Selective protein removal and desalting using microchip CE.

This paper describes the on-line sample pretreatment and analysis of proteins and peptides with a poly(methylmethacrylate) (PMMA) microfluidic device (IonChip). This chip consists of two hyphenated electrophoresis channels with integrated conductivity detectors. The first channel can be used for sample preconcentration and sample clean-up, while in the second channel the selected compounds are separated. Isotachophoresis (ITP) combined with zone electrophoresis (CZE) was used to preconcentrate a myoglobin sample by a factor of about 65 before injection into the second dimension and to desalt a mixture of six proteins with 100 mM NaCl. However, ITP-CZE could not be used for the removal of two proteins from a protein/peptide sample since the protein zone in the ITP step was too small to remove certain compounds. Therefore, we used CZE-CZE for the removal of proteins from a protein/peptide mixture, thereby injecting only the peptides into the second CZE separation channel.

Animals↗

High throughput therapeutic drug monitoring of clozapine and metabolites in serum by on-line coupling of solid phase extraction with liquid chromatography-mass spectrometry.

The characteristics of automated on-line solid phase extraction with liquid chromatography-mass spectrometry (SPE-LC-MS) are very amenable for flexibility and throughput in therapeutic drug monitoring (TDM). We demonstrate this concept of automated, on-line SPE-LC-MS for the analysis of clozapine and metabolites (desmethylclozapine and clozapine-N-oxide) in serum. Method development, optimisation and validation are described and a comparison with previously published methods for the determination of clozapine and metabolites in serum and plasma is made. Optimisation of chromatographic and SPE conditions for increased throughput resulted in SPE-LC-MS cycle times of only about 2.2 min, demonstrating the great potential of automated on-line SPE-LC-MS for TDM. The new method is shown to be clearly favourable, in particular in terms of ease of sample handling, throughput and detection limits. Recovery is essentially quantitative. Detection limits are at about 0.15-0.3 ng ml(-1), depending on the ionisation source used. Calibration follows a quadratic model for clozapine and its N-oxide and a linear model for the desmethyl metabolite (all cases: R > 0.99). Accuracy, evaluated at three concentration levels spanning the whole therapeutic range, shows that bias is less than 10%. Precision (intra - and inter assay) ranges from about 5% R.S.D. at the high end of the therapeutic range (700-1,000 ng ml(-1)) to about 20% R.S.D. (OECD defined limit) at the lower limit of quantitation ( approximately 50 ng ml(-1)). The lower limit of quantitation is well below the low end of the therapeutic range at 350 ng ml(-1).

Antipsychotic Agents↗

Simultaneous resolution of overlapping peaks in high-performance liquid chromatography and micellar electrokinetic chromatography with diode array detection using augmented iterative target transformation factor analysis.

In this paper, augmentation has been applied to data matrices, which originate from hyphenated methods that share the same mode of detection, but use different separation methods, HPLC-DAD and MEKC-DAD. A novel method, wavelength shift eigenstructure tracking (WET), has been proposed for the alignment between the wavelength scale of both detectors. WET proves to be suitable for the detection as well as correction of wavelength shift between both detectors. After correction of the wavelength scale, data obtained on both systems have been augmented and submitted to iterative target transformation factor analysis. Augmented curve resolution provides significantly better estimates of the chromatographic and electrophoretic profiles and spectra than the use of non-augmented curve resolution on HPLC and MEKC data separately. It is particularly useful when the pure fraction of a chromatographic peak is less than 0.10. Finally, the relative weight of MEKC versus HPLC in augmentation may be increased using intensity and noise normalisation. However, since noise normalisation and its accompanying decrease in signal-to-noise ratio leads to a loss of information, and, since intensity normalisation may cause a failure of the augmented curve resolution algorithm, benefits and drawbacks of normalisation should be weighed on a case-by-case basis.

Algorithms↗

An improved coating for the isolation and quantitation of interferon-gamma in spiked plasma using surface plasmon resonance (SPR).

