Design and evaluation of a flow phantom.
RATIONALE AND OBJECTIVES: We constructed a near-anatomically correct large-vessel phantom to perform repeatable flow dynamics research examinations by angiography, magnetic resonance (MR) angiography, and computed tomography (CT) angiography. METHODS: An internal carotid artery was constructed within a head phantom. The internal carotid artery branches into a middle and an anterior cerebral artery; the former trifurcates and ends in the superior sagittal sinus, and the latter ends in the inferior sagittal sinus. A transverse and sigmoid sinus drains the model. All four vessels connecting the arterial and venous vessels have variable flow-constricting ligatures placed around them. These ligatures are accessible on the skull surface. The skull cavity is filled with a silicone polymer that is isodense to brain on CT scans and isointense on most MR images. RESULTS: The flow in the phantom's vessels may be varied in a repeatable manner. Multiple scan sequences may be performed without the image degradation caused by patient motion. The homogeneity of the filler polymer allows visualization of flow-related artifacts that may be hidden by complex human anatomy. CONCLUSION: Preliminary images of each modality show promise for use of the phantom in imaging research on large-vessel flow dynamics.