Search PubMed⌕ Search

Biomedical subjects

D Z Kocovic

Publications and source records attributed to D Z Kocovic.

At least 19 recordsLinked to original sources

Unipolar defibrillator?

An unusual case of "unipolar" pacing and myopotential over-sensing leading to an inappropriate ICD shock in a patient with an implanted defibrillator is reported. The reasons for unipolar behavior in a system using a committed bipolar device are discussed.

Algorithms↗

Pacing alternans: an unusual pattern of pacemaker behavior.

We report an unusual pattern of pacemaker function related to the "autocapture" feature of a recently released pacemaker model. The electrocardiogram reveals pacing alternans. This report discusses the differential diagnosis and the correct explanation.

Algorithms↗

An electrocardiographic masquerade.

A 60-year-old woman underwent successful pulmonary embolectomy for a massive pulmonary embolism. On postoperative day 2, while receiving intravenous dopamine for hypotension, she developed the tachycardia, and a 12-lead ECG was obtained (Fig. 1). Figure 2 shows the 12-lead ECG obtained on the previous day, when she was in sinus rhythm with frequent premature

Bundle-Branch Block↗

Characteristics of electrograms recorded at reentry circuit sites and bystanders during ventricular tachycardia after myocardial infarction.

OBJECTIVES: The purpose of this study was to determine the relation of isolated potentials (IPs) recorded during ventricular tachycardia (VT) to reentry circuit sites identified by entrainment. BACKGROUND: Reentry circuits causing VT late after myocardial infarction are complex. Both IPs and entrainment have been useful for identifying successful ablation sites, but the relation of IPs to the location in the reentry circuit as determined by entrainment has not been completely defined. METHODS: Data from catheter mapping of 70 monomorphic VTs in 36 patients with prior myocardial infarction were retrospectively analyzed. Entrainment followed by radiofrequency current (RF) ablation was performed at 384 sites. On the basis of entrainment, sites were classified as reentry circuit exit, central-proximal, inner or outer loop sites. Sites outside the circuit were divided into remote and adjacent bystanders. RESULTS: Isolated potentials were recorded at 50% (51 of 101) of reentry circuit exit, central and proximal sites as compared with only 8% (11 of 146, p < 0.001) of inner loop and outer loop sites and only 1.8% (2 of 106) of remote bystander sites (p < 0.001). Isolated potentials were also present at 45% of adjacent bystander sites. At central and proximal sites the presence of an IP increased the incidence of tachycardia termination by RF to 47.5% from 24% (p = 0.05). At exit sites tachycardia termination occurred frequently regardless of the presence or absence of IPs (45% vs. 48%, p = NS). Isolated potentials at exit, central and proximal sites had a shorter duration at sites where ablation terminated VT than at sites without termination (20.9 +/- 9.6 ms vs. 35.7 +/- 15.3 ms, p < 0.001). CONCLUSIONS: Isolated potentials are a useful guide to sites in the central-proximal region of the reentry circuit, but often fail to identify exit sites where ablation is successful. Entrainment and analysis of electrograms provide complementary information during mapping of VT.

Aged↗

Histopathologic effects of radiofrequency catheter ablation in previously infarcted human myocardium.

INTRODUCTION: The use of catheter-based radiofrequency (RF) ablation for the treatment of ventricular tachyarrhythmias due to previous myocardial infarction has been steadily increasing. The histopathologic changes caused by this technique are not well described in humans. METHODS AND RESULTS: Three patients with hemodynamically tolerated ventricular tachycardias (VTs) due to previous myocardial infarction underwent endocardial mapping and catheter based RF ablation. All patients received between 5 and 11 RF lesions each of 60-second duration. One patient underwent myocardial resection of a left ventricular aneurysm 1 day following RF ablation, one expired 7 days after RF ablation, and one expired 9 months after RF ablation. None of the deaths occurred as a result of RF ablation. Pathologic specimens obtained early after RF ablation revealed areas of focal acute inflammation and fibrin deposition. Later specimens revealed several focal areas of fibrosis and granulation tissue. Specimens obtained late after RF ablation revealed a dense band of fibrosis, measuring 17 x 17 x 5 mm (1,250 mm3). CONCLUSION: Catheter-based RF ablation of ischemic VT in humans causes lesions that initially resemble coagulation necrosis. This is followed by the development of an inflammatory infiltrate and, finally, the development of fibrosis. Repeated application of RF ablation may result in much larger lesions than have been previously reported.

Aged↗

Cardiac death and stored electrograms in patients with third-generation implantable cardioverter-defibrillators.

