Two-year outcomes after sirolimus-eluting stent implantation: results from the Sirolimus-Eluting Stent in de Novo Native Coronary Lesions (SIRIUS) trial.

OBJECTIVES: The purpose of this study was to examine the two-year clinical outcomes in patients enrolled in the Sirolimus-Eluting Stent in De Novo Native Coronary Lesions (SIRIUS) study. BACKGROUND: The SIRIUS study was a double-blinded randomized study which demonstrated that sirolimus-eluting stents (SES) significantly improved angiographic results (at 8 months) and clinical outcomes (at 9 and 12 months) compared with bare-metal stents (BMS). METHODS: Patients with de novo native coronary artery lesions randomized to either SES (533 patients) or control BMS (525 patients) were followed for two years. RESULTS: Between one and two years, there were infrequent additional clinical events that were equally distributed between the sirolimus and control groups. After two years, target lesion revascularization was 5.8% and 21.3% in SES and control patients, respectively (p < 0.001), and major adverse cardiovascular events and target vessel failure rates were 10.1% versus 24.4% and 12.0% versus 26.7%, respectively (p < 0.0001 for both). There were no differences in death, myocardial infarction, and stent thrombosis between the two groups. CONCLUSIONS: Clinical outcomes two years after implantation of SES continue to demonstrate significant reduction in the need for repeat target lesion (and vessel) revascularization compared with BMS without evidence for either disproportionate late restenosis or late stent thrombosis.

Coronary Angiography↗

Safety and feasibility of transendocardial autologous bone marrow cell transplantation in patients with advanced heart disease.

The present report contains the final results of a Phase I study that evaluated the feasibility, safety, and potential efficacy of intramyocardial injection of autologous bone marrow (BM) in "no-option" patients with refractory angina and myocardial ischemia. Twenty-seven patients underwent electromechanic mapping-guided transendomyocardial injections (n = 12, 0.2 ml each) of unfractionated autologous BM cells directed to ischemic, noninfarcted myocardial territory. Patients were injected with 28 +/- 27 x 10(6)/ml nucleated cells containing 2.2 +/- 1.4% CD34+ cells. The autologous BM injection procedure was successful in all patients and was associated with no adverse events. At 3 months, the Canadian Cardiovascular Society angina score (3.2 +/- 0.5 vs 2.0 +/- 0.91, p = 0.001) and treadmill exercise duration (418 +/- 136 vs 489 +/- 142 seconds, p = 0.017) had improved significantly. The stress-induced ischemia score within the injected territories (118 segments) had also improved (2.2 +/- 0.8 vs 1.7 +/- 1.1, p < 0.001). At 1 year, the clinical improvement was sustained, although 5 patients had undergone revascularization procedures. The number of total injected nucleated cells (CD45+), progenitor cells (CD34+), and the magnitude of secreted vascular endothelial growth factor and macrophage chemoattractant protein-1 by cultured BM cells failed to predict the clinical response. In conclusion, the 3- and 12-month study results have indicated the safety of catheter-based transendocardial delivery of autologous BM cells in patients with advanced symptomatic ischemic heart disease and may suggest sustained potential efficacy. The cellular and humeral characteristics of autologous BM cells did not predict the clinical response, underscoring the advisability of additional mechanistic exploration.

Angina, Unstable↗

Serial angiographic follow-up of sirolimus-eluting stents for unprotected left main coronary artery revascularization.

