Search PubMed⌕ Search

Biomedical subjects

J Guttmann

Publications and source records attributed to J Guttmann.

At least 19 recordsLinked to original sources

Alveolar recruitment in acute lung injury.

Alveolar recruitment is one of the primary goals of respiratory care for acute lung injury. It is aimed at improving pulmonary gas exchange and, even more important, at protecting the lungs from ventilator-induced trauma. This review addresses the concept of alveolar recruitment for lung protection in acute lung injury. It provides reasons for why atelectasis and atelectrauma should be avoided; it analyses current and future approaches on how to achieve and preserve alveolar recruitment; and it discusses the possibilities of detecting alveolar recruitment and derecruitment. The latter is of particular clinical relevance because interventions aimed at lung recruitment are often undertaken without simultaneous verification of their effectiveness.

Humans↗

Reduced activation of immunomodulatory transcription factors during positive end-expiratory pressure adjustment based on volume-dependent compliance in isolated perfused rabbit lungs.

BACKGROUND: Repeated alveolar collapse and cyclic alveolar overdistension with associated activation of inflammatory signalling cascades contribute to ventilator-induced lung injury (VILI). The appropriate positive end-expiratory pressure (PEEP) which prevents or ameliorates VILI is unknown. In the isolated perfused lung, repeated adjustments of PEEP based on the continuously analysed intratidal compliance-volume curve have previously been shown to result in full end-expiratory alveolar recruitment and low risk of cyclic alveolar overdistension. Accordingly, we tested the hypothesis that such ventilatory management reduces intrapulmonary activation of the immunomodulatory transcription factors nuclear factor kappaB (NF-kappaB), activator protein 1 (AP-1) and cAMP-responsive element binding protein (CREB) which induce the expression of various chemokines and cytokines. METHODS: Isolated perfused rabbit lungs were randomly allocated to one of three groups: zero end-expiratory pressure (ZEEP) to induce repeated alveolar collapse (n=6), high PEEP to induce cyclic alveolar overdistension (n=6) and repeated PEEP adjustments based on intratidal compliance-volume curve analysis by the slice method to minimize repeated alveolar collapse and overdistension (n=9). All lungs were ventilated with a tidal volume of 6 ml kg(-1) bodyweight for 120 min. Thereafter, activation of transcription factors NF-kappaB, AP-1 and CREB in lung tissue was analysed by electrophoretic mobility shift assay. RESULTS: High PEEP was associated with the highest activation of NF-kappaB and AP-1 and repeated PEEP adjustments with the lowest activation when compared with the other two study groups (P<0.001). In contrast, activation of CREB did not differ between groups. Activated NF-kappaB and AP-1 protein complexes consisted mainly of the transactivators p50/p65 and c-Fos/Jun, respectively. CONCLUSIONS: In isolated perfused rabbit lungs, repeated adjustments of PEEP based on the continuously analysed intratidal compliance-volume curve were associated with less activation of early steps of inflammatory signalling cascades than ventilation with ZEEP or high PEEP.

Animals↗

Good short-term agreement between measured and calculated tracheal pressure.

BACKGROUND: Tracheal pressure (P(tr)) is required to measure the resistance of the tracheal tube and the breathing circuit. P(tr) can either be measured with a catheter or, alternatively, calculated from the pressure-flow data available from the ventilator. METHODS: Calculated P(tr) was compared with measured P(tr) during controlled ventilation and assisted spontaneous breathing in 18 healthy and surfactant-depleted piglets. Their lungs were ventilated using different flow patterns, tidal volumes (V(T)) and levels of positive end-expiratory pressure. RESULTS: In terms of the root mean square error (RMS), indicating the average deviation of calculated from measured P(tr), the difference between calculated and measured P(tr) was 0.6 cm H(2)O (95%CI 0.58-0.65) for volume-controlled ventilation; 0.73 cm H(2)O (0.72-0.75) for pressure support ventilation; and 0.78 cm H(2)O (0.75-0.80) for bi-level positive airway pressure ventilation. CONCLUSION: The good agreement between calculated and measured P(tr) during varying conditions, suggests that calculating P(tr) could help setting the ventilator and choosing the appropriate level of support.

Air Pressure↗

Extubation after breathing trials with automatic tube compensation, T-tube, or pressure support ventilation.