OBJECTIVES: We sought to utilize terminal stored intracardiac electrograms (EGMs) to study the electrophysiologic events that accompany mortality in patients with third-generation implantable cardioverter-defibrillators (ICDs). BACKGROUND: Despite their ability to effectively terminate ventricular tachyarrhythmias, cardiac mortality in patients with ICDs remains high. The mechanisms and modes of death in these patients are not well understood. METHODS: We retrospectively analyzed clinical data and stored EGMs from patients enrolled in the clinical trial of the Ventritex Cadence ICD. Of the 1,729 patients 119 died during 6 years of follow-up. The final recorded EGM was reviewed. Postimplant EGMs as well as 50 control EGMs were used to define normal EGM characteristics. RESULTS: There were 36 noncardiac deaths (30%) and 83 cardiac deaths (70%). Of the cardiac deaths, 55 (66%) were nonsudden and 28 (34%) were sudden. When cardiac deaths were analyzed, 46 (55%) had no stored EGMs within 1 h of death, implying that the deaths were not directly related to tachyarrhythmias. In 37 cardiac deaths (18 nonsudden, 19 sudden), stored EGMs were present within 1 h of death. In these 37 deaths, the final EGM recorded was wide (>158 ms) in 33 (89%). Wide EGMs were interpreted as ventricular tachycardia in 27 and ventricular fibrillation in 6. In 13 of the 33 patients (39%) with wide EGMs, therapy was not delivered by the ICD, as it incorrectly detected a spontaneous termination of the arrhythmia. EGMs were significantly wider if recorded within 1 h, as compared with those recorded from 1 to 48 h before death (261+/-124 vs. 181+/-93 ms, p=0.04). CONCLUSIONS: Only 37 patients (31%) who died after placement of an ICD had a stored EGM within 1 h of the time of death, suggesting that the majority of deaths (69%) were not the immediate result of a tachyarrhythmia. When EGMs were recorded, they were wide in 89% of patients. These wide EGMs most likely represent intracardiac recordings of electromechanical dissociation. Thus, of the 119 deaths, 112 (94%) were not the immediate result of a tachyarrhythmia.

Aged↗

Catheter ablation of ventricular tachycardia after myocardial infarction: relation of endocardial sinus rhythm late potentials to the reentry circuit.

OBJECTIVES: We sought to determine whether endocardial late potentials during sinus rhythm are associated with reentry circuit sites during ventricular tachycardia (VT). BACKGROUND: During sinus rhythm, slow conduction through an old infarct region may depolarize tissue after the end of the QRS complex. Such slow conduction regions can cause reentry. METHODS: Endocardial catheter mapping and radiofrequency ablation were performed in 24 patients with VT late after myocardial infarction. We selected for analysis a total of 103 sites where the electrogram was recorded during sinus rhythm and, without moving the catheter, VT was initiated and radiofrequency current applied in an attempt to terminate VT. RESULTS: Late potentials were present at 34 sites (33%). During pace mapping, the stimulus-QRS complex was longer at late potential sites, consistent with slow conduction, than at sites without late potentials (p < 0.0001). Late potentials were present at 15 (71%) of 21 sites classified as central or proximal in the reentry circuit based on entrainment, but also occurred frequently at bystander sites (13 [33%] of 39) and were often absent at the reentry circuit exit (3 [23%] of 13). Late potentials were present at 20 (54%) of 37 sites where ablation terminated VT, compared with 14 (21%) of 66 sites where ablation did not terminate VT (p = 0.004). Ablation decreased the amplitude of the late potentials present at sites where ablation terminated VT. CONCLUSIONS: Although sites with sinus rhythm late potentials often participate in VT reentry circuits, many reentry circuit sites do not have late potentials. Late potentials can also arise from bystander regions. Late potentials may help identify abnormal regions in sinus rhythm but cannot replace mapping during induced VT to guide ablation.

Action Potentials↗

Sustained bundle branch reentry in a patient with hypertrophic cardiomyopathy and nondilated left ventricle.

Ventricular tachycardia is a well-known complication in patients with hypertrophic cardiomyopathy. We report the case of a patient with hypertrophic cardiomyopathy with easily inducible monomorphic ventricular tachycardia. Electrophysiology study demonstrated that bundle branch reentry was the mechanism of the tachycardia. The tachycardia was rendered non-inducible by radiofrequency ablation of the right bundle branch.

Adult↗

Autonomic dysfunction after catheter ablation.

Autonomic dysfunction may occur as a consequence of radiofrequency (RF) catheter ablation of a variety of supraventricular tachycardias. Effects suggestive of autonomic dysfunction that may be seen acutely during the ablation procedure include sudden profound slowing of the sinus rate or transient AV block. These abnormalities may occur during application of RF current, typically along the tricuspid or mitral annulus, at sites distant from both the sinus and AV nodes; they resolve quickly when RF current delivery is terminated. The most common long-term indication of autonomic dysfunction after ablation is inappropriate sinus tachycardia. This complication, rarely a lasting significant clinical problem, is seen after AV node modification and after ablation of accessory pathways. It usually resolves within several months. The mechanism appears to be loss of parasympathetic influence on the sinus node. Autonomic dysfunction after ablation of ventricular tachycardia has not yet been described, but could occur as newer catheter technologies capable of producing larger lesions are perfected.

Autonomic Nervous System↗