Tissue phantom ratios (TPR), based on a normalization depth of 5 cm, have been measured in water for field sizes from 5 x 5 cm2 to approximately 40 x 40 cm2 and for depths from 1 to 40 cm for a Varian Clinac 4/100. These TPR's have been compared with those calculated from percent depth doses measured at the same time, and the two sets of data generally agree to better than 1%, with an average ratio of measured to calculated TPR of 0.999 +/- 0.013. Beam profiles have been measured for open and wedged fields, with particular concern for the often observed "horns," or the increase in dose at the corners of the field. The maximum dose at a depth of 1 cm, along the diagonal of the field for this machine, is approximately 5% higher than at the same depth on the central axis, whereas along the principal plane the maximum dose is only about 3% higher.

Models, Structural↗

A thermoregulated enclosure for controlling thermal drift in a radiation calorimeter.

The ability to control thermal drifts is essential in operating a calorimeter. We investigated a thermal enclosure, which envelops the calorimeter with temperature-regulated air, thus thermally isolating the calorimeter from the room. The desired temperature in the enclosure is controlled by a control circuit and a thermoelectric device, which works as a Peltier effect heat pump. In this report, the details of the enclosure design and construction are presented with actual performance evaluations.