[Technical feasibility of the implantation of a monorail stent system into the renal arteries without pre-dilatation].
PURPOSE: To evaluate the technical feasibility of the implantation of the monorail RX Herculink system into the renal arteries without pre-dilatation. MATERIALS AND METHODS: Forty-two patients (mean age 71 years) from four centers with a total of 44 renal artery stenoses underwent implantation of the RX Herculink stent. The mean grade of the stenosis was 83.8 %, the mean length 7.5 mm. The stenoses were ostial in 38 cases and in immediate proximity to the ostium in 6 cases. The mean follow-up-period was 57 weeks (24 - 176 weeks). RESULTS: In 42 cases, the implantation was successful without pre-dilatation. In 2 cases, pre-dilatation was carried out. In none of the cases, detachment of the stent from the balloon was observed. In one stenosis with a length of 17 mm, implantation of two stents was performed. In 9 cases, post-dilatation with a larger balloon or higher balloon pressure was necessary. Residual stenoses exceeding 30 % were not observed. Two patients developed local bleeding at the puncture site. During the follow-up, restenoses were observed in 5 stents after 26 to 126 weeks, which necessitated a second intervention in 3 cases (PTA in 2 cases, re-stenting in 1 case). The primary patency rate after 6 and 12 months was 0.92 +/- 0.056 according to Kaplan-Meier, the secondary patency rate after 6 and 12 months was 1.0 +/- 0.0. CONCLUSION: Implantation of the RX Herculink stent system into the renal arteries without pre-dilatation is technically feasible and safe. Even without pre-dilatation, the stent-system can be advanced through the stenosis without detachment. The complication rate is low. Our clinical results are comparable to previous studies.