Search PubMedSearch

Biomedical subjects

E H Bush

Publications and source records attributed to E H Bush.

5 recordsLinked to original sources

High-frequency ventilation: oscillatory dynamics.

OBJECTIVES: To determine the influence of the dynamic properties of the oscillator on the oscillatory volume delivered through the endotracheal tube to the lung or lung surrogate (delivered volume) under conditions of high-frequency ventilation. In particular, the relation between the tidal volume of the pump (oscillator) and the delivered volume was analyzed. PaCO2 was measured further as a function of the delivered volume in a number of experiments performed with healthy dogs. DESIGN: Laboratory study. SETTING: Engineering and animal laboratory. SUBJECTS: Lung surrogates and healthy dogs. INTERVENTIONS: An experimental oscillatory system was connected to various lung surrogates. In addition, six beagle dogs received high-frequency ventilation with different delivered volumes during the study. Control of the mean airway pressure was achieved by a peripheral pressure chamber located at the exhaust port of the bias flow tube. RESULTS: The delivered volume, which is the quantity of interest from a physiologic point of view, can deviate considerably from the tidal volume of the pump due to dynamic (particularly resonance) effects. Because the delivered volume and the mean airway pressure have to be controlled independently, two independent quantities are necessary for control purposes (e.g., the tidal volume of the pump and the mean pressure at the exhaust port). Furthermore, it was found that a minimal condition for adequate gas exchange is a delivered volume that exceeds the machine-related deadspace. For this reason, and in order to maximize the CO2 gradient, the exhaust tube must be as short as possible. CONCLUSIONS: a) The delivered volume has to be monitored under clinical conditions; b) however, because the impedance of the endotracheal tube in general considerably exceeds the impedance of the lung, the influence of the impedance of the lung on the delivered volume is generally small, and thus an in vitro calibration may serve as a useful approximation; c) at least two independent quantities are needed for an adequate oscillatory control; d) a necessary (not necessarily sufficient) condition for adequate CO2 removal is that the delivered volume must exceed the machine-related deadspace; e) in a clinical environment involving extremely pathologic lung conditions, e.g., adult respiratory distress syndrome, mechanical lung characteristics may deviate substantially from those characteristics used in this study (i.e., the results obtained may not necessarily be applicable under all clinical situations).

Animals

Augmentation of CO2 elimination during high frequency oscillation by removing the bias tube--an in vitro study.

In clinical applications of high frequency oscillation (HFO), sufficient CO2 elimination (VCO2) may represent a problem mainly at higher oscillation frequencies. With the intention of examining how to increase VCO2 a modified bias flow system was investigated in vitro with wash-out experiments. In bias flow systems, long tubes have been used in order to minimize the loss of oscillatory volume; however, a distinct increase of VCO2 was achieved in the present study by removing the bias tube. This improvement occurred over the whole frequency range of 2-60 Hz, although the oscillatory volume, effectively delivered to the lungs was smaller with the HFO circuit without bias tube (HFO-BT) as compared to the arrangement with bias tube (HFO + BT). A long bias tube flattens the CO2 concentration gradient from the alveoli to the atmosphere. Removing the bias tube results in a steeper CO2 concentration gradient and in a correspondingly enhanced VCO2. Furthermore, the large oscillatory volume at the exit of the bias flow system in HFO-BT supports VCO2 as an additional wash-out mechanism. Based upon longitudinal tracer gas concentration measurements between the alveoli and the atmosphere during HFO16,17, an increase of gas transport up to 20% can be expected for in vivo applications by removing the bias tube.

Carbon Dioxide

Lung surrogates.

In the development and evaluation of mechanical ventilation on the basis of high-frequency oscillation, appropriate surrogates of the lung are important, because they allow the measurement and control of various parameters which are not accessible in animal models. Yet, criteria have to be established according to which results obtained with a surrogate may be assessed with a view to extrapolation to humans. Theoretical considerations and impedance measurements are used for this purpose. It is found that for each given frequency a model can be made which exhibits realistic properties. However, no uniformly valid surrogate in the entire frequency range of 10-50 Hz is available at present.

Animals

Flow separation, an important mechanism in the formation of mean pulmonary pressure during high-frequency oscillation.

Mean pressures within the lungs and lung volume, respectively, are clinically important parameters. During ventilation by way of high-frequency oscillation (HFO), these parameters have been shown to be strongly frequency dependent. To identify mechanisms leading to mean pressure formation during HFO, findings of the theory of stationary flow were extended to oscillatory flow by a quasi-stationary approach. To confirm the theoretical findings, in-vitro experiments on HFO-models were performed. Flow separation was found to be an important mechanism in the formation of mean pressure. Flow separation causes a significant flow resistance, which may be distinctly different for in- and outflow. During oscillatory flow, a mean pressure difference thus results. This mechanism is of particular importance in bifurcations, which are present in the HFO-circuit as well as in the airways. With the direction-dependent flow separation, a general mechanism was found, which accounts for differing mean pressure values within the lungs with different HFO-circuits. This mechanism also contributes to interregionally different mean pressure values within the lungs.

High-Frequency Ventilation