Search PubMed⌕ Search

Biomedical subjects

M Dolovich

Publications and source records attributed to M Dolovich.

At least 19 recordsLinked to original sources

Anti-inflammatory effects of once daily low dose inhaled ciclesonide in mild to moderate asthmatic patients.

BACKGROUND: Ciclesonide exhibits clinical efficacy at 160 microg (ex-actuator) once daily but the anti-inflammatory effects at this dose are not known. We wished to know whether 4 weeks therapy with ciclesonide pMDI 160 microg once daily in the morning exhibited significant anti-inflammatory effects. METHODS: Seventeen patients with mild persistent asthma (FEV(1) 3.35 l) were recruited into a double-blind placebo-controlled randomized crossover study. Measurements were made after ciclesonide and placebo treatment as well as after run-in and washout periods, for adenosine monophosphate (AMP) bronchial challenge (primary variable), exhaled nitric oxide (NO) and induced sputum (in a subgroup). RESULTS: The mean (SEM) AMP bronchial challenge PC(20) following ciclesonide (140 (63) mg/ml) was significantly (P < 0.001) increased compared with placebo (17 (8) mg/ml), run-in (13 (5) mg/ml) and washout (9 (3) mg/ml) periods. This amounted to an eightfold (CI: 5.3-12.0) for ciclesonide vs placebo. Likewise, there were significant improvements in exhaled NO levels and a significant reduction in induced sputum eosinophil cell counts. CONCLUSION: We have shown that inhaled ciclesonide given at 160 microg once daily in the morning exhibits significant anti-inflammatory effects that are in keeping with the previously described clinical effects.

Administration, Inhalation↗

Distribution of intranasal instillations in mice: effects of volume, time, body position, and anesthesia.

Intranasal instillation techniques are used to deliver various substances to the upper and lower respiratory tract (URT and LRT) in mice. Here, we quantify the relative distribution achieved with intranasal delivery of a nonabsorbable tracer, (99m)Tc-labeled sulfide-colloid. Relative distribution was determined by killing mice after instillation and quantifying the radioactivity in dissected tissues using gamma scintigraphy. A significant effect of delivery volume on relative distribution was observed when animals were killed 5 min after instillation delivered under gas anesthesia. With a delivery volume of 5 microl, no radiation was detected in the LRT; this increased to a maximum of 55.7 +/- 2.5% distribution to the LRT when 50 microl were delivered. The majority of radiation not detected in the LRT was found in the URT. Over the course of the following 1 h, radiation in the LRT remained constant, while that in the URT decreased and appeared in the gastrointestinal tract. Instillation of 25 microl into anesthetized mice resulted in 30.1 +/- 6.9% distribution to the LRT, while only 5.3 +/- 1.5% (P < 0.05) of the same volume was detected in the LRT of awake mice. Varying the body position of mice did not affect relative distribution. When using intranasal instillation, the relative distribution between the URT and LRT and the gastrointestinal tract is heavily influenced by delivery volume and level of anesthesia.

Administration, Intranasal↗

Delivery of HFA and CFC salbutamol from spacer devices used in infancy.

The aim of the study was to compare the in vitro delivery of four salbutamol pressurized metered-dose inhalers (pMDIs) via the three spacer devices commonly used in European infants: Aerochamber-Infant, Babyhaler, and metallic NES-spacer. Emitted dose (ED) and fine particle dose (FPD, particles<5.8 microm) of each combination of spacer device and pMDI (chlorofluorocarbon-based Ventoline, Eolène, Spréor, and hydrofluoroalkane-based Airomir were measured respectively using unit dose sampling tubes (n=30 per combination) and an 8-stage cascade impactor (n=6 per group). The results were compared by analysis of variance and the Student-Newman-Keuls method. ED of Airomir was always greater than for Ventoline (P<0.05). FPD obtained with Ventoline was the lowest, with Eolène>Airomir=Spréor>Ventoline (P<0.05). Only Airomir produced a similar FPD with all three spacer devices. Chlorofluorocarbon-salbutamol pMDIs are not generics when used with spacer devices. The three spacer devices may be used interchangeably with Airomir.