OBJECTIVES: This study was performed to evaluate the clinical and serial angiographic outcomes of patients undergoing sirolimus-eluting stent (SES) implantation for unprotected left main coronary artery (LMCA) stenosis. BACKGROUND: The efficacy of SES has led to their expanded use for off-label indications, including LMCA disease. METHODS: Unprotected LMCA intervention with SES was attempted in 50 patients. Surveillance angiography was performed at three and nine months' follow-up. RESULTS: The target lesion involved the distal LMCA in 47 patients (94%). In-lesion restenosis occurred in 21 patients (42%), was focal in 85% of cases, and in 82% involved the branch ostia, sparing the LMCA itself. Target lesion revascularization (TLR) occurred in 19 patients (38%) over a mean follow-up of 276 +/- 57 days; TLR was ischemia-driven in 7 patients (14%). Late loss was significantly greater within the left circumflex (LCX) ostium compared to the parent vessel (PV) of the LMCA bifurcation (0.83 +/- 0.89 mm vs. 0.49 +/- 0.72 mm, p = 0.04). Late loss continued to increase between three- and nine-month follow-up. Final minimal luminal diameter and maximal balloon pressure were independent predictors of restenosis of the PV. CONCLUSIONS: Restenosis is a frequent finding when serial angiographic follow-up is performed after SES implantation for unprotected distal LMCA lesions. Restenosis is usually focal, most often involves the LCX ostium, and often occurs without symptoms.

Aged↗

Intravascular ultrasound assessment of fibrous cap remnants after coronary plaque rupture.

BACKGROUND: Although intravascular ultrasound (IVUS) can detect plaque rupture, the fibrous cap remnant has not previously been studied in detail. The aim of the present study is to assess the fibrous cap remnants by IVUS in ruptured plaques. METHODS: In 53 patients, a ruptured plaque with a fibrous cap remnant was studied by IVUS. RESULTS: In 36 (68%) patients, the rupture of the fibrous cap appeared to have occurred at the shoulder. The absolute length of the fibrous cap remnant was significantly longer in the center rupture site compared with the shoulder rupture site (1.37 +/- 0.56 vs 0.84 +/- 0.34 mm, P = .001); however, the estimated length of the original fibrous cap did not differ between the 2 rupture site groups (2.28 +/- 0.66 vs 2.11 +/- 0.69, P = not significant). In none of the patients did the remnants of the fibrous cap cover the entire mouth of the cavity. The estimated absolute length of the missing part of the fibrous cap correlated significantly with the cavity area (r = 0.517, P < .001), the lesion external elastic membrane area (r = 0.330, P = .016), the lumen area (r = 0.289, P = .036), the maximum plaque thickness (r = 0.364, P = .007), and the length of the estimated original fibrous cap (r = 0.709, P < .001). CONCLUSION: In general, the postrupture fibrous cap does not cover the entire mouth of the ruptured plaque cavity in its postrupture state. Potential explanations include the following: (1) part of the fibrous cap may be too thin to be visualized with IVUS, (2) part of it may have embolized, or (3) the prerupture fibrous cap may have been stretched and/or there were postrupture changes in lesion geometry.

Aged↗

Transcatheter aortic valve replacement in patients with critical aortic stenosis: rationale, device descriptions, early clinical experiences, and perspectives.

The development of lesser invasive transcatheter techniques for aortic valve replacement (AVR) to treat high surgical risk patients with severe aortic stenosis (AS) has engendered controversy among traditional cardiovascular therapists. Presently, there are two catheter-based treatment systems (the Cribier-Edwards Aortic Bioprosthesis and the CoreValve Revalving System) utilizing either a balloon-expandable or a self-expanding stent (or cage) platform which unfolds a pericardial tissue valve within the displaced diseased aortic valve. After ex vivo durability testing and animal studies, several clinical registries with these transcatheter AVR systems in almost 300 patients worldwide have demonstrated the following: (1) good acute hemodynamic performance with reduction in mean aortic valve gradients to <10 mm Hg; (2) frequent para-valvular regurgitation, which has improved with self-expanding devices and the use of larger (26 mm) valve sizes; (3) acceptable periprocedural (30-day) mortality (<10%) with the newest generation devices and improved operator techniques. Enlightened interdisciplinary treatment teams incorporating surgeons, interventionalists, and medical therapists as well as rigorously conducted randomized clinical trials will be required to determine if these innovative transcatheter AVR approaches will represent a viable therapy for high-risk patients with severe AS in the future.

Aortic Valve Insufficiency↗

Overview of pharmacology and clinical trials program with the zotarolimus-eluting endeavor stent.