BACKGROUND: Automatic tube compensation (ATC) is a new option to compensate for the pressure drop across the endotracheal or tracheostomy tube (ETT), especially during ventilator-assisted spontaneous breathing. While several benefits of this mode have so far been documented, ATC has not yet been used to predict whether the ETT could be safely removed at the end of weaning, from mechanical ventilation. METHODS: We undertook a systematic trial using a randomized block design. During a 2-year period, all eligible patients of a medical intensive care unit were treated with ATC, conventional pressure support ventilation (PSV, 5 cmH2O), or T-tube for 2-h. Tolerance of the breathing trial served as a basis for the decision to remove the endotracheal tube. Extubation failure was considered if reintubation was necessary or if the patient required non-invasive ventilatory assistance (both within 48 h). RESULTS AND CONCLUSIONS: After the inclusion of 90 patients (30 per group) we did not observe significant differences between the modes. Twelve patients failed the initial weaning trial. However, half of the patients who appeared to fail the spontaneous breathing trial on the T-tube, PSV, or both, were successfully extubated after a succeeding trial with ATC. Extubation was thus withheld from four and three of these patients while breathing with PSV or the T-tube, respectively, but to any patient breathing with ATC. It seems that ATC can be used as an alternative mode during the final phase of weaning from mechanical ventilation. Furthermore, this study may promote a larger multicenter trial on weaning with ATC compared with standard modes.

Device Removal↗

Automatic tube compensation (ATC).

Automatic tube compensation (ATC) is a new option to compensate for the non-linearly flow-dependent pressure drop across an endotracheal or tracheostomy tube (ETT) during inspiration and expiration. ATC is based on a closed-loop working principle. ATC is not a true ventilatory mode but rather a new option which can be combined with all conventional ventilatory modes. ATC compensates for the tube-related additional work of breathing. As of yet, ATC has been associated with certain benefits for the tracheally intubated spontaneously breathing patient. Among these, reduced work of breathing, preservation of the natural "noisy" breathing pattern, enhanced synchronization between the patient and the ventilator, and improvement of respiratory comfort seem to be most important. Moreover, sufficient spontaneous breathing with ATC alone, i.e. without any additional ventilatory assist, might help to predict more accurately readiness for extubation in the last phase of weaning from mechanical ventilation. Furthermore, it has been shown in patients with acute lung injury that ATC unloaded the inspiratory muscles and increased alveolar ventilation without adversely affecting cardiorespiratory function. It is the purpose of this article to describe the working principle of ATC and to give a review of the actual scientific discussion concerning ATC.

Bronchi↗

Analysis of forced expired volume signals using multi-exponential functions.

Patients with pulmonary disease are often unable to complete forced expiration manoeuvres. The aim of the study is to evaluate whether forced vital capacity (FVC), the volume exhaled at the end of completed forced expiration, can be estimated by extrapolating volume-time curves obtained from uncompleted manoeuvres. The suitability of mono-, bi-, and tri-exponential functions to characterise complete volume-time curves from 50 subjects is investigated. Mono-exponential modelling is insufficient, whereas bi-exponential fitting yields an adequate description for 47 data sets. Tri-exponential models lead to overfitting in all but three cases (normalised sum of least squares: 50.2 +/- 34.5 for mono-, 2.76 +/- 4.11 for bi-, 2.74 +/- 4.19 for tri-exponential modelling; condition number of the correlation matrix: 1.0025 +/- 0.0004 for mono-, 1.08 +/- 0.08 for bi-, 34.7 +/- 100.1 for tri-exponential fitting (mean +/- SD)). Thus, FVC is estimated by the extrapolation of 27 uncompleted spirograms using bi- or tri-exponential models, depending on their accordance with measured data and on the identifiability of their parameters. This algorithm yields unbiased estimates (difference from measured inspiratory vital capacity: 0.01 +/- 0.21 L). This method can be used for investigation of the lung function of subjects who cannot complete the forced expiration manoeuvre.

Adolescent↗

[Automatic tube compensation (ATC)].

