Search PubMed⌕ Search

Biomedical subjects

H Bosmans

Publications and source records attributed to H Bosmans.

At least 19 recordsLinked to original sources

Evaluation of the diagnostic value of a computed radiography system by comparison of digital hard copy images with screen-film mammography: results of a prospective clinical trial.

The purpose of the study was to determine prospectively the diagnostic value of a computed radiography (CR) system by comparing mammographic hard copy images with screen-film mammography (SFM). A series of 100 patients, who came for diagnostic investigation, underwent two-view SFM (Lorad M-IV Platinum) and digital mammography with a CR system (AGFA CR system). The images were obtained by double exposure, i.e. same view without removing compression of the corresponding breast. The CR images were processed with dedicated processing for mammography. Six radiologists read sets of SFM and CR images. The primary efficacy parameter was the overall diagnostic value. The secondary efficacy parameters were lesion conspicuity and lesion details (for masses and micro-calcifications), tissue visibility at chest wall and at skin line, axillary details, overall density and sharpness impression and the overall noise impression. These parameters were scored by a 7-point scoring system. "CR non-inferior to SFM" was concluded if the lower confidence interval bound exceeded 80%. The confidence interval for the overall diagnostic value was between 96.4% and 100%. Pooled analysis of the ten features for image quality comparison demonstrated for all but one feature (lesion details of the calcifications) CR non-inferiority to SFM.

Adult↗

Characteristics of dosemeter types for skin dose measurements in practice.

A growing number of papers report deterministic effects in the skin of patients who have undergone interventional radiological procedures. Dose measurements, and especially skin dose measurements, are therefore increasingly important. Methods and acceptable dosemeters are, however, not clearly defined. This paper is the result of a literature overview with regard to assessing the entrance skin dose during radiological examinations by putting a dosemeter on the patient's skin. The relevant intrinsic characteristics, as well as some examples of clinical use of the different detector types, are presented. In this respect, thermoluminescence, scintillation, semiconductor and film dosemeters are discussed and compared with respect to their practical use.

Film Dosimetry↗

Relevant training issues for introduction of digital radiology: results of a survey.

Council Directive 97/43/Euratom establishes the need for adequate training of radiology staff. The transition to digital radiology implies changes in various imaging aspects, which are not sufficiently covered by current institutional training programmes. This work aimed to assess how professionals, experienced in digital imaging, acquired their expertise and hence, what form institutional training should take. Within DIMOND III, a survey on training and resources was performed among radiology professionals. A lack of institutional education for digital radiology was found. In the transition to digital radiology, 30-35% train on the job and another 23-28% receive training from digital equipment vendors. A general agreement exists on the need for new quality criteria and strategies for dose management. Issues relevant for conventional/digital transition are only sparsely covered in EC training programmes. Based on these results, a set of training issues was produced, to be included in future European guidelines.

Education, Medical, Continuing↗

Towards a proposition of a diagnostic (dose) reference level for mammographic acquisitions in breast screening measurements in Belgium.

A diagnostic reference level (DRL) is a dose level for a typical X-ray examination of a group of patients with standard body sizes and for broadly defined types of equipment. These levels are expected not to be exceeded for standard procedures when good and normal practice regarding diagnostic and technical performance is applied. In this paper, we have calculated DRLs for screening mammography in Belgium. The 95th percentile of the mean average glandular dose is 2.46 mGy. The DRL based on polymethyl methacrylate (PMMA) measurements was 2.08 mGy. Correlation coefficient (R) between doses from patient studies and phantom studies was 0.90, with an average underestimation of the phantom measurements of 15% for systems that use only Mo/Mo anode/filter. For the centres that use other anode/filters, there is not enough scientific evidence that a single phantom measurement of a standard PMMA block is representative for the patient dose.

Belgium↗

A study to validate the method based on DIMOND quality criteria for cardiac angiographic images.

A method based on image quality criteria (QC) for cine-angiography was developed to measure the quality of cine-angiograms (CA). A series of 30 CA for left ventriculography (LV) and left and right coronary angiography (LCA, RCA) have been scored and 172 readings were obtained. Standard deviation of quality scores indicated the reproducibility of the method. Each part of CA was examined separately, giving scores for LV, LCA and RCA and a total score (TS), with clinical (C) and technical (T) criteria defined and examined separately. In 83% of the studies TS was >0.8 and with standard deviation from 0.02 to 0.21. In general, LV had a lower score and greater disagreement compared with RCA and LCA. Disagreement was greater in T, compared with C. In conclusion, these results indicate that QC, translated into a scoring system, yields reproducible data on the quality of cardiac images.

