Search PubMed⌕ Search

Biomedical subjects

Julián Alonso

Publications and source records attributed to Julián Alonso.

3 recordsLinked to original sources

Microflow injection system based on a multicommutation technique for nitrite determination in wastewaters.

In this work a microflow structure, suitable for micro-FIA (micro flow injection analysis), will be described, evaluated and applied to real samples. Microchannels, the detector flow cell and input/output ports have been micromachined in silicon and sealed with anodically bonded Pyrex glass. The channels are defined by etching approximately 200 microm depth in silicon using a dry reactive ion etching (RIE) process. Optical windows integrated in the chip structure allow simple absorbance/transmission measurements to be made. The optical measurements were made using an LED as emitter (lambda=525 nm) and a photodiode as a detector. A Visual-Basic program has been developed to control an automatic burette, three-way solenoid valves and the data acquisition system. The micro-FIA for nitrite determination using the Griess-Ilosvay reaction has been implemented for the on-line monitoring of wastewater treatment plants (WWTPs). The multicommutation concept has been applied in order to enhance the mixing process inside the microsystem. Tandem streams of reagent and sample were generated and evaluated at different commutation frequencies. Two optimal frequencies, 400/200 ms and 150/450 ms, were found to be the most suitable ones. The first commutation ratio gave rise to wide linear working range (0-250 ppm), in spite of a high detection limit (0.35 ppm) and a low sensitivity (0.0041+/-0.0004 AU ppm-1). With the second ratio, the working linear range was smaller (0-50 ppm) but the detection limit (0.17 ppm) and the sensitivity (0.0091+/-0.0003 AU ppm-1) improved remarkably. Finally, real samples with a high nitrite concentration (0-1500 ppm) coming from a study of kinetic inhibition in the nitrification process at a WWTP has been analysed with the proposed micro-FIA system. The obtained results have allowed the corroboration of the model of inhibition by the nitrite ion with great exactitude.

Autoanalysis↗

Flow injection potentiometric system for the simultaneous determination of inositol phosphates and phosphate: phosphorus nutritional evaluation on seeds and grains.

A simple flow injection potentiometric (FIP) system, which uses a tubular cobalt electrode, has been developed for phosphorus nutritional evaluation of seeds and grains. Inorganic phosphorus, P(i), is determined using a 1 x 10(-2) mol.L(-1) potassium phthalate buffer solution adjusted at pH 4. A sensitivity of 47 mV/decade and an operating range from 10 to 1000 mg.L(-1) (1 x 10(-4)-1 x 10(-2) M) of dihydrogen phosphate are obtained. The inositol phosphates amount, which is referred to the organic phosphorus, P(org), is directly determined from extracts using a 1 x 10(-2) mol.L(-1) Tris-HCl buffer solution adjusted at pH 8. A sensitivity of 127 mV/decade and an operating range of 10-1000 mg.L(-1) (2.5 x 10(-4)-5 x 10(-3) M) of P(org) (expressed as inositol hexakisphosphoric acid monocalcium) are achieved. Some samples of seed and grain are analyzed by an ICP-OES and a spectrophotometric method to compare results to the developed flow system; no significant differences at the 95% confidence level are observed using a paired t test. Other samples such as animal nursing feed, soybean meal, and corn are also analyzed with the proposed FIP system, showing a good correlation to the ICP-OES values.

Edible Grain↗

Improved integrated waveguide absorbance optodes for ion-selective sensing.

The first prototype of a technologically improved integrated waveguide absorbance optode (IWAO) was developed and tested with a membrane based on a new H+-selective ketocyanine dye and a cadmium ionophore. It was designed with curved instead of rectilinear planar waveguides. Results demonstrated the suitability of the new IWAOs to be employed as sensing platforms, which confer versatility, robustness, and mass production capabilities besides high sensitivity on conventional bulk optodes, as well as the usefulness of such dyes in developing ion-selective membranes in combination with a selective ionophore. The sensor integration as a detector in a flow injection system (FIA) was proposed to obtain an automated, simple, and sufficiently reproducible (RSD <5%) analytical methodology with a sample throughput of 55 h(-1). Very sensitive optodes were obtained, and detection limits on the order of 20 ppb were achieved. Because of the ionophore employed, the optode system showed excellent selectivity over alkali and alkaline-earth metals with the exception of samples containing lead and cadmium ions, where the membrane responded to both analytes. The proposed procedure combines all the advantages of the FIA systems, the simplicity of optical detection, ion recognition selectivity, and sensitivity of ketocyanine dyes, and the features achieved using the integrated device, which comprise an improved sensitivity and short response times as well as robustness, easy handling, and mass production.

Journal Article↗