The endotracheal tube (ETT) is a considerably flow-dependent and, therefore, variable mechanical load. Conventional modes of respiratory support cannot adequately compensate for the tube resistance in inspiratorion and not at all in expiration. Automatic tube compensation (ATC) compensates for the flow-dependent pressure drop across the tracheal tube by a positive pressure support in inspiration and by a negative pressure support in expiration. The pressure support closely follows the nonlinear pressure-flow curve of the ETT. Automatic tube compensation has an indirect closed-loop working principle since the target tracheal pressure is not directly measured but rather calculated from continuously measured airway pressure and flow rate. It is not an own ventilatory mode but rather a component of flow-proportional pressure support which can be combined with all conventional ventilatory modes, and provides a rational basis for subdividing the pressure support to overcome the mechanical load of the tubing and to overcome that of the respiratory system. Partial tube obstructions, which could decrease the effectivity of ATC, could be detected automatically by analysing the expiratory flow signal using a software, which could be easily implemented into the ventilator. The effectivity of ATC during long-term application can be maintained by intermittent short-term measurement of the tracheal pressure. Up to now there is no commercially available ventilator which allows complete expiratory ATC. Studies in volunteers and in mechanically ventilated patients have convincingly shown that ATC reduces work of breathing and increases respiratory comfort. In addition, successful extubation could be better predicted with this mode in difficult-to-wean patients compared to other modes. There are no special rules in the clinical application of ATC. However, to prevent overassist the support level of the ventilatory mode which is combined with ATC should be reduced.

Algorithms↗

Automatic tube compensation.

In this article automatic tube compensation (ATC) is described with respect to working principle, to technical realization, and to clinical experience. ATC, based on an indirect closed-loop working principle, compensates for the flow-dependent pressure drop across the tracheal tube during both inspiration and expiration. ATC reduces patient work of breathing, increases respiratory comfort, and allows prediction of successful extubation. ATC is not a stand-alone ventilatory mode, but rather a component of flow-proportional pressure support that can be combined with all conventional ventilatory modes.

Humans↗

Is pulmonary resistance constant, within the range of tidal volume ventilation, in patients with ARDS?

When managing patients with acute respiratory distress syndrome (ARDS), respiratory system compliance is usually considered first and changes in resistance, although recognized, are neglected. Resistance can change considerably between minimum and maximum lung volume, but is generally assumed to be constant in the tidal volume range (V(T)). We measured resistance during tidal ventilation in 16 patients with ARDS or acute lung injury by the slice method and multiple linear regression analysis. Resistance was constant within V(T) in only six of 16 patients. In the remaining patients, resistance decreased, increased or showed complex changes. We conclude that resistance within V(T) varies considerably from patient to patient and that constant resistance within V(T) is not always likely.

Adolescent↗

Volume-dependent compliance and ventilation-perfusion mismatch in surfactant-depleted isolated rabbit lungs.

OBJECTIVE: Volume-dependent alterations of lung compliance are usually studied over a very large volume range. However, the course of compliance within the comparably small tidal volume (intratidal compliance-volume curve) may also provide relevant information about the impact of mechanical ventilation on pulmonary gas exchange. Consequently, we determined the association of the distribution of ventilation and perfusion with the intratidal compliance-volume curve after modification of positive end-expiratory pressure (PEEP). DESIGN: Repeated measurements in randomized order. SETTING: An animal laboratory. SUBJECTS: Isolated perfused rabbit lungs (n = 14). INTERVENTIONS: Surfactant was removed by bronchoalveolar lavage. The lungs were ventilated thereafter with a constant tidal volume (10 mL/kg body weight). Five levels of PEEP (0-4 cm H2O) were applied in random order for 20 mins each. MEASUREMENTS AND MAIN RESULTS: The intratidal compliance-volume curve was determined with the slice method for each PEEP level. Concurrently, pulmonary gas exchange was assessed by the multiple inert gas elimination technique. At a PEEP of 0-1 cm H2O, the intratidal compliance-volume curve was formed a bow with downward concavity. At a PEEP of 2 cm H2O, concavity was minimal or compliance was almost constant, whereas higher PEEP levels (3-4 cm H2O) resulted in a decrease of compliance within tidal inflation. Pulmonary gas exchange did not differ between PEEP levels of of 0, 1, and 2 cm H2O. Pulmonary shunt was lowest and perfusion of alveoli with a normal ventilation-perfusion was highest at a PEEP of 3-4 cm H2O. Deadspace ventilation did not change significantly but tended to increase with PEEP. CONCLUSIONS: An increase of compliance at the very beginning of tidal inflation was associated with impaired pulmonary gas exchange, indicating insufficient alveolar recruitment by the PEEP level. Consequently, the lowest PEEP level preventing alveolar atelectasis could be detected by analyzing the course of compliance within tidal volume without the need for total lung inflation.