Cineangiography↗

Image quality measurements and metrics in full field digital mammography: an overview.

This paper gives an overview of test procedures developed to assess the performance of full field digital mammography systems. We make a distinction between tests of the individual components of the imaging chain and global system tests. Most tests are not yet fully standardised. Where possible, we illustrate the test methodologies on a selenium flat-panel system.

Contrast Media↗

Current challenges of full field digital mammography.

Full field digital mammography (FFDM) has advantages over screen-film mammography (SFM), but some important challenges remain. The first challenge is related to the specific characteristics of FFDM. It remains unclear, which shape and limiting values of the modulation transfer function have the most influence on the performance of a detector, such as the effect of the image display on the overall image quality and the effect of processing on cancer detection. In order to assess the image quality of FFDM, we have set up a scoring system. The second challenge is related to screening mammography: is the quality of an image the same when it is viewed on different monitors and with different processing algorithms? Is Computer Aided Diagnosis necessary in a screening environment? In FFDM, the effect of different detectors, processing and display possibilities on the image and on cancer detection are not clearly investigated.

Breast Neoplasms↗

Data analysis from a multi-centre, comparative study of angiographic examinations leading to practical guidelines for the optimisation of patient doses.

In this study, patient doses were analysed against exposure parameters and procedure protocol. Patient doses were measured in seven hospitals for a standard diagnostic vascular examination. Image quality was assessed using the contrast-detail phantom from Nijmegen (CD-DISC 2.0). A link between dose and image quality was investigated. Image quality strongly depends on the preset dose level. The higher the dose level, the better the image quality, but also the higher the patient dose. However, no clear correlation was found between total dose-area product (DAP) or effective dose and image quality. A large range in patient dose was found: DAP (22-130 Gy cm2) and effective dose (3.9-16.8 mSv). A difference in number of frames was also found in the different centres, owing to different frame rates (3-2-1 frames per second) and use of oblique projections. Differences in doses and exposure settings offer the possibility of optimising patient doses in angiography and interventional radiology.

Angiography↗

On the use of DICOM cine header information for optimisation: results from the 2002 European DIMOND cardiology survey.

The paper explores the level of information contained within the DICOM header in images from various cardiology systems. Data were obtained in the European DIMOND survey on image quality (Italy, Ireland, Belgium, Greece and Spain). Images from five standard diagnostic cardiology procedures carried out in six European hospitals have been analysed. DICOM header information was extracted to a database in order to analyse how it could help in the optimisation of the procedures. The level of data contained in the headers differs widely between cardiology systems. None of the X-ray systems in the 2002 survey archives the dosimetric data in the DICOM header. The mean number of runs per procedure ranges between 7.5 and 15.4 and the mean number of frames per procedure between 575 and 1417. Differences in kVp, mA, pulse time, distances and C-arm angulations are substantial and suggest that there exists a wide range for optimisation.

Cardiology↗

Quality control of equipment used in digital and interventional radiology.

Digital and interventional radiology are increasingly important areas of radiology. Quality control (QC) of such equipment is of particular importance to avoid unnecessary high doses and to help to achieve good image quality. Within the DIMOND III project, equipment requirements and specifications for digital and interventional radiology have been formulated. A protocol for QC tests has been drafted based on various national and international recommendations. Tests are included for various parts of the imaging chain, i.e. X-ray tube and generator, X-ray tube control system, laser printer and display station, and image quality and patient dose. Preliminary tolerance levels have been set for the various tests, after initial measurements. To check the suitability of QC tests and stated tolerance levels, measurements were made at the University Hospital Gasthuisberg in Leuven for equipment used for paediatric radiology and a unit used for chest examinations. The results of the various tests are reported.

Humans↗

Image quality criteria in cardiology.

Image quality evaluation plays a key role in the process of optimisation in radiological procedures. Image quality criteria for cardiac cine-angiography were recently agreed as part of a European Research Project, and a scoring system based on these criteria has been developed to allow an 'objective' measurement of the quality of cardiac angiograms. Two studies aimed at the evaluation of the methodology have been completed, demonstrating that the method can be applied to cardiac images and translated into a scoring system that yields reproducible data. Based on the results of these studies, quality criteria have been further reviewed by DIMOND III panel and the updated version is presented in this paper.

Cardiology↗

Radiation dose levels during interventional cardiology procedures in a tertiary care hospital.

The aim of our investigation was to prospectively measure the patient radiation exposure from different cardiological procedures performed in the Catheterisation laboratory of the University Hospital Gasthuisberg in Leuven. The following local reference values were proposed: 40, 47 and 80 Gycm2 for coronary angiography (CA) or angioplasty (PTCA and stent implantation for elective patients), radio frequency ablation with angiographic images and CA plus ad hoc PTCA, respectively.