Despite considerable benefits associated with current drug-eluting stents (DES), continued attention to the safety, efficacy, and deliverability of first-generation DES has led to the development of new antiproliferative agents with alternative stent platforms and different drug carrier systems. Zotarolimus is a recently developed pharmacologic agent with both antiproliferative and anti-inflammatory properties. The Endeavor drug-eluting stent (Medtronic Vascular, Santa Rosa, CA) represents the combination of zotarolimus, a low-profile cobalt alloy stent platform, and a biocompatible phosphorylcholine drug carrier system. At present, four clinical trials examining the safety and efficacy of the Endeavor stent have been performed. Although these studies have enrolled patients with similar clinical and angiographic characteristics, they have differed in trial design and study population size and have been performed across a broad geographic and physician distribution. Despite these differences, the results of these trials demonstrate consistently low rates of angiographic restenosis and repeat revascularization in addition to a favorable safety profile, with no occurrences of late stent thrombosis through 1 year of follow-up. This review describes the pharmacology and design on the Endeavor stent, summarizes results from recent clinical trials evaluating the Endeavor stent, and provides an overview of ongoing and future directions for clinical investigation.

Anti-Inflammatory Agents↗

A multicenter, randomized, dose-finding study of gamma intracoronary radiation therapy to inhibit recurrent restenosis after stenting.

OBJECTIVES: The objective of this double-blind, randomized study was to determine the safety and efficacy of intracoronary radiation therapy (ICRT) with a dose of 17 Gray (Gy) compared to the currently recommended dose prescription of 14 Gy for the treatment of in-stent restenosis within bare metal stents. BACKGROUND: While gamma ICRT for in-stent restenosis has been proven efficacious, the optimal dose is unknown, and radiation failure due to recurrent neointimal hyperplasia remains a significant clinical problem for some patients. A higher radiation dose may improve outcomes, but may potentially increase adverse events. METHODS: Following coronary intervention, 336 patients with in-stent restenosis were randomly assigned to receive ICRT with either 14 Gy or 17 Gy at 2 mm from an 192-iridium source. RESULTS: At 8-month follow up, fewer patients in the 17 Gy group underwent target lesion revascularization (TLR = 15.2% versus 27.2%; p = 0.01), target vessel revascularization (21.3% versus 33.1%; p = 0.02), or reached the composite endpoint of death, myocardial infarction, thrombosis, or TLR (17.1% versus 28.4%; p = 0.02). There were no differences in late thrombosis or mortality between treatment groups. There was a strong trend toward reduced in-lesion late loss (0.36 +/- 0.63 mm vs. 0.51 +/- 0.64 mm; p = 0.09) and a significantly lower rate of binary restenosis (23.9% versus 38.1%; p = 0.031) in the high dose group. CONCLUSIONS: Gamma ICRT with 17 Gy is safe and, compared to 14 Gy, reduces recurrent stenosis and clinical events at 8-month follow up. An increase in the currently recommended gamma radiation dose prescription from 14 Gy to 17 Gy should be strongly considered.

Aged↗

Frequency and determinants of black holes in sirolimus-eluting stent restenosis.

BACKGROUND: In-stent neointimal hyperplasia typically has a homogenous echoreflective appearance during intravascular ultrasound (IVUS) imaging. However, the neointimal hyperplasia of some brachytherapy failure lesions contains a very echolucent, dark zone that has been termed black hole. We sought to investigate the frequency and determinants of black holes (BH) after sirolimus-eluting stent (SES) implantation. METHODS: A prospective, consecutive series of 33 intrastent SES restenosis (> 50% neointimal hyperplasia with a minimal lumen area < 4.0 mm2) was studied with IVUS. Patients were divided into 2 groups: 8 patients with BH versus 25 patients without BH. RESULTS: Fifty percent of the BH cases occurred in saphenous vein graft lesions. BH tissue was noticed in 8% of all patients with in-stent restenosis. Three patients in each group had previous vascular brachytherapy failure (p = 0.1). Compared to non-BH cases, a greater proportion of BH cases occurred after SES treatment of bare-metal stent restenosis (75% vs. 32%; p = 0.035). BH cases presented earlier (89.9 +/- 34.3 vs. 161.3 +/- 78.8 days; p = 0.001) with more severe in-stent restenosis, as indicated by greater absolute and relative amounts of neointimal hyperplasia (41.3 +/- 10.0 vs. 26.9 +/- 19.5 mm3; p = 0.012, and 19.6 +/- 6.8 vs. 10.4 +/- 7.9%; p = 0.0001, respectively). CONCLUSIONS: Echolucent restenotic tissue (black hole) is more often observed in SES restenosis after treatment of saphenous vein graft lesions or treatment of bare-metal stent restenosis. It occurs earlier and is more severe than typical SES failures.