Animals↗

Breathing pattern and perception at different levels of volume assist and pressure support in volunteers.

OBJECTIVE: Volume assist (VA) amplifies the breathing effort whereas pressure support ventilation (PSV) provides a fixed, effort-independent ventilatory support. According to the concept of VA, its level should compensate for the pathologically increased (additional) elastance (Eadd). However, it is unclear whether breathing subjects prefer an exact compensation of Eadd and whether they are able to adjust the support level by themselves. DESIGN: Prospective, interventional study. SETTING: Laboratory. SUBJECTS: Twelve healthy volunteers, nine females, three males, aged 21-33 yrs. INTERVENTIONS: Increased Eadd was generated by banding of the thorax and abdomen. Volunteers breathed via a mouthpiece with VA or PSV using a positive end-expiratory pressure of 5 cm H2O (0.5 kPa). The study was subdivided into two parts. In part I, volunteers were instructed to adjust the level of VA and PSV themselves starting from three different, randomly applied levels in each mode (2, 8, 14 cm H2O or cm H2O/L; 0.2, 0.8, 1.4 kPa[/L]). In part II, 20 levels of VA and PSV (1-20 cm H2O or cm H2O/L, 0.1-2 kPa[/L]) were randomly selected by an investigator and estimated by the volunteers using a visual analog scale. Additionally, the breathing pattern was characterized. MEASUREMENTS AND MAIN RESULTS: Eadd (7.1 +/- 1.5 cm H2O/L [0.7 +/- 0.2 kPa/L], mean +/- sd) corresponded almost exactly to the "self-adjusted" VA level of part I (7.0 +/- 3.3 cm H2O/L [0.7 +/- 0.3 kPa/L]) and to the adequate level of part II (8-9 cm H2O/L [0.8-0.9 kPa/L]). The accordant PSV levels were 5.7 +/- 2.6 cm H2O (0.6 +/- 0.3 kPa) and 6-7 cm H2O (0.6-0.7 kPa). The breathing pattern was less influenced by changes of the support level with VA compared with PSV, which may explain in part the greater comfort of VA. CONCLUSIONS: We confirmed the theoretical assumption that VA should be adapted to Eadd. Furthermore, we demonstrated that conscious subjects are able to adjust the level of VA and PSV themselves.

Adult↗

CO(2)-controlled sampling of alveolar gas in mechanically ventilated patients.

A newly designed gas-sampling device using end-tidal CO(2) to separate dead space gas from alveolar gas was evaluated in 12 mechanically ventilated patients. For that purpose, CO(2)-controlled sampling was compared with mixed expiratory sampling. Alveolar sampling valves were easily controlled via CO(2) concentration. Concentrations of four volatile substances were determined in the expired and inspired gas. Isoflurane and isoprene, which did not occur in the inspired air, had ratios of end-tidal to mixed expired concentrations of 1.75 and 1.81, respectively. Acetone and pentane, found in both the inspired and expired air, showed ratios of 0.96 and 1.0, respectively. Precision of concentration measurements was between 2.4% (isoprene) and 11.2% (isoflurane); reproducibility (as coefficient of variation) was 5%. Because the only possible source of isoflurane and isoprene in this setting was patients' blood, selective enrichment of alveolar gas was demonstrated. By using the new sampling technique, sensitivity of breath analysis was nearly doubled.

Acetone↗

Dynamic respiratory system mechanics in infants during pressure and volume controlled ventilation.

Dynamic respiratory system mechanics can be determined using multiple linear regression (MLR) analysis. There is no need for a particular ventilator setting or for a special ventilatory manoeuvre. The purpose of this study was to investigate whether or not different ventilator modes and the flow-dependent resistance of the endotracheal tube (ETT) influence the determination of resistance and compliance by MLR. Ten paediatric patients who were on controlled mechanical ventilation for various disorders were investigated. The ventilator modes were changed between pressure control (PC) and volume control (VC). Flow and airway pressure were measured and tracheal pressure was continuously calculated. Each mode was applied for 3 min, and 10 consecutive breaths at the end of each period were analysed. Respiratory mechanics were determined by MLR based on either airway pressure, thus including the resistance of the ETT, or tracheal pressure. Resistance was found to be slightly higher in PC than in VC. There was no effect on determination of compliance between the different modes. Elimination of the flow-dependent resistance of the ETT preserved the differences between the modes. The authors conclude that using multiple linear regression compliance is not affected by the actual ventilator mode, whereas resistance is.

