventricular tachycardia; and inducible, nonsuppressible ventricular tachyarrhythmia on electrophysiologic study were randomly assigned to receive an implanted defibrillator (n = 95) or conventional medical therapy (n=101). We used a two-sided sequential design with death from any cause as the end point. RESULTS: The base-line characteristics of the two treatment groups were similar. During an average follow-up of 27 months, there were 15 deaths in the defibrillator group (11 from cardiac causes) and 39 deaths in the conventional-therapy group (27 from cardiac causes) (hazard ratio for overall mortality, 0.46; 95 percent confidence interval, 0.26 to 0.82; P=0.009). There was no evidence that amiodarone, beta-blockers, or any other antiarrhythmic therapy had a significant influence on the observed hazard ratio. CONCLUSIONS: In patients with a prior myocardial infarction who are at high risk for ventricular tachyarrhythmia, prophylactic therapy with an implanted defibrillator leads to improved survival as compared with conventional medical therapy.

Adult↗

Prevention of recurrent atrial fibrillation with chronic dual-site right atrial pacing.

OBJECTIVES: We investigated 1) the feasibility, safety and efficacy of multisite right atrial pacing for prevention of atrial fibrillation (AF); and 2) the ability of atrial pacing in single- and dual-site modes to increase arrhythmia-free intervals in patients with drug-refractory AF. BACKGROUND: We recently developed and applied a novel technique of dual-site right atrial pacing in an unselected group of consecutive patients with AF requiring demand pacing. A prospective crossover study design was used to evaluate single- and dual-site right atrial pacing modes. METHODS: The frequency of AF during the 3 months before pacemaker implantation was analyzed. Consecutive consenting patients underwent insertion of two atrial leads and one ventricular lead with a DDDR pulse generator. Patients were placed in a dual-site pacing mode for the first 3 months and subsequently mode switched to single site pacing for 3 months. Mode switching was repeated at 6-month intervals thereafter. RESULTS: Atrial pacing resulted in a marked decline in AF recurrences (p < 0.001). During dual-site pacing with an optimal drug regimen, there was no AF recurrence in any patient compared with five recurrences in 12 patients during single-site pacing (p = 0.03). The mean (+/-SD) arrhythmia-free interval before pacing (14 +/- 14 days) was prolonged with dual- (89 +/- 7 days, p < 0.0001) and single-site pacing (76 +/- 27 days, p < 0.0001). Symptomatic AF episodes showed a declining trend during dual- and single-site pacing compared with those during the preimplantation period (p = 0.10). Mean antiarrhythmic drug use for all classes declined from 4 +/- 1.9 drugs before implantation to 1.5 +/- 0.5 (p < 0.01) drugs after implantation. Twelve (80%) of 15 patients remained in atrial paced rhythm at 13 +/- 3 months. CONCLUSIONS: We conclude that multisite right atrial pacing is feasible, effective and safe for long-term application. Atrial pacing significantly prolongs arrhythmia-free intervals in patients with drug-refractory paroxysmal AF. Dual-site right atrial pacing may offer additional benefits and should be considered either as the primary mode or in patients unresponsive to single-site pacing.

Adult↗

Picroliv protects against alcohol-induced chronic hepatotoxicity in rats.

Daily administration of ethyl alcohol (3.76 g/kg, p.o.) for 45 days resulted in significant changes in several biochemical parameters of the liver and serum of albino rats. After exposure to alcohol for 30 days when Picroliv (12 mg/kg, p.o.), an iridoid glycoside fraction of Picrorhiza kurroa, was administered for 15 days along with alcohol, the degree of change in most of the parameters was reduced.

Alcohol Dehydrogenase↗

New generations of implantable pacemaker defibrillators for ventricular and atrial tachyarrhythmias.

Implantable defibrillation devices have now been extensively applied to patients requiring cardioversion and defibrillation of sustained ventricular tachyarrhythmias. The focus of new developments is in improving technology, achieving physiologic operation in the atrium and ventricle, seeking new indications and identifying patient populations amenable to this therapy. Ventricular application technology is focusing on simplifying and improving robustness of lead systems yet seeking lower defibrillation thresholds. Dual chamber pacing, sensing and defibrillation are being developed. New populations for ventricular application include non-sustained ventricular tachycardia patients with coronary artery disease, dilated cardiomyopathy at risk for sudden death, long QT syndrome, pediatric patients with risk of sudden death and high risk postoperative coronary bypass patients. New applications include atrial defibrillation combined with atrial pacing in future devices. These devices are planned to have capabilities of ventricular defibrillation as a backup to address proarrhythmia concerns. It can be anticipated that implantable cardioverter defibrillator devices will be used for arrhythmia reversion in an expanding group of patients in the future.

Arrhythmias, Cardiac↗

Prognostic markers in amebic liver abscess: a prospective study.

