Mixing intensity modulated electron and photon beams: combining a steep dose fall-off at depth with sharp and depth-independent penumbras and flat beam profiles.

For application in radiotherapy, intensity modulated high-energy electron and photon beams were mixed to create dose distributions that feature: (a) a steep dose fall-off at larger depths, similar to pure electron beams, (b) flat beam profiles and sharp and depth-independent beam penumbras, as in photon beams, and (c) a selectable skin dose that is lower than for pure electron beams. To determine the required electron and photon beam fluence profiles, an inverse treatment planning algorithm was used. Mixed beams were realized at a MM50 racetrack microtron (Scanditronix Medical AB, Sweden), and evaluated by the dose distributions measured in a water phantom. The multileaf collimator of the MM50 was used in a static mode to shape overlapping electron beam segments, and the dynamic multileaf collimation mode was used to realize the intensity modulated photon beam profiles. Examples of mixed beams were generated at electron energies of up to 40 MeV. The intensity modulated electron beam component consists of two overlapping concentric fields with optimized field sizes, yielding broad, fairly depth-independent overall beam penumbras. The matched intensity modulated photon beam component has high fluence peaks at the field edges to sharpen this penumbra. The combination of the electron and the photon beams yields dose distributions with the characteristics (a)-(c) mentioned above.

Electrons↗

Verification of compensator thicknesses using a fluoroscopic electronic portal imaging device.

A method is presented for verification of compensator thicknesses using a fluoroscopic electronic portal imaging device (EPID). The method is based on the measured transmission through the compensator, defined by the ratio of the portal dose with the compensator in the beam and the portal dose without the compensator in the beam. The transmission is determined with the EPID by dividing two images, acquired with and without compensator inserted, which are only corrected for the nonlinear response of the fluoroscopic system. The transmission has a primary and a scatter component. The primary component is derived from the measured transmission by subtracting the predicted scatter component. The primary component for each point is only related to the radiological thickness of the compensator along the ray line between the focus and that point. Compensator thicknesses are derived from the primary components taking into account off-axis variations in beam quality. The developed method has been tested for various compensators made of a granulate of stainless steel. The compensator thicknesses could be determined with an accuracy of 0.5 mm (1 s.d.), corresponding to a change in the transmitted dose of about 1% for a 10 MV beam. The method is fast, accurate, and insensitive to long-term output and beam profile fluctuations of the linear accelerator.

Biophysical Phenomena↗

Dosimetric verification of intensity modulated beams produced with dynamic multileaf collimation using an electronic portal imaging device.

Dose distributions can often be significantly improved by modulating the two-dimensional intensity profile of the individual x-ray beams. One technique for delivering intensity modulated beams is dynamic multileaf collimation (DMLC). However, DMLC is complex and requires extensive quality assurance. In this paper a new method is presented for a pretreatment dosimetric verification of these intensity modulated beams utilizing a charge-coupled device camera based fluoroscopic electronic portal imaging device (EPID). In the absence of the patient, EPID images are acquired for all beams produced with DMLC. These images are then converted into two-dimensional dose distributions and compared with the calculated dose distributions. The calculations are performed with a pencil beam algorithm as implemented in a commercially available treatment planning system using the same absolute beam fluence profiles as used for calculation of the patient dose distribution. The method allows an overall verification of (i) the leaf trajectory calculation (including the models to incorporate collimator scatter and leaf transmission), (ii) the correct transfer of the leaf sequencing file to the treatment machine, and (iii) the mechanical and dosimetrical performance of the treatment unit. The method was tested for intensity modulated 10 and 25 MV photon beams; both model cases and real clinical cases were studied. Dose profiles measured with the EPID were also compared with ionization chamber measurements. In all cases both predictions and EPID measurements and EPID and ionization chamber measurements agreed within 2% (1 sigma). The study has demonstrated that the proposed method allows fast and accurate pretreatment verification of DMLC.

Algorithms↗

Multiple two-dimensional versus three-dimensional PTV definition in treatment planning for conformal radiotherapy.