Albuterol↗

Therapeutic equivalence of three metered-dose inhalers containing salbutamol (Albuterol) in protecting against methacholine-induced bronchoconstriction in children with asthma.

Many pharmaceutical companies sell salbutamol in metered-dose inhalers (MDI) for the treatment of asthma. However, the therapeutic equivalence of the more recently released generic products has not been compared with the original patented product in children. Twenty children with mild to moderate asthma, presently asymptomatic and with normal lung function, were randomly allocated to receive 200 microg of inhaled salbutamol (Albuterol) from three MDIs prepared by different manufacturers: the original Glaxo product and two generic products. The three drug formulations and placebo were given 10 min before a methacholine challenge test to determine the degree of protection provided against methacholine-induced bronchoconstriction (MIB) by each salbutamol aerosol. Tests were performed on 4 consecutive days. Doubling concentrations of methacholine were inhaled until the forced expired volume in 1 sec (FEV(1)) decreased by 20% from its baseline value. Compared to placebo, all patients increased significantly the provocation concentration that decreased FEV(1) by 20% (PC(20)) by more than one doubling concentration after inhaling each of the three salbutamol aerosols. The effectiveness was not significantly different between medications (P = 0.8). There was a small but significant difference among MDIs in aerosol particle size and total and fine-particle dose released per actuation. However, no relation was found between aerosol particle size or released dose and the protective effect. This study shows that the three tested brands of salbutamol MDI protected asthmatic children equally from MIB. When prescribing these salbutamol MDIs to prevent symptoms triggered by nonspecific stimuli in asthmatic children, the selection may be based on cost-benefit criteria.

Adolescent↗

Sputum induction: effect of nebulizer output and inhalation time on cell counts and fluid-phase measures.

BACKGROUND: A knowledge of the factors that can affect induced sputum results is essential in order to standardize the procedure. OBJECTIVE: We investigated the influence of nebulizer output on sputum cell counts and fluid phase measurements at increasing times of sputum induction. METHODS: Eighteen adults with stable asthma inhaled an aerosol of 3% hypertonic saline to induce sputum after 7, 14 and 21 min on 2 days separated by 48 h. On one day, in random order, the ultrasonic nebulizer used had a relatively low output of 0.87 mL/min (particle size 5.58 microm mass median aerodynamic diameter, MMAD) and, on the other, a higher output of 1.90 mL/min (particle size 4.14 microm MMAD). The sputum was selected from each expectorate and examined blind to the induction procedures. RESULTS: With both nebulizers, the 14- and 21-min samples were lower in weight, neutrophils, eosinophils, eosinophil cationic protein (ECP) and interleukin (IL)-8 and higher in macrophages. The higher output nebulizer induced sputum with higher cell viability and lower ECP and IL-8. CONCLUSION: The results identify that the volume of hypertonic saline inhaled in sputum induction influences the fluid-phase measurements. The duration of induction does alter the cell counts and suggests that the later expectorated sputum samples originate from more peripheral airways. The results draw attention to the need to standardize the volume and time of nebulization to accurately interpret and compare results.

Administration, Inhalation↗

Delivery of salbutamol pressurized metered-dose inhaler administered via small-volume spacer devices in intubated, spontaneously breathing rabbits.

Little is known about the ability of small-volume valved spacer devices to deliver a significant amount of an aerosolized drug to the lungs of babies. This study compared the in vitro delivery of salbutamol administered via Aerochamber-Infant (145 mL), Babyhaler (350 mL), and metallic NES-spacer (250 mL), as well as the in vivo delivery using an animal model. The lung deposition study of technetium-99m-labeled salbutamol was conducted in six anesthetized, intubated (3.0-mm endotracheal tube simulating oropharyngeal deposition), spontaneously breathing New Zealand White rabbits, a model for 3-kg babies. Each rabbit was studied on three separate occasions, once with each spacer device. The amount of radioactivity deposited in the spacer device, the endotracheal tube, the lungs, or the body was measured by a gamma camera and expressed as a percentage of the emitted labeled dose. The emitted dose and particle size distribution of salbutamol via the three spacer devices were measured using unit dose sampling tubes and an eight-stage Anderson cascade impactor, respectively. The results were compared by ANOVA or Student-Newman-Keuls test when indicated. In vitro, the NES-spacer and Babyhaler were equivalent for delivering particles <5.8 microm in diameter (NES-spacer = Babyhaler > Aerochamber-Infant; p < 0.05). In vivo, the lung and body deposition was low with all spacer devices (range: 0.52-5.40% of the delivered dose) but greater with the NES-spacer than with the Aerochamber-Infant or the Babyhaler (5.40 +/- 2.40%, 2.91 +/- 0.86%, 0.52 +/- 0.46%, respectively; p = 0.002). These results suggest the metal-valved spacer device may be preferable for delivering pressurized aerosols to spontaneously breathing infants.

