Validation of scanning laser Doppler flowmetry against single point laser Doppler flowmetry in the measurement of human gastric serosal/muscularis perfusion.
BACKGROUND: Single point laser Doppler (LDF) has been employed intra-operatively to measure gastro-intestinal tissue perfusion, but despite reports that it may have a role in predicting post-operative complications consequent upon hypo-perfusion, methodological drawbacks have prevented its routine clinical application. Scanning laser Doppler flowmetry (SLDF) may address these drawbacks, but this new development has not been previously validated in human gastro-intestinal tissue. AIMS: This study assessed the ability of Scanning laser Doppler flowmetry to measure human gastric perfusion. METHODS: 10 patients undergoing oesophageal resection were studied. In each 6 measurements of gastric corporal serosal/muscularis were made over 30 seconds using the previously validated single point LDF technique. A single 1 cm2 SLDF scan was made of the same area prior to any vascular division. These measurements were repeated after partial gastric devascularisation and the two techniques' measurements then compared. RESULTS: Correlation between the 2 methods was good (coefficient of correlation was 0.955 (p<0.01). The mean bias (scan - single point PUs) was -3 PUs (95% CIs 16.3 to 45 PUs) and the limits of agreement were -64 to 58 PUs. The coefficients of variation for the 6 sets of single point measurements made in each patient ranged between 3.0% and 30.4% (mean = 12.5%). CONCLUSION: These results validate SLDF as a method of measuring gastric tissue perfusion intra-operatively.