OBJECTIVES: Amebic liver abscess (ALA) is being increasingly recognized with the progressive spread of AIDS. The prognosis of ALA needs to be determined to decide whether aggressive intervention therapy should be used. A prospective study was conducted to determine the factors that predicted mortality in patients with ALA. METHODS: The study population consisted of 135 consecutive patients with ALA who were treated with 80 mg/kg/day of metronidazole for 10 days if they survived. Needle aspiration or open surgical drainage was performed in patients who deteriorated despite drug therapy or had an abscess that clinically appeared to be at risk of impending rupture. Survivors and nonsurvivors were compared by univariate and multivariate analysis to identify predictors of outcome. These predictors were then prospectively evaluated in a subsequent cohort of patients with ALA. RESULTS: Twenty-four patients died during the acute phase. Significant differences between survivors and nonsurvivors were observed. A stepwise logistic regression suggested that a bilirubin level >3.5 mg/dl, encephalopathy, volume of abscess cavity, hypoalbuminemia (serum albumin level <2.0 g/dl), and the number of abscesses were independent risk factors for mortality. The duration of symptoms and type of treatment did not influence mortality. The regression equation derived was then applied prospectively to 64 subsequent patients with ALA, and the validity of the prediction rule was confirmed. The qualities of simplicity, availability, low cost of derivation, and good discriminating power suggest that this index would be useful in assessing prognosis in patients with ALA.

Adult↗

Clinical efficacy and safety of atrial defibrillation using biphasic shocks and current nonthoracotomy endocardial lead configurations.

We undertook a prospective randomized clinical trial evaluating efficacy and safety of internal atrial defibrillation in patients with drug-refractory atrial fibrillation (AF). Consecutive patients with paroxysmal or chronic AF were randomly tested with 3 internal atrial defibrillation lead configurations and biphasic shocks. Patients with implanted cardiac pacemakers were tested with the right atrium (RA) and left pulmonary artery or coronary sinus (CS) configuration. Shocks were initially delivered without anesthesia to assess patient tolerance. The need for backup ventricular defibrillation and pacing support was evaluated. Eighteen patients with (n = 15) or without (n = 3) structural heart disease, mean left ventricular ejection fraction 36 +/- 14%, and mean left atrial diameter 4.5 +/- 0.6 cm were studied. The mean defibrillation threshold in the best randomized lead configuration was 9.9 +/- 7.7 J. Mean defibrillation threshold for the right ventricle (RV) and superior vena cava configuration was 13.3 +/- 5 J, which was significantly lower than the RA and axilla configuration (20.1 +/- 7.4 J, p < 0.04) but not the RV to RA configuration (16.5 +/- 11 J, p > 0.2). The mean defibrillation threshold using the RA-left pulmonary artery/CS configuration was 8.9 +/- 9 J (p > 0.2 vs RV-superior vena cava). There was a bimodal distribution of defibrillation thresholds. Low atrial defibrillation thresholds correlated with absence of heart disease, higher ejection fraction, and smaller left ventricular end-diastolic diameter. Shocks were hemodynamically well tolerated, but 2 of 18 patients (11%) had nonsustained ventricular tachycardia after shock delivery. Six of 18 patients (33%) had postshock bradyarrhythmias. Fourteen of 16 patients perceived shocks > or = 3 J as intolerable.(ABSTRACT TRUNCATED AT 250 WORDS) [corrected]

Adult↗

Efficacy and safety of radiofrequency catheter ablation of left-sided accessory pathways through the coronary sinus.

Radiofrequency catheter ablation of left-sided accessory pathways (APs) with the use of an endocardial technique carries all potential risks of left heart catheterization. We analyzed the determinants of success, efficacy, and safety of radiofrequency catheter ablation from the coronary sinus (CS), as a potential alternative to the endocardial technique in these patients. Thirteen patients (mean age 40 +/- 20 years) with 15 left-sided APs and a history of symptomatic supraventricular tachycardia were included in the study. Nine APs were localized in the left posteroseptal region, and the remaining 6 in the left free wall. Ablation from CS was attempted in 12 patients with 14 APs. In 1 patient ablation within the CS was not deemed safe because of a small venous lumen. All 14 APs were successfully ablated using either CS ablation alone or combined with the endocardial technique. Efficacy of the CS ablation as a primary technique was 56% (5 of 9 APs). In 5 additional APs, ablation in the CS eliminated pathway conduction after failed endocardial attempts. CS ablation either as a primary or a secondary technique eliminated conduction in 10 of 14 APs (71.4%) (group 1). In the remaining 4 APs (group 2), the primary CS attempt was unsuccessful and APs were ablated with a subsequent endocardial approach. Determinants of success for the CS method were local AP to atrial and/or ventricular electrogram amplitude ratios > or = 1 (p < 0.05). The success rate of CS ablation was 83% in the left posteroseptal APs adjoining the branching point of the middle cardiac vein or a CS anomaly.(ABSTRACT TRUNCATED AT 250 WORDS)

Adolescent↗