Rebreathing characteristics of the Bain anesthesia circuit.

The humidity output and CO2 elimination of the Bain circuit were tested on a simulated adult patient. The moisture content of inspired gases was found to be adequate when the circuit was used with a minute volume of 8.4 L/min and a fresh gas inflow of 4.9 L/min (65 percent relative humidity at room temperature at the onset of experimentation, rising to 100 percent after 80 minutes). However, the mean inspired CO2 concentration increased from 0.8 percent to 5.5 percent when the fresh gas inflow was decreased from 8 L/min to 3.5 L/min. It is recommended, therefore, that the circuit should not be used for long periods of time without measuring arterial CO2 tension or in situations where intentional hypocarbia is desired.

Adult↗

An infant circuit with a water vaporizer warmed by carbon dioxide neutralization.

A pediatric anesthesia system was constructed by connecting the machine end of a modified Mapleson D system to a soda lime canister containing a water vaporizer heated through the reaction of neutralization of the lime by the CO2 emerging from the breathing tube. The vaporizer was placed in line with the fresh-gas delivery tube of the circuit. The humidity output of the system varied from 15 +/- 1 mg H2O/L with a CO2 output of 15 ml/L to 27 +/- 1.5 mg H2O/L with a CO2 output of 60 ml/min. Rebreathing of CO2 was within acceptable limits as long as fresh-gas inflow remained at 5 L/min. The system compares favorably in respect to its humidity output with all other circuits availabe on the market for pediatric anesthesia.

Anesthesiology↗