Adaptive sensitivity encoding incorporating temporal filtering (TSENSE).

A number of different methods have been demonstrated which increase the speed of MR acquisition by decreasing the number of sequential phase encodes. The UNFOLD technique is based on time interleaving of k-space lines in sequential images and exploits the property that the outer portion of the field-of-view is relatively static. The differences in spatial sensitivity of multiple receiver coils may be exploited using SENSE or SMASH techniques to eliminate the aliased component that results from undersampling k-space. In this article, an adaptive method of sensitivity encoding is presented which incorporates both spatial and temporal filtering. Temporal filtering and spatial encoding may be combined by acquiring phase encodes in an interleaved manner. In this way the aliased components are alternating phase. The SENSE formulation is not altered by the phase of the alias artifact; however, for imperfect estimates of coil sensitivities the residual artifact will have alternating phase using this approach. This is the essence of combining temporal filtering (UNFOLD) with spatial sensitivity encoding (SENSE). Any residual artifact will be temporally frequency-shifted to the band edge and thus may be further suppressed by temporal low-pass filtering. By combining both temporal and spatial filtering a high degree of alias artifact rejection may be achieved with less stringent requirements on accuracy of coil sensitivity estimates and temporal low-pass filter selectivity than would be required using each method individually. Experimental results that demonstrate the adaptive spatiotemporal filtering method (adaptive TSENSE) with acceleration factor R = 2, for real-time nonbreath-held cardiac MR imaging during exercise induced stress are presented.

Algorithms↗

Ghost artifact cancellation using phased array processing.

In this article, a method for phased array combining is formulated which may be used to cancel ghosts caused by a variety of distortion mechanisms, including space variant distortions such as local flow or off-resonance. This method is based on a constrained optimization, which optimizes SNR subject to the constraint of nulling ghost artifacts at known locations. The resultant technique is similar to the method known as sensitivity encoding (SENSE) used for accelerated imaging; however, in this formulation it is applied to full field-of-view (FOV) images. The method is applied to multishot EPI with noninterleaved phase encode acquisition. A number of benefits, as compared to the conventional interleaved approach, are reduced distortion due to off-resonance, in-plane flow, and EPI delay misalignment, as well as eliminating the need for echo-shifting. Experimental results demonstrate the cancellation for both phantom as well as cardiac imaging examples.

Artifacts↗

Low-latency temporal filter design for real-time MRI using UNFOLD.

To improve real-time control of interventional procedures such as guidance of catheters, monitoring of ablation therapy, or control of dosage during drug delivery, the image acquisition and reconstruction must be high speed and have low latency (small time delay) in processing. A number of different methods have been demonstrated which increase the speed of MR acquisition by decreasing the number of sequential phase-encodes. A design and implementation of the UNFOLD method which achieves the desired low latency with a recursive temporal filter is presented. The recursive filter design is characterized for this application and compared with more commonly used moving average filters. Experimental results demonstrate low-latency UNFOLD for two applications: 1) high-speed, real-time imaging of the heart to be used in conjunction with cardiac interventional procedures; and 2) the injection of drugs into muscle tissue with contrast enhancement, i.e., monitoring needle insertion and injection of a drug with contrast enhancement properties. Proof-of-concept was demonstrated by injecting a contrast agent. In both applications the UNFOLD technique was used to double the frame rate.

Animals↗