Administration, Inhalation↗

Safety and efficiency of metered dose inhaler delivery of salbutamol in the intubated rabbit.

OBJECTIVE: To determine the safety and efficiency of metered dose inhaler salbutamol delivered to the intubated rabbit. DESIGN: Prospective, comparative, five-group laboratory investigation. SETTING: Animal laboratory, Department of Nuclear Medicine. SUBJECTS: A total of 30 adult, anesthetized New Zealand White rabbits. INTERVENTIONS: Three groups of rabbits underwent tracheal intubation through a tracheostomy and received 5 puffs of 99mTcO4 salbutamol delivered at the elbow connector (group 1) or via a catheter with its distal tip positioned at the midpoint (group 2) or bevel of the endotracheal tube (group 3). No intervention was provided for the rabbits in the fourth group. A fifth group underwent tracheal intubation through the mouth and received salbutamol (5 puffs) delivered at the bevel of the endotracheal tube. MEASUREMENTS: Delivery efficiency was expressed as the ratio of radioactivity emitted from lungs and trachea to the total radioactivity of the administered dose. Histopathologic injury scores were assigned to each trachea or lung specimen. MAIN RESULTS: Delivery efficiency was 30 times greater in groups 3 and 5 (full catheter) than in group 1 (elbow). The injury scores were similar in all groups. CONCLUSION: We conclude that the increased efficiency obtained by administration of metered dose inhaler salbutamol at the distal tip of endotracheal tube is not necessarily associated with increased epithelial injury.

Adrenergic beta-Agonists↗

Estimation of pulmonary deposition of aerosol using gamma scintigraphy.

Following delivery of technetium 99m-labeled aerosols through a ventilator circuit, the amount of radioactivity in the lungs of 58 ventilated rabbits was estimated first by gamma scintigraphy via gamma camera and later by direct counting of the excised lungs (n = 116 specimens) with a gamma counter. The in situ radioactivity measured via scintigraphy was closely correlated with the gamma counter ex vivo tissue counts of the radioactivity (R2 = 0.997, P < 0.001). Overall, gamma scintigraphy gave slightly lower values of activity than the tissue counts from the gamma counter, but the limits of agreement between the two measurements were narrow enough for us to consider that the tissue and scintigraphy methods were in agreement. We conclude that gamma scintigraphy provides a convenient and noninvasive means for the accurate estimation of aerosol deposition in the lungs of small animals and possibly in small infants.

Aerosols↗

New propellant-free technologies under investigation.

Interest in the use of solution inhalers and dry powder inhalers (DPIs) has increased in recent years, broadening from systems for the delivery of traditional respiratory medications to the delivery of inhaled proteins, peptides, and analgesics. This has led to numerous developments and a range of inhaler designs and technology, each with advantages and distinctive features. Novel devices for aerosolization of solutions can be broadly divided into three main classes: adaptive aerosol delivery devices, metered dose liquid inhalers (MDLIs), and breath-actuated nebulizers. Similarly, the wide range of DPIs in development can be classified generally according to their means of storing and providing the drug, that is, as single capsules, a bulk reservoir, or multiunit-dose devices. Choice of a system for delivering a liquid solution is based on a number of factors (e.g., ease of use, convenience, aerosol properties, dose delivery, and drug deposition), the combination of which will have an impact on the single most important factor: clinical benefit. Ultimately, the patient must be able to use the device easily, maintain it, and derive clinical benefit from the drug delivered by the system. Physicians and patients must recognize that if one system does not work, an alternative must be tried.