PURPOSE: To demonstrate the need for a fully three-dimensional (3D) computerized expansion of the gross tumour volume (GTV) or clinical target volume (CTV), as delineated by the radiation oncologist on CT slices, to obtain the proper planning target volume (PTV) for treatment planning according to the ICRU-50 recommendations. MATERIALS AND METHODS: For 10 prostate cancer patients two PTVs have been determined by expansion of the GTV with a 1.5 cm margin, i.e. a 3D PTV and a multiple 2D PTV. The former was obtained by automatically adding the margin while accounting in 3D for GTV contour differences in neighbouring slices. The latter was generated by automatically adding the 1.5 cm margin to the GTV in each CT slice separately; the resulting PTV is a computer simulation of the PTV that a radiation oncologist would obtain with (the still common) manual contouring in CT slices. For each patient the two PTVs were compared to assess the deviations of the multiple 2D PTV from the 3D PTV. For both PTVs conformal plans were designed using a three-field technique with fixed block margins. For each patient dose-volume histograms and tumour control probabilities (TCPs) of the (correct) 3D PTV were calculated, both for the plan designed for this PTV and for the treatment plan based on the (deviating) 2D PTV. RESULTS: Depending on the shape of the GTV, multiple 2D PTV generation could locally result in a 1 cm underestimation of the GTV-to-PTV margin. The deviations occurred predominantly in the cranio-caudal direction at locations where the GTV contour shape varies significantly from slice to slice. This could lead to serious underdosage and to a TCP decrease of up to 15%. CONCLUSIONS: A full 3D GTV-to-PTV expansion should be applied in conformal radiotherapy to avoid underdosage.

Humans↗

Sharpening the penumbra of high energy electron beams with low weight narrow photon beams.

BACKGROUND AND PURPOSE: High energy (20-50 MeV) electron beams, available from the MM50 Racetrack Microtron, can be used for the treatment of deep-seated tumors. A disadvantage is the increasing penumbra width as a function of depth. By the addition of a narrow (typically 1 cm wide) photon beam near the field edge, the 50-90% penumbra width of the electron beam is reduced, yielding a significantly increased effective field size. MATERIALS AND METHODS: For rectangular electron beams in a water phantom (energies 25 and 40 MeV, field sizes 5 x 5-15 x 15 cm2) a computer program was used to optimize the photon beam parameters (position, weight and width) to obtain a combined beam with the sharpest penumbra at the optimization depth and a beam flatness within certain constraints. The study furthermore included penumbra sharpening of an irregular multileaf collimator-shaped field. RESULTS AND CONCLUSION: At optimization depths near R90, photon beam addition reduces the penumbra width by 40-50% (from 15-20 mm to 8-10 mm). Beam flatness at the optimization depth is within +/-5% and hot-spots are < or =120% for all depths. By the addition of narrow photon beams around the rectangular or irregular field, the electron field width can be reduced by 1-3 cm, while the effective field size is maintained.

Electrons↗

In vivo dosimetry for prostate cancer patients using an electronic portal imaging device (EPID); demonstration of internal organ motion.

PURPOSE: To investigate the use of a commercially available video-based EPID for in vivo dosimetry during treatment of prostate cancer patients. METHODS: For 10 prostate cancer patients, the inter-fraction variation within measured portal dose images (PDIs) was assessed and measured PDIs were compared with corresponding predicted PDIs based on the planning CT scan of the patient. RESULTS: For the lateral fields, the average standard deviation in the measured on-axis portal doses during the course of a treatment was 0.9%; for the anterior fields this standard deviation was 2.2%. The difference between the average on-axis measured portal dose and the predicted portal dose was 0.3+/-2.1% (1 SD) for the lateral fields and 0.7+/-3.4% (1 SD) for the anterior fields. Off-axis differences between measured and predicted portal doses were regularly much larger (up to 15%) and were caused by frequently occurring gas pockets inside the rectum of the patients during treatment or during acquisition of the planning CT scan. The detected gas pockets did sometimes extend into the gross tumour volume (GTV) area as outlined in the planning CT scans, implying a shift of the anterior rectum wall and prostate in the anterior direction (internal organ motion). CONCLUSIONS: The developed procedures for measurement and prediction of PDIs allow accurate dosimetric quality control of the treatment of prostate cancer patients. Comparing measured PDIs with predicted PDIs can reveal internal organ motion.

Follow-Up Studies↗

Leaf trajectory calculation for dynamic multileaf collimation to realize optimized fluence profiles.

An algorithm for the calculation of the required leaf trajectories to generate optimized intensity modulated beam profiles by means of dynamic multileaf collimation is presented. This algorithm iteratively accounts for leaf transmission and collimator scatter and fully avoids tongue-and-groove underdosage effects. Tests on a large number of intensity modulated fields show that only a limited number of iterations, generally less than 10, are necessary to minimize the differences between optimized and realized fluence profiles. To assess the accuracy of the algorithm in combination with the dose calculation algorithm of the Cadplan 3D treatment planning system, predicted absolute dose distributions for optimized fluence profiles were compared with dose distributions measured on the MM50 Racetrack Microtron and resulting from the calculated leaf trajectories. Both theoretical and clinical cases yield an agreement within 2%, or within 2 mm in regions with a high dose gradient, showing that the accuracy is adequate for clinical application.

