Search PubMedSearch

Biomedical subjects

F A DiBianca

Publications and source records attributed to F A DiBianca.

5 recordsLinked to original sources

Space charge effects in a kinestatic charge detector.

The effects of space charge in a kinestatic charge detector (KCD) were examined using computer solutions to Poisson's equation. The KCD is a strip-beam parallel-plate drift chamber used for digital radiography. It was assumed that there is negligible electron attachment, i.e. there are no negative ions formed. The ionization rate per mA as a function of x-ray interaction depth was calculated for a detector filled with xenon at 25.3 x 10(5) Pa. Solution of Thomson's equations gave the positive ion density at the cathode, also as a function of depth. Water filtration values ranging from 0 to 30 cm were used in order to estimate the range of ion density values expected in a clinical KCD. The case of steady-state x-ray illumination was simulated for ionization rates less than the zero field limit (above which space charge changes the polarity of the electric field). Line spread responses were found for varying ionization rates to show the effect of space charge due to electric field distortion on the spatial resolution performance in the drift direction. The effect of imaging ideal edges with a KCD was calculated and the expected output signal was plotted for densities up to the zero field limit. Space charge dependence on the selection of KCD design and operating parameters is discussed. Because of the dependence of the KCD drift-direction spatial resolution on the uniformity of the electric field, space charge effects impose an upper limit on the detector entrance exposure and define the dynamic range of the device.

Humans

The use of contrast -- detail -- dose evaluation of image quality in a computed tomographic scanner.

Limiting-perception measurements of hole patterns in computed tomography reconstructions of a low-contrast phantom are used to generate contrast -- detail -- dose diagrams for various contrasts, hole diameters, dose, and system modulation transfer functions (MTF). These curves are then divided into MTF-dominated, transition, and noise-dominated regions. The low-contrast perceptibility and image noise are parameterized and compared. A limiting-perception factor is difined and is shown to be nearly independent of contrast, hole diameter, dose, and MTF in the noise-dominated region.

Data Display

Use of a low ionization potential dopant in a kinestatic charge detector: experimental findings.

The broadening of the line spread function (LSF) in the drift direction with increasing drift distance in the kinestatic charge detector is substantially reduced when small amounts (less than 1%) of trimethylamine [(CH3)3 N] are added to the x-ray detection medium (krypton or xenon). The LSF of a mixture of Kr and 0.01% trimethylamine (TMA) was measured as a function of distance at 15, 25, and 35 atm absolute pressure. The full width at half-maximum (FWHM) of the LSF was reduced from about 1.0 mm to less than 0.5 mm at a drift distance of 4.0 mm for the three pressures. The LSF's of mixtures of xenon and TMA at concentrations ranging from 0.0004% to 0.4% in one run and 0.06% to 4.0% in a second run were measured at a constant pressure of 20 atm. The FWHM of the LSF was reduced from 0.6 to 0.4 mm at 4.0 mm for the xenon measurements. The optimum concentration of TMA in Xe was found to be in the neighborhood of 0.1%. The use of TMA reduced the drift distance-dependent LSF broadening to the level expected from ionic diffusion, space charge repulsion, and electric field nonuniformity, and it may be possible to reduce the 0.4-mm FWHM plateau through the use of an improved Frisch grid design. Observation of negative charge carriers showed that electron attachment increases with increasing TMA concentration, although this could be caused by impurities in the TMA. The implications of these results are discussed in terms of extending the maximum drift distance attainable in a kinestatic charge detector.

Humans