A study was initiated to investigate the use of surface plasmon resonance (SPR) for the detection in plasma of a high pI model protein, recombinant human interferon-gamma (IFN-gamma). Initially a number of self-assembled monolayers (SAMs) and hydrogel-derivatised SAM-coatings were characterised for the adsorptive and desorptive properties of plasma components. Next a monoclonal anti-IFN-gamma antibody, MD-2, was covalently attached to dextran-modified mercaptoundecanoic acid surfaces that performed best. On coatings consisting of carboxyl-modified dextran (CMD) a difference in interaction behaviour was observed when IFN-gamma was injected in either buffer or diluted plasma. During the injection of IFN-gamma in buffer, an acceleration of the interaction process was observed and the signal continued to increase after the injection plug had passed. Upon injection of diluted plasma spiked with IFN-gamma, the response increased without acceleration of the binding process. After the injection was finished, some of the bound material desorbed as expected, resulting in a signal decrease. On non-charged dextrans, the interaction between the antibody-modified surface and IFN-gamma in either plasma or buffer was similar. During sample injection the response increased with a binding rate depending on the concentration of IFN-gamma present in solution. When the injection was finished, some of the bound material was washed away from the surface and only a minor contribution of non-specific adsorbed plasma components was noticeable. From the coatings tested, the non-modified dextran-coated SPR sensor disks prove to be best suited for the detection of IFN-gamma in complex matrices like plasma. The interaction of IFN-gamma in both diluted plasma and buffer is comparable and concentrations of IFN-gamma of 250 ng ml-1 and higher can be detected in both buffer and 100x-diluted plasma. The non-specific adsorption of plasma components is low, whereas the specific IFN-gamma response is relatively high.

Animals↗

On the response of a label-free interferon-gamma immunosensor utilizing electrochemical impedance spectroscopy.

The research on our flow-injection, label-free, non-faradaic impedimetric immunosensor for interferon-gamma (IFN-gamma) has been extended. The sensor is prepared by immobilization of anti-IFN-gamma antibodies on a self-assembled monolayer (SAM) of acetylcysteine, deposited on polycrystalline gold. A multi-frequency impedance method is described, which allows time-resolved measurement of Nyquist plots. To these plots, an equivalent circuit was fitted, which is discussed in terms of a two-layer structure (inner and outer layer) of the interfacial region. Because binding of IFN-gamma mainly causes a decrease of Q (a constant-phase element), this element is considered as the outer layer. Several aspects of the impedimetric sensor response are studied, including the dependence on detection frequency, target concentration and applied dc potential. For quantitative detection of IFN-gamma, an optimum of the signal-to-noise (S/N) ratio of the out-of-phase impedance component (Z'') was found at about 100 Hz. At a dc-potential of +0.2 V versus a saturated calomel reference electrode, the sensor response is higher than at 0.0 V. Logarithmic dose-response curves of IFN-gamma in the concentration range of 10(-18) to 10(-9) M were obtained using two procedures: by successive injections over a single electrode, and by using freshly prepared electrodes for each measurement. Using the latter method, the repeatability is impaired. The need for in situ complementary techniques for a correct interpretation of the studied parameters is discussed.

Antibodies↗

Optimisation of high-performance liquid chromatography with diode array detection using an automatic peak tracking procedure based on augmented iterative target transformation factor analysis.

An automated method for the optimisation of high-performance liquid chromatography is developed. First of all, the sample of interest is analysed with various eluent compositions. All obtained data are combined into one augmented data matrix. Subsequently, augmented iterative target transformation factor analysis performs the integrated tasks of curve resolution and peak tracking. Since this type of curve resolution processes all data at once, it can deal with strong peak overlap and reveal the correspondence of compounds between runs, i.e. peak tracking. The retention time and peak width at half height for each component of the sample are determined for every eluent composition. Next, models are built for the retention time and the peak width at half height. These models are used to predict the resolution and the analysis time for each point in factor space. Finally, a multi-criterion decision-making method, Pareto optimality, is used to find the optimum. The method completes all calculations within a few minutes and without user intervention. By means of this procedure, a mixture of three benzodiazepines is successfully separated using a ternary mobile phase. There are two requirements for the automated optimisation method to work correctly. Firstly, all components of the sample must have sufficiently different spectra. Secondly, each compound should have the same spectrum under all experimental conditions.