Aged↗

Intravascular ultrasound characterization of the "black hole" phenomenon after drug-eluting stent implantation.

An intraluminal echolucent tissue, dubbed "black hole," has been identified by intravascular ultrasonography after intracoronary brachytherapy. This study reports the characteristics and incidence of the black hole in patients treated with drug-eluting stent implantation using a sirolimus-eluting stent (SES). We included intravascular ultrasound data from the Compassionate Use of Sirolimus-Eluting Stent (SECURE, n = 61 lesions) registry, a study involving patients in whom previous brachytherapy had failed, and the DIABETES trial (n = 165 lesions), a multicenter, randomized study comparing SES versus bare metal stents in diabetic patients. Intravascular ultrasound follow-up was scheduled at 8 months (SECURE trial, post-brachytherapy population) and 9 months (DIABETES trial). In the SECURE population, a black hole was observed in 10 patients (19.6%). Seven black hole segments had significant intimal hyperplasia (> 10%). A black hole accounted for 27% of total intraluminal tissue. In the DIABETES trial, 2 patients (2.5%) in the SES group and none in the bare metal stent group showed echolucent intimal hyperplasia. In conclusion, a black hole occurred frequently after implantation of a SES in patients in whom intracoronary brachytherapy had previously failed. Black holes were also identified in a nonirradiated population, although the incidence was lower than in the post-brachytherapy patients. Bare metal stents were not associated with this phenomenon.

Adult↗

A blinded, randomized, placebo-controlled trial of percutaneous laser myocardial revascularization to improve angina symptoms in patients with severe coronary disease.

OBJECTIVES: This study was a randomized, patient- and evaluator-blinded, placebo-controlled trial in patients treated using percutaneous myocardial laser revascularization. BACKGROUND: Previous studies using similar therapies have been confounded by placebo bias. METHODS: A total of 298 patients with severe angina were randomly assigned to receive low-dose or high-dose myocardial laser channels or no laser channels, blinded as a sham procedure. The primary end point was the change in exercise duration from baseline examination to six months. RESULTS: The incidence of 30-day death, stroke, myocardial infarction, coronary revascularization, or left ventricular perforation occurred in two patients in the placebo, eight patients in the low-dose, and four patients in the high-dose groups (p = 0.12); 30-day myocardial infarction incidence was higher in patients receiving either low-dose or high-dose laser (nine patients) compared with placebo (no patients, p = 0.03). At six months, there were no differences in the change in exercise duration between those receiving a sham (28.0 s, n = 100), low-dose laser (33.2 s, n = 98), or high-dose laser (28.0 s, n = 98, p = 0.94) procedure. There were also no differences in the proportion of patients improving to better than Canadian Cardiovascular Society class III angina symptoms at six months. The follow-up visual summed stress single-photon-emission computed tomography scores were not significantly different from baseline in any group and were no different between groups. The modest improvement in angina symptoms assessed by the Seattle Angina Questionnaire also was not statistically different among the arms. CONCLUSIONS: Treatment with percutaneous myocardial laser revascularization provides no benefit beyond that of a similar sham procedure in patients blinded to their treatment status.

Angina Pectoris↗

Association of plaque characterization by intravascular ultrasound virtual histology and arterial remodeling.