Airway Resistance↗

Detection of obstructive sleep apnea by analysis of phase angle using the forced oscillation signal.

Pharyngeal collapse in patients with obstructive sleep apnea syndrome (OSAS) is linked to decreased upper airway muscle activity. We hypothesised that decreased muscle activity causes decreased stiffness of the upper airway wall and assumed that a decrease in wall stiffness would result in a change point (CP) of the morphology of phase angle time series phi(t) obtained by forced oscillation technique (FOT). We developed an algorithm to detect CPs in &phi;(t) and investigated phi(t) data measured in parallel to all-night polysomnography in seven patients. A total of 2724 CPs were detected by algorithm. The CPs were marked on the polysomnograms and compared with polysomnogram scoring performed jointly by three sleep expert physicians. A total of 791 (67%) of the 1172 handscored respiratory events showed a CP in phi(t) within a time interval of 8 sec before and 6 sec after the marked onset. A total of 672 (57%) respiratory events were detected at an earlier time by algorithm, and 119 (10%) were algorithmically detected later. The average detection time by the CP algorithm was 4.0+/-1.1 sec earlier than the manual scoring. We conjecture that a CP in phi(t) indicates a change in upper airway collapsibility and that phi(t) might be a potentially useful parameter for detection of impending upper airway obstruction.

Adult↗

Breathing pattern associated with respiratory comfort during automatic tube compensation and pressure support ventilation in normal subjects.

BACKGROUND: Automatic tube compensation (ATC) is a new option to support spontaneously breathing tracheally intubated patients. We have previously demonstrated an increased respiratory comfort compared to pressure support ventilation (PSV) in volunteers. Here we characterized the breathing pattern during ATC associated with respiratory comfort in comparison to PSV. Furthermore, we studied whether ATC can be substituted by a simple modification of PSV. METHODS: We exposed 10 volunteers breathing through a 7.5 mm endotracheal tube via mouthpiece to PSV with 1) immediate and 2) delayed pressure rise and to 3) ATC. Immediate changes of the respiratory pattern after mode shifts were analyzed in detail. Furthermore, the volunteers were instructed to indicate changes in comfort after transitions between these modes as increased, unchanged, or decreased. RESULTS: Decreased comfort was associated with a substantial increase of tidal volume, minute ventilation, gas flow, and pressure. No differences in respiratory comfort were perceived between immediate and delayed pressure rise during PSV. CONCLUSION: PSV resulted in excessive tidal volumes and airflow, which was perceived as discomfort. This cannot be avoided by a delayed pressure rise but can be by the more comfortable ATC. ATC seems to adapt better to the ventilatory demand than PSV.

Adult↗

Total versus tube-related additional work of breathing in ventilator-dependent patients.

BACKGROUND: In tracheally intubated or tracheostomized spontaneously breathing patients, tube resistance can highly increase the patient's work of breathing. In this study we focused upon the relationship between total (WOBtot) and tube-related additional inspiratory work of breathing (WOBadd) and compared different ventilatory modalities for proper tube compensation. METHODS: In ten tracheostomized spontaneously breathing patients we measured WOBtot and WOBadd in the continuous positive airway pressure (CPAP) mode, under inspiratory pressure support of 5, 10, and 15 cmH2O in the pressure support ventilation (PSV) mode, and under flow-adjusted pressure support in the automatic tube compensation (ATC) mode. WOBadd and WOBtot were calculated on the basis of measured tracheal pressure and esophageal pressure, respectively. Inspiratory peak tracheal pressure above PEEP was taken as an estimate of pressure support beyond mere tube compensation (i.e., overcompensation). RESULTS: The percentage of the tube-related WOBadd on WOBtot in the CPAP mode was 52%. It decreased with increasing pressure support in the PSV mode from 32% (PSV 5 cmH2O) to 17% (PSV 15 cmH2O). WOBadd was only 15% of WOBtot in the ATC mode. In contrast to the other ventilatory modes, reduction of WOBadd in the ATC mode was achieved with the smallest amount of overcompensation, i.e. with minimal pressure assist beyond mere tube compensation. CONCLUSION: In tracheally intubated or tracheostomized spontaneously breathing patients, adequate compensation of tube resistance (i.e. with minimal overcompensation and minimal undercompensation) is best done by the ATC mode.

Aged↗