Benzodiazepines↗

On-line coupling of solid-phase extraction with mass spectrometry for the analysis of biological samples. III. Determination of prednisolone in serum.

Solid-phase extraction (SPE) was directly coupled to mass spectrometry (MS) to assess the feasibility of the system for the rapid determination of prednisolone in serum. A C(18) stationary phase allowed washing of the cartridge with 25% methanol. Elution was performed by switching the methanol percentage from 25% in the washing step to 50% during elution. The high flow-rates during the extraction (5.0 ml/min) combined with ion-trap MS detection resulted in a total analysis time of 4 min. Some tailing of the prednisolone peak was observed. However, the tailing was found acceptable, since by this elution procedure most matrix compounds were prevented from eluting from the cartridge. Some matrix interference was still observed with a triple-quadrupole MS, even in the multiple reaction monitoring mode. This resulted in a detection limit (LOD) of about 10 ng/ml. The matrix interference and the LOD were similar for atmospheric pressure chemical ionisation and atmospheric pressure photo ionisation. Applying an ion-trap MS in the MS-MS mode resulted in cleaner chromatograms. Due to extensive fragmentation of prednisolone, the LOD was not lower than about 5 ng/ml prednisolone in serum, and a limit of quantitation of about 10 ng/ml (relative standard deviation <15%) was observed.

Glucocorticoids↗

Investigations into the stabilisation of drugs by sugar glasses: II. Delivery of an inulin-stabilised alkaline phosphatase in the intestinal lumen via the oral route.

In this study the possibility to deliver the acid-sensitive enzyme alkaline phosphatase (AP) from calf intestine (CIAP) to the intestinal system by oral administration was investigated. Tablets were prepared and in vitro evaluated. Final proof of concept studies were performed in rats. This acid labile enzyme is potentially useful in the treatment of sepsis, a serious condition during which endotoxins can migrate into the blood stream. The CIAP was freeze-dried with inulin and subsequently compacted into round biconvex tablets with a diameter of 4mm and a weight of 25-30 mg per tablet. The tablets were coated with an enteric coating in order to ensure their survival in the stomach. In vitro evaluation of tablets containing alkaline phosphatase from bovine intestine (BIAP) was the first step in the development. It was found that tablets without enteric coating dissolved rapidly in 0.10 M HCl with total loss of enzymatic activity of the alkaline phosphatase. Tablets that were coated were stable for at least 2 h in 0.10 M HCl, but dissolved rapidly when the pH was increased to 6.8. Furthermore, it was shown that the enzymatic activity of the released BIAP was fully preserved. The in vivo test clearly showed that the oral administration of enteric coated tablets resulted in the release of enzymatically active CIAP in the intestinal lumen of rats. The location of the enhanced enzymatic activity of AP in the intestines varied with the time that had passed between the administration of the tablets and the sacrificing of the rats. Also, the level of enzymatic activity increased with an increasing number of tablets that were administered.

Administration, Oral↗

Potential of capillary electrophoresis for the monitoring of the stability of placental alkaline phosphatase.