Positive remodeling is more often observed in lesions of patients who have acute coronary syndromes or vulnerable (rupture-prone) plaques. However, there are few data that correlate plaque morphology, composition, and arterial remodeling in vivo. We evaluated coronary plaque characterization of lesions with positive remodeling using intravascular ultrasound (IVUS) radiofrequency data analysis. Seventy-seven nonbifurcation native coronary lesions (in 50 patients) were imaged in vivo using 30-MHz IVUS transducers. Lesions were classified into 4 plaque types, fibrous, fibrofatty, dense calcium, and necrotic core, by using processing of the radiofrequency signal validated in vitro. The remodeling index was calculated as the lesion external elastic membrane area divided by the proximal reference external elastic membrane area. Lesions were divided into 2 groups: positive remodeling (remodeling index>1.0, 26 lesions) and intermediate/negative remodeling (remodeling index<or=1.0, 51 lesions). Total plaque volume and the volume of each plaque type were averaged over the length of the lesion. Reference segment plaque compositions were similar. Mean lesion fibrofatty plaque area was significantly larger in lesions with positive remodeling than in lesions with intermediate/negative remodeling (1.2+/-0.7 vs 0.8+/-0.4 mm2, p=0.001; 26.3+/-6.6% vs 19.8+/-5.7%, p<0.001, of total plaque volume). The same results were obtained at the minimum lumen site and in the subgroup of patients who had acute coronary syndromes. Further, there was a linear relation between remodeling index and fibrofatty plaque area (r=0.26, p=0.02). In conclusion, in vivo IVUS radiofrequency data analysis demonstrates that positive remodeling occurs in lesions with more fibrofatty plaque.

Coronary Artery Disease↗

Impact of asymmetric stent expansion on neointimal hyperplasia following sirolimus-eluting stent implantation.

To assess whether asymmetric stent expansion affects suppression of neointimal hyperplasia after sirolimus-eluting stent implantation, 64 patients in the SIRolImUS-coated Bx Velocity stent trial who underwent single 18-mm stent implantation and 3-dimensional intravascular ultrasonography at 8-month follow-up were enrolled. To assess the longitudinal stent asymmetric expansion, 2 cross sections with a maximal/minimal stent area were chosen in each patient. To assess for tomographic stent asymmetric expansion, stent eccentricity was determined by dividing the minimum stent diameter by the maximum stent diameter. At the 2 cross sections with a maximal/minimal stent area, a sirolimus-eluting stent reduced neointimal hyperplasia significantly with no interaction between the treatment and stent areas. A sirolimus-eluting stent also significantly reduced neointimal hyperplasia in the concentric and eccentric stent groups.

Angioplasty, Balloon, Coronary↗

Late incomplete stent apposition after sirolimus-eluting stent implantation: a serial intravascular ultrasound analysis.

OBJECTIVES: We sought to identify the frequency of incomplete stent apposition (ISA) in sirolimus-eluting stents (SES) and clarify its findings and clinical sequelae. BACKGROUND: Late-acquired ISA has been reported in bare-metal stents (BMS) and brachytherapy and recently in drug-eluting stents. However, the characteristics of late ISA in SES have not been clarified. METHODS: From the SIRIUS trial, a randomized, multicenter study comparing SES and BMS, serial qualitative intravascular ultrasound (IVUS; at stent implantation and eight-month follow-up) was available in 141 patients (BMS: n = 61; SES: n = 80). The IVUS images were reviewed for the presence of ISA. RESULTS: Incomplete stent apposition at follow-up was observed in 19 patients (BMS: n = 6 [9.8%]; SES: n = 13 [16.3%]; p = NS). Among these, 12 had ISA after intervention and at follow-up (persistent ISA). Late-acquired ISA was seen in the remaining seven cases, all from the SES group (BMS: n = 0; SES: n = 7 [8.7%]; p < 0.05). In late-acquired ISA, there was an increase in external elastic membrane area (after intervention: 16.2 +/- 2.7 m2; follow-up: 18.9 +/- 3.6 mm2; p < 0.05). The location of stent-vessel wall separation was primarily at the stent edges in persistent ISA cases, whereas late-acquired ISA in SES occurred mostly in the mid portion of the stent. There were no negative clinical events reported for any ISA cases at 12-month clinical follow-up. CONCLUSIONS: Late ISA was observed in 8.7% of patients after SES implantation. There were no negative clinical events associated with this IVUS finding at 12-month clinical follow-up; however, careful long-term follow-up will be necessary.

