Search PubMed⌕ Search

Biomedical subjects

J P Hoggatt

Publications and source records attributed to J P Hoggatt.

6 recordsLinked to original sources

Clinicopathologic study of the effect of radial tears and loop fixation on intraocular lens decentration.

PURPOSE: The purpose of this study is to determine the effect of loop fixation and anterior capsular tears on intraocular lens (IOL) decentration. METHODS: A retrospective measurement of IOL decentration was performed on 144 human eyes with posterior chamber (PC) IOLs obtained after death. RESULTS: Decentration in eyes with asymmetrical bag-sulcus fixation (mean +/- standard deviation, 0.64 +/- 0.39 mm) was significantly higher than eyes with symmetrical fixation. In the presence of radial tears, symmetrically fixated IOLs in either the capsular bag or the ciliary sulcus decentered to a similar degree, 0.35 +/- 0.25 mm and 0.4 +/- 0.26 mm, respectively. The least decentration was observed with capsular fixation and no radial tears (0.18 +/- 0.09 mm). This was significantly less decentration than with any other form of fixation in the presence of radial tears. CONCLUSION: This study shows that capsular fixation with no radial tears, as can be achieved by using the continuous curvilinear capsulorhexis, is associated with the least decentration.

Adult↗

Posterior continuous curvilinear capsulorhexis. An experimental study with clinical applications.

The posterior continuous curvilinear capsulorhexis technique has been advocated in cases of posterior capsule rupture during extracapsular cataract extraction. The authors compared posterior continuous curvilinear capsulorhexis with posterior capsular sharp-edged tears. Two different types of forces were experimentally created on the posterior capsule of 30 human eyes obtained after death: (1) implantation and dialing of posterior chamber intraocular lenses (PC IOLs) and (2) increased intravitreal pressure by injection of balanced salt solution. All posterior capsular tears extended toward the equator, causing major capsular defects. In contrast, the posterior continuous curvilinear capsulorhexis remained intact in all cases. This experimental study proves that in cases where an inadvertent posterior capsular tear occurs, a posterior continuous curvilinear capsulorhexis is useful in preventing further capsular damage. Also, in cases where a posterior capsulotomy is indicated, a smooth edge created by a posterior continuous curvilinear capsulorhexis may be useful to maintain the integrity of the capsular bag for PC IOL capsular implantation.

Aged↗

Prospective experimental study of factors related to posterior chamber intraocular lens decentration.

The effect of posterior chamber intraocular lens (IOL) dimensions, design, style, loop fixation, and anterior capsular tears on decentration were investigated in an experimental model. Nine posterior chamber IOLs of various designs and styles with loop diameters between 12.0 and 14.0 mm and optic diameters between 5.0 and 7.0 mm were implanted in human eyes obtained post mortem. Symmetrical and asymmetrical fixation were investigated in eyes with and without radial tears using the Miyake posterior view technique. Location of IOL loops proved to be the most significant factor in IOL decentration. Decentration was least with symmetrical bag/bag fixation and no radial tears (mean = 0.20 +/- 0.05 mm). Asymmetrical bag/sulcus fixation in the presence of anterior capsular tears was associated with the highest decentration rate (mean 0.68 +/- 0.28 mm). Optic size and total loop diameter had no apparent effect on IOL centration in the immediate postoperative period.

Cataract Extraction↗

Scanning electron microscopic study of modern silicone intraocular lenses.

Silicone, as manufactured today, appears to be a biocompatible material. The safety and efficacy of silicone lenses are primarily related to the intraocular lens (IOL) design. We compared the edge finish of two three-piece polypropylene loop foldable silicone IOL designs (Allergan Medical Optics) and three one-piece designs (Staar Surgical Co., CooperVision-Cilco). Except for an early Staar one-piece design, all lenses including the more recent Staar lenses had acceptably smooth edges with minimal molding flash.

Biocompatible Materials↗

Posterior capsule opacification.

A complication of extracapsular cataract extraction with or without posterior chamber intraocular lens (PC-IOL) implantation is posterior capsule opacification. This condition is usually secondary to a proliferation and migration of residual lens epithelial cells. Opacification may be reduced by atraumatic surgery and thorough cortical clean-up. Clinical, pathological and experimental studies have shown that use of hydrodissection, the continuous curvilinear capsulorhexis and specific IOL designs may help reduce the incidence of this complication. Capsular-fixated, one-piece all-polymethylmethacrylate PC-IOLs with a C-shaped loop configuration and a posterior convexity of the optic are effective. Polymethylmethacrylate loops that retain "memory" create a symmetric, radial stretch on the posterior capsule after in-the-bag placement, leading to a more complete contact between the posterior surface of the IOL optic and the taut capsule. This may help form a barrier against central migration of epithelial cells into the visual axis. Various pharmacological and immunological methods are being investigated but conclusive data on these modalities are not yet available.

Cataract↗