Alkaline phosphatase (AP) is a potential therapeutic agent in the treatment of sepsis. In this paper the potential of capillary zone electrophoresis (CZE) for the monitoring of the degradation of placental alkaline phosphatase (PLAP) was investigated. To induce degradation PLAP samples were exposed to high temperatures, low and high pH and freeze-drying. The samples were then analyzed by CZE and enzymatic activity assay. Upon exposure to temperatures above 65 degrees C, PLAP lost its activity exponentially over time, while CZE revealed both a linear decrease of the area of the main peak and a rise of degradation products. At acidic pH the enzyme appeared to lose its activity. CZE revealed a decrease of the area of the main peak, but no degradation products could be detected. At pH 12 the enzymatic activity and the area of the main peak both decreased linearly over time and, in addition, formation of degradation products could be detected by CZE. Activity and CZE profile of PLAP remained unchanged upon freeze-drying in the presence of inulin. Prolonged storage of freeze-dried samples at room temperature caused a slight decrease of enzymatic activity, while the potential formation of oligomers was revealed by CZE analysis. The examples in this study show that, in combination with activity assays, CZE can provide useful complementary information, especially on the status of the protein and the presence of degradation products.

Alkaline Phosphatase↗

Ion suppression in the determination of clenbuterol in urine by solid-phase extraction atmospheric pressure chemical ionisation ion-trap mass spectrometry.

Ion suppression effects were observed during the determination of clenbuterol in urine with solid-phase extraction/multiple-stage ion-trap mass spectrometry (SPE/MS(3)), despite the use of atmospheric pressure chemical ionisation. During SPE, a polymeric stationary phase (polydivinylbenzene) was applied. Post-cartridge infusion of analyte to the SPE eluate after the extraction of blank urine was performed to obtain a profile of the suppression. Single and multiple-stage MS were performed to provide insight in the suppressing compounds. The ion suppression was mainly ascribed to two m/z values, but still no identification of the compounds was achieved from the multiple-stage MS data. No ionisable and non-ionisable complexes and/or precipitation of clenbuterol with matrix compounds were observed. A concentration dependence of the percentage of suppression was observed. Up to 70% of the signal was suppressed upon post-cartridge infusion of 0.22 microg/mL (at 5 microL/min) clenbuterol into the eluate, and this decreased to about 4% at infusion of 22 microg/mL clenbuterol. Molecularly imprinted polymers were used to enhance the selectivity of the extraction. Although matrix components were still present after extraction, no interference of these compounds with the analyte was observed. However, the bleeding of the imprint from the polymer (brombuterol) caused significant ion suppression.

Animals↗

Investigations into the stabilisation of drugs by sugar glasses: I. Tablets prepared from stabilised alkaline phosphatase.

The aim of this study was to investigate the formulation of sugar glass stabilised alkaline phosphatase from bovine intestine (BIAP) into tablets. Two major subjects of tablet formulation were investigated. First, the compaction behaviour of the inulin sugar glass was investigated. Secondly, the effect of the compaction process on the physical stability of sugar glass stabilised BIAP was investigated, comparing inulin and trehalose glass. The tabletting properties of freeze-dried inulin without BIAP were studied first. Freeze-dried inulin conditioned at either 20 degrees C/0% relative humidity (RH) or 20 degrees C/45% RH was compacted at various pressures. As expected, the yield pressure of the material conditioned at 0% RH was higher (68 MPa) than after conditioning at 45% RH (39 MPa). Tablets made of the material stored at 0% RH showed severe capping tendency, especially at high compaction pressures. In contrast, material conditioned at 45% RH gave tablets without any capping tendency and a friability of less than 1%. Sugar glasses of BIAP and either inulin or trehalose were prepared by freeze-drying (BIAP/sugar 1/19 (w/w)). The material was subsequently compacted. Tablets and powders were stored at 60 degrees C/0% RH. The activity of the incorporated BIAP was measured at various time intervals. It was found that inulin was by far superior to trehalose as stabiliser of BIAP in tablets. The poor stabilising capacities of trehalose after compaction are explained by crystallisation of trehalose induced by the compaction process and moisture in the material. The results clearly show that inulin is an excellent stabiliser for BIAP. The tabletting properties are adequate, showing sufficient tablet strengths and low friability. Furthermore, the good (physical) stability of inulin glass with respect to exposure to high relative humidities makes it practical to work with.

Alkaline Phosphatase↗

Characterisation of reversed-phase stationary phases for the liquid chromatographic analysis of basic pharmaceuticals by thermodynamic data.