Administration, Inhalation↗

Randomised controlled study of early use of inhaled corticosteroid in preterm infants with respiratory distress syndrome.

AIM: To investigate the therapeutic efficacy of inhaled fluticasone propionate, started on day 1 of age, on ventilated preterm infants with respiratory distress syndrome. METHODS: Starting within 24 hours of age, ventilated preterm infants (gestation < 32 weeks, birthweight < 1.5 kg) with respiratory distress syndrome were given a 14 day course (two puffs, 12 hourly) of either fluticasone propionate (250 microg/puff) (group 1, n=27) or placebo (group 2, n=26) with a metered dose inhaler-spacer device. Response to treatment was assessed by the rate of successful extubation by days 7 and 14 of age, changes in respiratory system mechanics, death, occurrence of chronic lung disease, and other neonatal complications. RESULTS: More infants in the treatment group were successfully extubated by 14 days of age than those in the placebo group (17/27 vs 8/26; p = 0.038). The treated infants also showed a more significant improvement in respiratory system compliance during the first 14 days of life. The two groups, however, did not differ significantly in their need for systemic steroids after day 14 of age, death, or the occurrence of chronic lung disease. The treatment was not associated with any increase in neonatal complications, including those attributable to steroid induced side effects. CONCLUSION: These results provide preliminary evidence that early treatment with inhaled corticosteroids may be beneficial to ventilated preterm infants with respiratory distress. Further study of its use in a large scale randomised trial is warranted.

Administration, Inhalation↗

A randomized controlled trial on the effect of inhaled corticosteroids on airways inflammation in adult cigarette smokers.

STUDY OBJECTIVE: To determine whether inhaled corticosteroid treatment can reduce airways inflammation in adult cigarette smokers. DESIGN: This was a randomized, placebo-controlled, double-blinded clinical trial. SETTING: The subjects were recruited from the community by advertising. PARTICIPANTS: Seventy-one adults with a > or = 5 pack-year history who were current smokers, had a normal FEV1, and produced sputum daily. INTERVENTION: Sixty subjects were randomized to receive four puffs of placebo or beclomethasone dipropionate ([BDP]; total dosage, 1,000 microg/d) using a metered-dose aerosol inhaler with a valved holding chamber (AeroChamber; Trudell Medical; London, Ontario, Canada) for 28 days. MEASUREMENTS AND RESULTS: Eleven subjects were not randomized because of poor compliance. The primary outcome was fractional airway neutrophilia, as assessed by a differential cell count of sputum. Additional outcome measures were spirometry, measurement of airway responsiveness by methacholine challenge, and lung epithelial permeability measured by the clearance of radiolabeled diethylenetriamine pentaacetic acid. There were no significant differences between the two groups in any outcome measurement after 4 weeks of treatment. CONCLUSIONS: With normal spirometry, we found no benefit of treatment with inhaled BDP, 1,000 microg/d, on noninvasive measures of airways inflammation in adult smokers. This indicates that cigarette smoke-induced inflammation in its early stages (before a demonstrable airflow obstruction) is not steroid sensitive. This may occur because the site of involvement is not accessible to inhaled medications or because the inflammatory process is resistant to moderate doses of inhaled corticosteroids.

Administration, Inhalation↗

Aerosol delivery to children: what to use, how to choose.

Careful in vitro assessment of aerosol delivery devices using realistic breathing patterns can be advantageous in choosing an efficient and efficacious delivery system for infants and children. These modeling investigations become more important in assessing delivery to the very young patient, more so if the treatment is not as successful as anticipated. The choices in devices to use for inhalant therapy are increasing and physicians and patients should have a good understanding of the functional differences between them.

Aerosols↗

Effect of continuous lateral rotational therapy on lung mucus transport in mechanically ventilated patients.