Angioplasty, Balloon, Coronary↗

Optimal beam quality selection in digital mammography.

An experimental method of determining the optimal beam quality for digital mammography systems was applied to two systems (Fuji Profect and GE Senographe 2000D). The mean glandular dose (MGD) and contrast-to-noise ratio (CNR) were measured using Perspex breast phantoms simulating breasts from 20 mm to 90 mm thick. For each thickness, four combinations of tube voltage and target/filter were tested. Optimal beam quality was defined as giving a target CNR for the lowest MGD and was similar for the two systems. For breasts with a thickness of 21 mm or 32 mm, a tube voltage of either 25 kV or 28 kV and a Mo/Mo target/filter combination was optimal. For breast thicknesses of 45 mm and greater, the combination that had the highest X-ray energy (34 kV Rh/Rh) was optimal. Optimization using the higher energy beam quality required greater detector dose to compensate for the lower contrast. Thus for a 75 mm thick breast the 34 kV Rh/Rh combination required about a 90% greater detector dose than 28 kV Mo/Mo to achieve the same CNR because of the 25% reduction in contrast. Nonetheless, the MGD was reduced by 32% by choosing the higher energy spectra and achieving the same CNR. Current automatic exposure control (AEC) designs that aim for a fixed detector dose are not optimal and greater use of higher energy spectra should be accompanied by higher detector doses at all breast thicknesses which are average or above. This may result in slightly higher doses, but better image quality for these breasts.

Breast↗

Skin dose measurements on patients for diagnostic and interventional neuroradiology: a multicentre study.

The present study reports on the skin dose measurements on patients undergoing diagnostic and interventional neuroradiology procedures. Grids of thermoluminescence dosemeters were attached on the head of the patient. The exposure parameters of the X-ray systems and the clinical procedures were fully documented. While for the diagnostic procedure, the dose threshold of 2 Gy for deterministic effects was not reached, this situation was much different for the embolisations. For angiography of the carotid arteries, no skin doses were measured >320 mGy. For the cerebral embolisations, maximum skin doses up to 5.4 Gy were measured. Doses to the patients varied largely between different hospitals and within the same hospital for similar procedures. On the one hand, the complexity of the pathology for interventional procedures was responsible for the large variability in dose. On the other hand, large differences in clinical protocol and technical parameters of the X-ray systems, explaining the dose variations, were also observed. A correlation was found between the maximum skin dose measured on a patient and the total dose-area product (DAP) value for cerebral embolisations. This correlation makes it possible to estimate the maximum skin dose from these DAP values and to determine a trigger level. In conclusion, management of patient doses in interventional radiology requires training, specialisation and well-documented procedure guidelines.

Brain↗

Effective doses in angiography and interventional radiology: calculation of conversion coefficients for angiography of the lower limbs.

The present study reports on investigations that we have performed to allow the calculation of effective doses (E) in interventional radiology. The use of published conversion tables might not allow sufficient guidance for the establishment of optimization strategies for procedures in interventional radiology. With the Monte Carlo N-Particle transport code (MCNP4B), conversion coefficients, linking dose-area product (DAP) measurements with E, are calculated for angiography of the lower limbs in six hospitals. The influence of various parameters on the calculation of these conversion coefficients is studied in a systematic way using the 2(n) factorial design. In this design the effect of different parameters and their pair-wise interactions on a certain variable is explored. In our study, the relevant parameters are tube potential, total filtration and field size and position. We concluded that the influence of radiation spectrum (kVp + filtration) is large and that the effect of field position and size is moderate, except when differences are observed in respect of the gonads. In that case, the variation in conversion coefficients is large. The results of this statistical analysis are then applied to the differences observed between the conversion coefficients, calculated for angiography of the lower limbs in the six hospitals. Recommendations for optimization of patient doses are given.

Adolescent↗

The value of fast MR imaging as an adjunct to ultrasound in prenatal diagnosis.