Algorithms↗

Accurate portal dose measurement with a fluoroscopic electronic portal imaging device (EPID) for open and wedged beams and dynamic multileaf collimation.

Measuring portal dose with an electronic portal imaging device (EPID) in external beam radiotherapy can be used to perform routine dosimetric quality control checks on linear accelerators and to verify treatments (in vivo dosimetry). An accurate method to measure portal dose images (PDIs) with a commercially available fluoroscopic EPID has been developed. The method accounts for (i) the optical 'cross talk' within the EPID structure, (ii) the spatially nonuniform EPID response and (iii) the nonlinearity of the EPID response. The method is based on a deconvolution algorithm. Measurement of the required input data is straightforward. The observed nonlinearity of the EPID response was largely due to the somewhat outdated EPID electronics. Nonlinearity corrections for more modern systems are expected to be smaller. The accuracy of the method was assessed by comparing PDIs measured with the EPID with PDIs measured with a scanning ionization chamber in a miniphantom, located at the same position as the fluorescent screen. For irradiations in open, wedged and intensity modulated 25 MV photon beams (produced with dynamic multileaf collimation) EPID and ionization chamber measurements agreed to within 1% (1 SD).

Fluoroscopy↗

Portal dose image (PDI) prediction for dosimetric treatment verification in radiotherapy. I. An algorithm for open beams.

A method is presented for calculation of transmission functions for high energy photon beams through patients. These functions are being used in our clinic for prediction of portal dose images (PDIs) which are compared with PDIs measured with an electronic portal imaging device (EPID). The calculations are based on the planning CT-scan of the patient and on the irradiation geometry as determined in the treatment planning process. For each beam quality, the required input data for the algorithm for transmission prediction are derived from a limited number of measured beam data. The method has been tested for a PDI-plane at 160 cm from the focus, in agreement with the fixed focus-to-detector distance of our fluoroscopic EPIDs. For 6, 23 and 25 MV photon beams good agreement (approximately 1%) has been found between calculated and measured transmissions through anthropomorphic phantoms.

Algorithms↗

Field margin reduction using intensity-modulated x-ray beams formed with a multileaf collimator.

PURPOSE: In axial, coplanar treatments with multiple fields, the superior and inferior ends of a planning target volume (PTV) are at risk to get underdosed due to the overlapping penumbras of all treatment fields. We have investigated a technique using intensity modulated x-ray beams that allows the use of small margins for definition of the superior and inferior field borders while still reaching a minimum PTV-dose of 95% of the isocenter dose. METHODS AND MATERIALS: The applied intensity modulated beams, generated with a multileaf collimator, include narrow (1.1-1.6 cm) boost fields to increase the dose in the superior and inferior ends of the PTV. The benefits of this technique have been assessed using 3D treatment plans for 10 prostate cancer patients. Treatment planning was performed with the Cadplan 3D planning system (Varian-Dosetek). Dose calculations for the narrow boost fields have been compared with measurements. The application of the boost fields has been tested on the MM50 Racetrack Microtron (Scanditronix Medical AB), which allows fully computer-controlled setup of all involved treatment fields. RESULTS: Compared to our standard technique, the superior-inferior field length can be reduced by 1.6 cm, generally yielding smaller volumes of rectum and bladder in the high dose region. For the narrow boost fields, calculated relative dose distributions agree within 2% or 0.2 cm with measured dose distributions. For accurate monitor unit calculations, the phantom scatter table used in the Cadplan system had to be modified using measured data for square fields smaller than 4 x 4 cm2. The extra time needed at the MM50 for the setup and delivery of the boost fields is usually about 1 min. CONCLUSION: The proposed use of intensity modulated beams yields improved conformal dose distributions for treatment of prostate cancer patients with a superior-inferior field size reduction of 1.6 cm. Treatments of other tumor sites can also benefit from the application of the boost fields.

Humans↗

Dynamic multileaf collimation without 'tongue-and-groove' underdosage effects.

In all commercially available multileaf collimators, a 'tongue-and-groove'--or similar--construction is used for reduction of leakage radiation between adjacent leaves. These constructions can cause serious underdosages in intensity-modulated photon beams. A method for leaf trajectory calculation for dynamic multileaf collimation, which fully avoids these underdosage effects, is presented. The method is based on pairwise synchronizations of trajectories of adjacent leaf pairs, such that the delivered beam intensity in each 'tongue-and-groove' region is always equal to the smallest of the two prescribed intensities for the two corresponding leaf pairs. The effectiveness of the method has been proven for a large number of intensity-modulated fields, using the dynamic multileaf collimation mode of our MM50 Racetrack Microtron. Compared to dynamic multileaf collimation without synchronization, beam-on times are always equal or longer. For the cases that we studied, the beam-on time was typically increased by 5 to 15%.