PURPOSE: Continuous lateral rotational therapy (CLRT) <40 degrees is a method of altering the position of the ventilated patient to help clear secretions from the lung. CLRT has not been shown to reduce the incidence of atelectasis or pneumonia but potentially offers a way to maximize positional drainage in these patients without producing adverse effects. Treatment intervention, bracketed by two (nonrotational) control periods. The purpose of this study was to determine if CLRT alters mucus transport in critically ill, intubated patients in the intensive care unit of a teaching hospital. MATERIALS AND METHODS: Thirteen critically ill, but stable, mechanically ventilated patients, mean age 74 years, were enrolled. They were placed supine on a Biodyne bed (KCI, San Antonio, Texas) and pressures in the cushions adjusted to patient's weight. A radiolabeled aerosol was delivered by bagging for 2 to 3 minutes and repeated measurements of lung radioactivity were obtained by imaging of the thorax over the following 3 hours. A 90-minute period of rotation of the bed, 30 degrees to either side was preceded and followed by two 45-minute control periods during which the patient remained supine and stationary on the bed. Coughs and suctions were recorded and blood gases obtained pre and post study. RESULTS: (1) The mucous clearance was slower than that reported in normal subjects and in ambulatory patients with COPD; (2) there was a slight, but not significant, increase in clearance during CLRT; (3) clearance reverted to pre-oscillation levels following therapy. Lack of significant effect may be attributed to too shallow an angle for rotation or too short an intervention period. CONCLUSION: Positional drainage effected by short duration CLRT did not appear to stimulate significant mucous removal from the lung in critically ill patients but also did not cause any adverse effects.

Administration, Inhalation↗

Nebulisation of surfactants in an animal model of neonatal respiratory distress.

AIMS: To evaluate pulmonary deposition and gas exchange following nebulisation of two surfactants by either a jet or an ultrasonic nebuliser. METHOD: After bronchoalveolar lavage (BAL), 19 rabbits were ventilated in four groups. Group A1 (n = 5) and A2 (n = 6) received Technetium-99m labelled Exosurf, and groups B1 (n = 4) and B2 (n = 4) received radiolabelled Survanta. Groups A1 and B1 received jet nebuliser therapy, whereas groups A2 and B2 received ultrasonic nebuliser. Pulmonary deposition, distribution, and blood gases were determined. RESULTS: Pulmonary deposition as per cent of initial dose and mg lipid) was 0.28(0.10)% or 0.59(0.21) mg in group A1, 1.05(0.23)% or 2.21(0.48) mg in group A2, 0.08(0.02)% or 0.30(0.08) mg in group B1, and 0.09(0.02)% or 0.34(0.08) mg in group B2. Deposition in group A2 was greater than in other groups (p = 0.001). Group A2 showed a small improvement in blood gases. CONCLUSIONS: Even the highest deposition--ultrasonic nebuliser with Exosurf--achieved limited clinical effect. The aerosol route is currently not effective for surfactant treatment.

Aerosols↗

Delivery of metered dose inhaler aerosols to paralyzed and nonparalyzed rabbits.

OBJECTIVE: To assess whether paralysis alters pulmonary deposition of albuterol delivered by metered dose inhaler and spacer to small animals. DESIGN: A parallel group study of intubated and ventilated rabbits. INTERVENTIONS: Animals in group 1 (n = 7) were paralyzed with intravenous pancuronium, and ventilated at a rate of 30 breaths/ min. The animals in group 2 (n = 6) were ventilated at a rate of 10 breaths/min under light anesthesia without paralysis. In this latter group, spontaneous respiration continued at a rate of 40 to 50 breaths/min. Both groups were maintained at PaCO2 of 35 to 40 torr (4.7 to 5.3 kPa), and other ventilatory settings were identical. MEASUREMENTS AND MAIN RESULTS: Technetium-99m labeled albuterol aerosol was delivered by metered dose inhaler via a spacer device to both groups. Pulmonary deposition of the aerosol, determined by measuring the radioactivity in the lung tissues at autopsy, was expressed as percent of the total radioactivity dispensed by the metered dose inhaler. Group 2 showed significantly greater lung deposition than group 1 (0.510 +/- 0.076 [SEM]% vs. 0.226 +/- 0.054%, p = .0094). Deposition in the airway, the endotracheal tube, and the ventilator circuit did not differ significantly. CONCLUSION: Metered dose inhaler delivery of aerosolized medications to ventilated rabbits is significantly enhanced if respiration is not controlled. This observation might have implications for the delivery of therapeutic aerosols to newborns and young infants receiving slow, intermittent, mandatory ventilation.

Aerosols↗