The aim of this study was to evaluate the role of MR imaging of the fetus to improve sonographic prenatal diagnosis of congenital anomalies. In 40 fetuses (not consecutive cases) with an abnormality diagnosed with ultrasound, additional MR imaging was performed. The basic sequence was a T2-weighted single-shot half Fourier (HASTE) technique. Head, neck, spinal, thoracic, urogenital, and abdominal fetal pathologies were found. This retrospective, observational study compared MR imaging findings with ultrasonographic findings regarding detection, topography, and etiology of the pathology. The MR findings were evaluated as superior, equal to, or inferior compared with US, in consent with the referring gynecologists. The role of these findings in relation to pregnancy management was studied and compared with postnatal follow-up in 30 of 40 babies. Fetal MRI technique was successful in 36 of 39 examinations and provided additional information in 21 of 40 fetuses (one twin pregnancy with two members to evaluate). More precise anatomy and location of fetal pathology (20 of 40 cases) and additional etiologic information (8 of 40 cases) were substantial advantages in cerebrospinal abnormalities [ventriculomegaly, encephalocele, vein of Galen malformation, callosal malformations, meningo(myelo)cele], in retroperitoneal abnormalities (lymphangioma, renal agenesis, multicystic renal dysplasia), and in neck/thoracic pathology [cervical cystic teratoma, congenital hernia diaphragmatica, congenital cystic adenomatoid lung malformation (CCAM)]. This improved parental counseling and pregnancy management in 15 pregnancies. In 3 cases, prenatal MRI findings did not correlate with prenatal ultrasonographic findings or neonatal diagnosis. The MRI provided a more detailed description and insight into fetal anatomy, pathology, and etiology in the vast majority of these selected cases. This improved prenatal parental counseling and postnatal therapeutic planning.

Adult↗

Radiofrequency ablation for eradication of pulmonary tumor in rabbits.

BACKGROUND: Radiofrequency ablation (RFA) has emerged as an alternative for surgery in clinical oncology. This animal experiment was conducted to evaluate the feasibility of RFA in the treatment of pulmonary tumor. METHODS: Eighteen rabbits with pulmonary implantation of VX2 tumors were divided into two groups. Group A (n = 12) was treated with RFA by using a cooled-tip electrode technique. Group B (n = 6) received sham operation. The therapeutic efficacy was evaluated by survival rate, magnetic resonance imaging (MRI), postmortem microangiography, and histology. RESULTS: All animals in group B died within 3 months after tumor implantation. Tumor eradication was achieved in 9 of 12 rabbits (75.0%) in group A, of which 4 rabbits survived longer than 3 months free of disease and another 5 rabbits were found free of viable tumor when sacrificed. One rabbit was subjected to incomplete tumor ablation and two rabbits suffered from local tumor relapse and/or lung metastasis. The 3-month survival rate of RFA-treated rabbits was significantly higher (P < 0.01) than that of control rabbits. The typical MRI appearances of the acute RFA lesion consisted of five characteristic concentric zones, which corresponded to central needle track (zone A), tumor coagulation (zone B), pulmonary parenchyma coagulation (zone C), peripheral hemorrhage (zone D), and inflammatory layer (zone E) on histology. CONCLUSIONS: Eradication of pulmonary tumor could be achieved with current RFA technique in rabbits. MRI is a useful modality for assessment of lung tumor ablation.

Angiography↗

Intracoronary delivery of Gd-DTPA and Gadophrin-2 for determination of myocardial viability with MR imaging.

The aim of this study was to compare intracoronary (i.c.) administration of Gadophrin-2, a necrosis-avid contrast agent (NACA), and nonspecific agent Gd-DTPA for determination of myocardial viability (MV) in acute myocardial infarction (AMI) with MRI. Reperfused AMI was induced in 12 dogs by transcatheter balloon occlusion of coronary artery. In 6 dogs each, Gd-DTPA at 0.1 mmol/kg or Gadophrin-2 at 0.005 mmol/kg was administered into coronary artery by fast bolus (n = 3) or slow infusion (n = 3). Serial ECG-triggered cardiac MRI of T1-weighted segmented turbo fast low-angle shot (FLASH) sequence was conducted and compared with triphenyltetrazolium chloride (TTC) histochemical staining. The contrast ratio and infarct size were quantified and analysed statistically. No cardiovascular side effects were found with local delivery of both agents. After i.c. administration, Gadophrin-2 induced a strong (CR > or = 1.78) and persistent (> or = 10 h) contrast enhancement of infarcted region. The infarct size defined with Gadophrin-2 was almost identical to that with TTC staining throughout the postcontrast period. With a dose 20 times higher, Gd-DTPA also strongly enhanced infarct-to-normal contrast; however, the enhancement diminished with time, i.e. from early strong to later faint enhancement and eventual loss of contrast. The delineated infarct size was also unstable, i.e. from early overestimation to later underestimation and eventual disappearance of the enhanced infarct. In combination with PTCA procedure, i.c. administration of MRI contrast agents may prove useful for post-procedure verification of diagnosis. The NACA-enhanced MRI may serve as an in vivo surrogate of postmortem histochemical staining for determination of MV. Although applicable in clinical setting, cardiac MRI with nonspecific Gd-DTPA is less reliable and should be performed within less than 1 h after contrast.

Animals↗