This paper describes the characterisation of reversed-phase liquid chromatography (RPLC) columns using thermodynamic measurements. Retention versus 1/T data were used to construct Van't Hoff plots. The slope of these plots indicates the standard enthalpy of transfer of the analyte from the mobile to the stationary phase. The standard entropy can be calculated from the intercept. Van't Hoff plots were linear for the investigated RPLC columns, meaning that for basic analytes over the temperature range studied no changes in the retention mechanism occurred. Enthalpies and entropies of transfer of basic analytes from the mobile to the stationary phase revealed information about the types of interaction of protonated and neutral compounds with the stationary phases. However, a clear view using the present set of basic compounds on how these thermodynamic data may explain the observed substantial differences in peak symmetry cannot be given. It is considered that addition of N,N-dimethyloctylamine (DMOA) to the eluent will results in a dynamically coating of the stationary phase. Addition of DMOA to the eluent resulted for protonated basic compounds in a reduction of both enthalpy and entropy. In practice, with DMOA in the eluent symmetrical peaks were obtained. It is assumed that this is due to blocking residual silanols and/or ion exclusion effects.

Chromatography, Liquid↗

Capillary electrophoresis with laser-induced fluorescence detection for fast and reliable apolipoprotein E genotyping.

The use of capillary electrophoresis (CE) with laser-induced fluorescence (LIF) detection for the rapid determination of apolipoprotein E (apoE) genotypes was studied. High resolution and sensitive detection of the concerned DNA restriction fragments was achieved using CE buffers with hydroxypropylmethylcellulose (HPMC) as sieving polymer and ethidium bromide (EB) as fluorescent intercalating agent. In order to achieve adequate resolutions in short analysis times, parameters such as concentration of HPMC and EB, separation voltage, and length and coating of the capillary were evaluated. Using a separation buffer with 0.8% (w/w) HPMC and 7 microM EB, characteristic DNA-fragment profiles could be obtained for all common apoE genotypes at an overall rate of ten samples per hour. The method allows direct injection of untreated PCR samples and the use of standard fused-silica capillaries which are effectively coated following a short, one-step rinse procedure. With a simple computerized algorithm based on migration-time ratios for pattern assignment, highly reliable apoE genotyping was achieved. Overall, in terms of speed, ease of use and objectivity the presented method provides a significant improvement over previously reported CE-based procedures for apoE genotyping.

Apolipoproteins E↗

Immunoaffinity chromatography for the sample pretreatment of Taxus plant and cell extracts prior to analysis of taxanes by high-performance liquid chromatography.

The application of immunoaffinity chromatography for the purification of Taxus plant and cell extracts prior to the HPLC analysis is described. Polyclonal antibodies raised against 10-deacetylbaccatin III (10-DAB III), paclitaxel's main precursor in plant, were characterised by enzymed-linked immunosorbent assay. Immunoglobulins from selected antisera were immobilised on CNBr-activated Sepharose 4B. The immunoaffinity column was used for the purification of plant and plant cell culture extracts prior to their analysis by HPLC. Immunoaffinity chromatography enabled the selective concentration of taxoids and enhanced sample clean-up.

Antibodies↗

Feasibility of the direct coupling of solid-phase extraction-pipette tips with a programmed-temperature vaporiser for gas chromatographic analysis of drugs in plasma.

Solid-phase extraction-pipette tips (SPE-PTs) were used for micro solid-phase extraction of lidocaine and diazepam from plasma. Off-line extraction was followed by on-line desorption. On-line desorption was carried out by direct coupling of the SPE-PTs with the liner of the programmed-temperature vaporiser. This coupling only required shortening of the liner by maximally 16 mm, cutting the SPE-PT, and equipping the remaining part with two O-rings. Due to the heating of the injector the SPE-PTs were heated as well, which resulted in a significant amount of impurities. Pre-heating and pre-washing was performed prior to the extraction to reduce the impurity level. The internal coupling device was applied successfully for the analysis of plasma samples with gas chromatography (GC) and mass-selective detection. Detection limits of 0.75 ng/ml and 2.5 ng/ml were obtained for lidocaine and diazepam, respectively, using 200 microl plasma. Recoveries for both compounds were about 80%. Although it is possible, the internal coupling device was not developed to be used as such. The main goal of this coupling was to show the feasibility of the integration of SPE-PTs with GC and to realize an important step to new automated SPE-GC systems.

