Hydrostatic and central venous pressure measurement by the IVAC 560 infusion pump.

The IVAC Variable Pressure Volumetric Pump, Model 560, was evaluated for hydrostatic and central venous pressure (CVP) measurement. In bench tests, 4 per cent of IVAC systems failed a +/- 2 mm Hg accuracy test. When measurements were inaccurate, errors were small (SD = 0.49 mm Hg). In simultaneous clinical measurements with a standard electronic transducer system, the correlation coefficient between the IVAC and transducer measurements was 0.95. A potential drawback of the IVAC 560 for CVP measurement is the lack of waveform display. The IVAC system is simple to operate and less expensive per use than the standard electronic system.

Central Venous Pressure↗

Accuracy of hydrostatic pressure measurement with a disposable dome transducer system.

The Gould-Statham P23id transducer with disposable dome was tested for accuracy between 0 and 20 mm Hg. With a criterion of +/- 2 mm Hg, 16 per cent of 74 systems failed. Inaccuracies were relatively large (SD, 6.02 mm Hg). The source of the unreliable measurements was poor sealing between the membrane of the disposable dome and the transducer diaphragm.