Immunoglobulin-complexed aspartate aminotransferase with a possible association with ulcerative colitis and its activity.

We demonstrate immunoglobulin-complexed aspartate aminotransferase (macro-AST) in a 14-year-old boy with rectitis-type ulcerative colitis by using both gel filtration and electrophoresis methods. The immunoglobulin complexed with AST in this case was identified as an IgG kappa both by electrosyneresis and immunoprecipitation reactions. The present case was noted to have a concomitant elevation of macro-AST associated with deterioration of ulcerative colitis. However, macro-AST has continued to exist when the activity of ulcerative colitis subsided, and the serum level of AST became normal. Thus, macro-AST might be related to an immunological component of ulcerative colitis as well as to the activity of the disease.

Adolescent

[Negative interference with peroxidase-coupled enzymatic serum creatinine assay by a hemostatic drug containing a hydroquinone structure].

We report a discrepancy between serum creatinine levels obtained by the method based on alkaline picric acid (Jaffé reaction) and that by automated analyzer employing enzymatic assay based on peroxidase-coupled reaction in a patient with renal dysfunction taking Ethamsylate, a capillary vessel stabilizer. Serum creatinine level obtained by the Jaffé reaction was 83 mg/l, and that by the enzymatic method was 22 mg/l. Recovery test of creatinine using the patient's serum showed 101 to 102% recovery by the Jaffé reaction and 77 to 89% by the enzymatic method. The same results were observed both in an in vitro test and in serum obtained from a healthy volunteer who had been given Ethamsylate previously. As the chemical structure of Ethamsylate includes a hydroquinone unit, it was thought that the compound may consume hydrogen peroxide produced by the reaction of peroxidase as the hydrogen donor and compete with the substrate used as the hydrogen donor for the color development. Thus evaluation of test results obtained by peroxidase coupled methods should be carefully interpreted in a patient who is consuming drugs which hydrogen peroxide such as hydroquinone.

Creatinine

[Decrease of CD16 antigen density on granulocytes in chronic myeloid leukemia].

We examined the fluorescence intensity of CD16 antigen, which represents the density of CD16 antigen, on granulocytes by flow cytometry in 15 healthy subjects and 15 patients with neutrophilia due to inflammatory diseases and 10 patients with chronic myeloid leukemia (CML). The fluorescence intensity of CD16 antigen was significantly lower in patients with CML than in healthy subjects and also than in patients with neutrophilia. These data indicate that 1) density of CD16 antigen on granulocytes decrease in CML, and 2) analysis on the density of granulocyte CD16 antigen is useful for differential diagnosis of CML from inflammatory diseases with neutrophilia.

Flow Cytometry

New enzymatic method with tryptophanase for determining potassium in serum.

We established a simple and rapid enzymatic method for measuring potassium ion in serum by using tryptophanase (EC 4.1.99.1) purified from Escherichia coli K12 strain (E. coli K12 IFO 3301). The presence of pyridoxal 5-phosphate promotes this enzymatic reaction, and potassium and (or) ammonium ions further accelerate it, with ammonium and potassium ions providing equivalent acceleration. We eliminated endogenous ammonium ion by using glutamate dehydrogenase (GLDH; EC 1.4.1.4), then produced ammonium ion in the presence of tryptophanase, tryptophan, and pyridoxal 5-phosphate. The concentration of formed ammonium ion, which was proportional to that of potassium ion in sample, was determined by adding GLDH to produce NADP+ in the presence of 2-oxoglutarate and NADPH; we then read the change of absorbance at 340 nm. The standard curve was linear for potassium ion concentrations up to 7.00 mmol/L. The within-assay variation (CV) was 0.89% at 5.51 mmol/L and 1.32% at 3.37 mmol/L. The day-to-day CVs were 0.99% at 6.85 mmol/L and 1.71% at 3.52 mmol/L. Analytical recoveries ranged from 98.7% to 108.9%. The correlation coefficient between values obtained with this enzymatic assay (y) and by flame photometry (x) was 0.995: y = 0.984x + 0.091 mmol/L (Sy.x = 0.105, n = 100). The presence of hemoglobin, bilirubin, or other cations little affects this system.