Humans↗

Film dosimetry in water in a 23 MV therapeutic photon beam.

The field size and water depth dependence of the measured optical density of Kodak XV-2 film, irradiated in a 23 MV photon beam has been investigated. The films were positioned in a water tank in a vertical plane containing the beam axis with the upper film edge parallel to the water surface at a depth of 0.3 mm. The observed field size and water depth dependence of the film sensitivity cannot be fully attributed to the usual variation of the photon spectrum with field size and water depth: measured optical densities do significantly depend on the amount of film material above the point of measurement and on the film orientation. A method for application of film for relative water dose measurements in a plane containing the beam axis in a 23 MV therapeutic photon beam is presented; the observed agreement between film and ionisation chamber measurements is very good: typically within 1% or 2 mm.

Absorptiometry, Photon↗

Granulate of stainless steel as compensator material.

Compensators produced with computer controlled milling devices usually consist of a styrofoam mould, filled with an appropriate material. We investigated granulate of stainless steel as filling material. This cheap, easy to use, clean and re-usable material can be obtained with an average granule diameter of 0.3 mm, enabling an accurate and reproducible filling. No wax or other sealing material is added. The density of the granulate is approximately 4.5 g/cm3, which allows an accurate production of compensators in a sufficiently wide transmission range without the compensators becoming too thick. Transmission and surface dose measurements show that the dosimetric properties of stainless steel granulate are suitable for use as compensator material.

Radiotherapy Dosage↗

Daily dosimetric quality control of the MM50 Racetrack Microtron using an electronic portal imaging device.

The MM50 Racetrack Microtron, suited for advanced three-dimensional conformal radiotherapy techniques, is a complex machine in various respects. Therefore, for a number of gantry angles, daily quality control of the absolute output and fluence profiles of the scanned beams are mandatory. For the applied photon beams, a fast method for these daily checks, based on dosimetric measurements with the Philips SRI-100 Electronic Portal Imaging Device (EPID), has been developed and tested. Open beams are checked for four different gantry angles; for gantry angle 0, a wedged field is checked as well. Performing and analyzing the measurements takes about 10 min. The applied EPID has favourable characteristics for dosimetric quality control measurements: absolute output measurements reproduce within 0.5% (1 SD) and the reproducibility of relative (2D) beam profile measurements is 0.2% (1 SD). The day-to-day sensitivity stability over a period of one month is 0.6% (1 SD). Measured grey scale values are within 0.2% linear with the applied dose. The 2D fluence profile of the 25 MV photon beam of the MM50 is very stable in time: during a period of 5 months a maximum fluctuation of 2.2% has been observed. Once, a deviation in the cGy/MU-value of 6% was detected. There is no interlock in the MM50-system that would have prevented patient treatment with this strongly deviating output. Based on the results of this study and on clinical requirements regarding acceptability of deviations of beam characteristics, a protocol has been developed including action levels for additional investigations and, if necessary, adjustment of the beam characteristics.

Calibration↗

Portal dose measurement in radiotherapy using an electronic portal imaging device (EPID).

Physical characteristics of a commercially available electronic portal imaging device (EPID), relevant to dosimetric applications in high-energy photon beams, have been investigated. The EPID basically consists of a fluorescent screen, mirrors and a CCD camera. Image acquisition for portal dose measurement has been performed with a special procedure, written in the command language that comes with the system. The observed day-to-day variation in local EPID responses, i.e. measured grey scale value (EPID signal) per unit of delivered portal dose, is 0.4% (1 SD); day-to-day variation in relative EPID responses (e.g. normalized to the on-axis response) are within 0.2% (1 SD). Measured grey scale values are linearly proportional to transmitted portal doses with a proportionality constant which is independent of the thickness of a flat, water-equivalent absorber in the beam, but which does significantly depend on the size of the applied x-ray beam. It is shown that the observed increased in EPID response with increasing field size is mainly due to contributions to the EPID signals from scattered light: visible photons produced by the x-ray beam in a point of the fluorescent screen not only generate a grey scale value in the corresponding point of the EPID image, but also lead (due to scatter from components of the EPID structure onto the CCD chip) to an increased grey scale value at all other points of the image. A point spread function, derived from measured data and describing the increase in EPID response at the beam axis due to off-axis irradiation of the fluorescent screen, has been successfully applied to connect portal doses with grey scale values measured with the EPID.

Analysis of Variance↗