Combined Modality Therapy↗

Impact of coronary culprit lesion calcium in patients undergoing paclitaxel-eluting stent implantation (a TAXUS-IV sub study).

Randomized clinical trials have shown that paclitaxel-eluting stents significantly reduce restenosis after percutaneous coronary intervention. The impact of lesion calcification on the efficacy of paclitaxel-eluting stents is unknown. In the TAXUS-IV trial, 1,314 patients who underwent percutaneous coronary intervention were randomly assigned to a bare-metal or paclitaxel-eluting stent. By core laboratory analysis, 247 lesions (19%) were moderately or severely calcified. At the 9-month angiographic follow-up examination, the paclitaxel-eluting stent had significantly reduced the amount of late loss compared with the control stent (0.26 +/- 0.56 vs 0.51 +/- 0.48 mm, p = 0.015) within the analysis segment in the calcific lesions. The analysis segment restenosis rate was similar in patients with calcified and noncalcified lesions after paclitaxel-eluting stent implantation (7.5% vs 8.0%, respectively; p = 1.0). The rate of ischemia-driven target lesion revascularization (TLR) at 1 year was reduced by 56% in patients with calcified lesions (11.9% vs 5.1%, p = 0.09) and by 75% in noncalcified lesions (15.7% vs 4.3%, p <0.0001). By interaction testing, the efficacy of the paclitaxel-eluting stent in reducing TLR at 1 year was similar in the calcified and noncalcified lesions (p = 0.30). Moreover, by multivariate analysis, implantation of the paclitaxel-eluting stent was a powerful independent predictor of freedom from TLR, with similar hazard ratios for efficacy in calcified and noncalcified lesions (0.30 and 0.26, respectively). In conclusion, implantation of paclitaxel-eluting stents in patients with de novo coronary lesions significantly reduced restenosis in patients with and without calcified lesions.

Angioplasty, Balloon, Coronary↗

Predictors of edge stenosis following sirolimus-eluting stent deployment (a quantitative intravascular ultrasound analysis from the SIRIUS trial).

To study the interaction of the sirolimus-eluting stent and vessel margins, we analyzed the intravascular ultrasound parameters in 317 edges of 167 stents having 18 edge stenoses at 8 months of follow-up from the SIRIUS trial. Of the baseline parameters, a larger reference percentage of plaque area and a larger edge stent area/reference minimum lumen area were associated with edge stenosis in the sirolimus-eluting stent cohort compared with the incidence of edge stenosis in the bare metal stent cohort. Thus, full lesion coverage and matching the stented segment properly to the adjacent segment using intravascular ultrasound guidance may improve sirolimus-eluting stent implantation efficacy further.

Blood Vessel Prosthesis Implantation↗

Impact of intravascular ultrasound lesion characteristics on neointimal hyperplasia following sirolimus-eluting stent implantation.

The effect of lesion characteristics on neointimal hyperplasia after sirolimus-eluting stent implantation was examined in 45 patients who underwent successful preinterventional intravascular ultrasound. There were no differences in neointimal hyperplasia between the moderate/severe calcified lesion group (calcium arc >120 degrees ) and the non/mild calcified lesion group or between the positive vessel remodeling group (external elastic membrane area at the minimal lumen area site larger than that at the proximal reference site) and negative vessel remodeling group. No correlation between preinterventional plaque burden and neointimal hyperplasia was found. In patients who have coronary artery disease, sirolimus-eluting stents continue to demonstrate striking suppression of neointimal proliferation, irrespective of lesion characteristics previously associated with greater restenotic risk.

Blood Vessel Prosthesis Implantation↗