Ammonia

[Mixed type cryoglobulinemia with discrepancy of platelet counts measured with an automated cell counter and by manual counting method].

We experienced an abnormal blood sample which showed larger numbers of white blood cells and platelets counted by the electric resistance system and of platelets in mechanical counting by the flow system, than those in manual counting. Many precipitates were observed in this blood sample at room temperature or 4 degrees C, but were not observed at 37 degrees C. The precipitates were clarified to be caused by the mixed type (IgA, IgG and IgM) cryoglobulin by using an immunofluorescence method and an ouchterlony method. These data indicate that precipitation of cryoglobulin at room temperature induce falsely high value of platelet count in mechanical counting systems.

Chemical Precipitation

[Examination of the inhibitor to factor VIII in non-haemophilic patient].

We examined an inhibitor to factor VIII in non-haemophilic patient who had been developed widely spread ecchymosis and intramuscular bleeding. He had no previous personal or family history of abnormal bleeding tendency. His laboratory data was all normal except examination for blood coagulation. Coagulation studies showed prolonged activated partial thromboplastin time (48 sec) and decreased factor VIII activity (8%). The activity of the inhibitor to factor VIII was demonstrated to be 4.0 Bethesda unit. By the studies of dilution and time response curve, this inhibitor was found to inhibit up to 90% of factor VIII activity but not 100%. This inhibitor was shown to be IgG by protein-A affinity chromatography. In addition, bleeding time was prolonged in the patient. The value of von Willebrand factor antigen was 200%, but that of Ristocetin cofactor was 93%. Since the gel filtration analysis indicated that this inhibitor also suppressed Ristocetin cofactor activity, the relatively low value of Ristocetin cofactor might occur through the action of the inhibitor. These data suggest that patient's inhibitor react to factor VIII high molecular subunit.

Blood Coagulation Disorders

Digoxin-like immunoreactivity in Chinese medicine.

Immunoreactive digoxin-like activity was observed in Chinese medicine, KYUSHIN tablet, taken popularly in Japan without prescription. The antibodies used in the assays of digoxin reacted with Chan-su, the major effective component of KYUSHIN, which contained cardiotonic steroids with similar chemical structure as digoxin. One tablet of KYUSHIN had digoxin-like immunoreactivity equivalent to 1.9, 1.5, and 72 micrograms of digoxin measured by three different commercial kits. These different equivalences may be attributed to differences in cross-reactivity of the antibody used in the immunoassays. Two healthy volunteers took two KYUSHIN tablets three times a day, a typical dose, and digoxin-like immunoreactivity reached almost 0.4 micrograms/L in 0.5 day. Therapeutic drug monitoring should be interpreted carefully in patients taking Chinese medicines, many of which contain the Chan-su component.

Cross Reactions

Urinary iodide excretion measured with an iodide-selective ion electrode: studies on normal subjects of varying ages and patients with thyroid diseases.

As a part of studies concerning clinical application of the measurement of urinary iodide with an iodide-selective ion electrode, we report here the reference values for the iodide content or concentration in morning spot urine specimens from normal subjects of varying ages and studies with patients with thyroid diseases in Japan. The number distribution of the iodide content or concentration in the morning specimens appeared logarithmic normal in adults, children and infants. Normal ranges found in 95 per cent of populations of healthy subjects were 5.3 to 62.0 mumoles/g Creatinine (Cr) for adults, 5.3 to 42 mumoles/g Cr for children, and 1.9 to 56 microM for infants, respectively. Urinary iodide concentrations in breast-fed infants varied widely compared with those in bottle-fed infants. Mean values were 16.4 microM for breast-fed infants and 8.6 microM for bottle-fed infants, and they were not statistically different. Urine samples from the infants with transient hypothyroidism, who had undergone amniofetography showed extraordinarily high iodide concentrations, even though they were measured at 20th, 29th and 30th days after birth. Although urinary iodide excretion in patients with simple goiter was within normal limits, the mean was statistically lower than that in normal controls (p less than 0.001). Because of the simplicity and rapidity of the electrode method, we strongly recommend it for use in examining iodide excretion in patients with various thyroid diseases.

Adolescent