Search PubMed⌕ Search

Biomedical subjects

Richard D Oleschuk

Publications and source records attributed to Richard D Oleschuk.

3 recordsLinked to original sources

Microchip-based capillary electrochromatography using packed beds.

Integration of a packed column onto a microchip for performance of capillary electrochromatography (CEC) is described. The quartz device incorporated a cross-injector, and a double weir trapping design for formation of 1, 2 and 5 mm long CEC columns. Three fluorescent dyes were baseline-resolved with plate numbers of 330,(330,000 plates/m; height equivalent to a theoretical plate, H = 3.0 microm) for BODIPY 493/503, 360 (360,000 plates/m; H = 2.8 microm) for rhodamine 123 and 244 (244,000 plates/m; H = 4.1 microm) for acridine orange (AO) with 500 V applied on a 1 mm long column. The 2 mm column yielded approximately 1.8 times more theoretical plates than did the 1 mm column, when operated at the same flow rate. Van Deemter plots were obtained for the three column lengths, showing increased plate height for the 5 mm length. A 2 mm column gave peak height and area relative standard deviation (RSD) values of 2.5 and 3.3%, respectively, as averages for the three dyes (n = 15). The RSD for the dye retention times was 1% (n = 6) over one day, and 3% (n = 30) over five days. Indirect fluorescence detection of thiourea and of amino acids was possible using a neutral indicator dye (BODIPY 493/503), with a detection limit of 10 microM for amino acids.

Arginine↗

Surface characterization using chemical force microscopy and the flow performance of modified polydimethylsiloxane for microfluidic device applications.

The widespread interest in micro total analysis systems has resulted in efforts to develop devices in cheaper polymer materials such as polydimethylsiloxane (PDMS) as an alternative to expensive glass and silicon devices. We describe the oxidation of the PDMS surface to form ionizable groups using a discharge from a Tesla coil and subsequent chemical modification to augment electroosmotic flow (EOF) within the microfluidic devices. The flow performance of oxidized, amine-modified and unmodified PDMS materials has been determined and directly compared to conventional glass devices. Exact PDMS replicas of glass substrates were prepared using a novel two step micromolding protocol. Chemical force microscopy has been utilized to monitor and measure the efficacy of surface modification yielding information about the acid/base properties of the modified and unmodified surfaces. Results with different substrate materials correlates well with expected flow modifications as a result of surface modification. Oxidized PDMS devices were found to support faster EOF (twice that of native PDMS) similar to glass while those derivatized with 3-aminopropyl triethoxysilane (APTES) showed slower flow rates compared to native PDMS substrates as a result of masking surface charge. Results demonstrate that the surface of PDMS microdevices can be manipulated to control EOF characteristics using a facile surface derivatization methodology allowing surfaces to be tailored for specific microfluidic applications and characterized with chemical force microscopy.

Dimethylpolysiloxanes↗

An integrated solid-phase extraction system for sub-picomolar detection.

A microchip structure etched on a glass substrate for packed column solid-phase extraction (SPE) and capillary electrochromatography (CEC) is described. A 200 microm long, octadecylsilane (ODS) packed column was secured using two different approaches: solvent lock or polymer entrapment. The former method was utilized for SPE while the latter approach was applied for CEC. In SPE, the ODS packed chamber gave a detection limit of 70 fM for a nonpolar BODIPY (493/503) dye when concentrated for 3 min at an electroosmotic flow rate of 4.14 nL/min, compared to 30 pM for this detector without the SPE step. SPE beds showed reproducible, linear calibration curves (R(2) = 0.9989) between 1 and 100 pM BODIPY at fixed preconcentration times. Breakthrough curves for the 330 pL (ODS-packed) bed indicated a capacity for BODIPY dye of 8.1 x 10(-14) mmol, or 0.25 mmol dye per liter of bed. The ODS-chamber could also be used to analyze dilute amino acid and peptide solutions. In the CEC format, two neutral dyes (BODIPY and acridine orange) were baseline-separated in an isocratic run with a theoretical plate count of 84 (420 000 plates/m) and a reduced plate height of about 1. A labeled peptide was also analyzed by CEC, using the acidic eluent (84% acetonitrile, and 26% aqueous trifluoroacetic acid (0.05%)) preferred for peptide separations on ODS-coated silica particles.

Acridine Orange↗