Chromatography, Gas↗

On-fiber derivatization for direct immersion solid-phase microextraction. Part II. Acylation of amphetamine with pentafluorobenzoyl chloride for urine analysis.

On-fiber derivatization was used for solid-phase microextraction (SPME) in order to increase the detectability and extractability of drugs in biological samples. Amphetamine, which was used as a model compound, was derivatized with pentafluorobenzoyl chloride (PFBCl) and subjected to gas chromatography with electron capture or mass spectrometric detection. Extraction was performed by direct immersion of a 100 microm polydimethylsiloxane-coated fiber into buffered human urine. On-fiber derivatization was performed either after or simultaneously with extraction. The former procedure gave cleaner chromatograms but the latter turned out to be superior with respect to linearity and repeatability. For the on-fiber derivatization of amphetamine an excess of reagent is required. Because a considerable part of the PFBCl loaded on to the fiber is used up by reaction with matrix compounds and water, a reagent loading time of 5 min was needed to obtain a linear range (r = 0.9756) from 250 pg mL(-1) to 15 ng mL(-1). Due to an interfering matrix compound, the limit of detection was also found to be dependent on the reagent loading time, i.e., the limit of detection for a PFBCl loading time of 5 min is 250 pg mL(-1) whereas that for a 1 min loading time it is 100 pg mL(-1). The relative standard deviation (n = 7) of the method was about 11% at an amphetamine concentration of 1 ng mL(-1). The applicability of the method for the determination of drugs in biological samples is shown.

Acylation↗

Non-equilibrium solid-phase microextraction coupled directly to ion-trap mass spectrometry for rapid analysis of biological samples.

To determine sub-ppb levels of drugs in biological samples, selective, sensitive and rapid analytical techniques are required. This work shows the possibilities for high-throughput analysis of solid-phase microextraction (SPME) directly coupled to an ion-trap mass spectrometer equipped with an atmospheric pressure chemical ionisation source. As no chromatographic separation is performed, the SPME procedure is the time-limiting step. Direct immersion SPME under non-equilibrium conditions permits the determination of lidocaine in urine within 10 min. After a 5 min sorption time with a 100 microm polydimethylsiloxane-coated fibre, the extraction yield of lidocaine from urine is about 7%. When applying 4 min desorption, using a mixture of ammonium acetate buffer (pH 4.5) and acetonitrile (85 + 15 v/v), about 10% of the analyte is retained on the fibre. An extra cleaning step of the fibre is therefore used to prevent carry-over. By use of tandem MS, no matrix interference is observed. The detection limit for lidocaine is about 0.4 ng ml(-1) and the intraday and interday reproducibility are within 14% over a concentration range of 2-45 ng ml(1).

Humans↗

Characterisation of reversed-phase liquid chromatography stationary phases for the analysis of basic pharmaceuticals: eluent properties and comparison of empirical test methods.

The reversed-phase liquid chromatographic analysis of basic pharmaceuticals can be problematic. Both the properties of the eluent and the stationary phase can influence the chromatographic performance. Therefore selection of suitable experimental conditions for the analysis of basic compounds can be difficult. This paper shows that the organic modifier and the nature of the buffer influence the eluent properties. Moreover, the nature and amount of modifier also influence the basicity of the analytes. Investigations showed that the nature of the buffer can have a significant influence on retention and peak shape of basic compounds. Test procedures using basic analytes as test probes provided relevant information with respect to selecting columns to analyse basic pharmaceutical compounds. Test procedures using compounds like aniline, phenol and benzene were found to be less suitable.

Chromatography, High Pressure Liquid↗